Navarre Opti-Drain NOD8PTRecalls & reported problems
Bard Access Systems, Inc. · Navarre Opti-Drain Multi-Use Drainage Catheter, Pigtail, 8 Fr x 30 cm · Class 1
All models of Navarre Opti-Drain (8)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00801741037108), received between Jan 15, 2020 and Jan 14, 2026 — classified as 10 malfunctions.
The most frequently reported specific device problem is “Fracture” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture5
- Defective Component4
- Fluid/Blood Leak4
- Nonstandard Device3
Patient problems
- No Clinical Signs, Symptoms or Conditions9
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 14, 2026Fracture
AN INTERVENTIONAL RADIOLOGIST INSERTED A NAVARRE DRAINAGE CATHETER AND CONNECTED A 3 WAY ADAPTER TO THE CATHETER HUB TO ALLOW ATTACHMENT TO A DRAINAGE BAG. THE FOLLOWING DAY, THE CATHETER HUB BROKE. SIMILAR INCIDENTS OF HUB BREAKAGE HAVE OCCURRED 2¿3 TIMES PREVIOUSLY. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24064958 ↗MalfunctionAug 29, 2025Defective Component; Defective Component
A PATIENT UNDERWENT A CHRONIC CATHETER PLACEMENT PROCEDURE USING THE NAVARRE OPTI DRAIN. IT WAS REPORTED THAT DURING THE PROCEDURE, CATHETER WAS ALLEGEDLY FOUND TOO LONG. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22920396 ↗MalfunctionAug 19, 2025Nonstandard Device; Defective Component; Nonstandard Device; Defective Component
A PATIENT UNDERWENT A DRAINAGE CATHETER PLACEMENT PROCEDURE BY USING NAVARRE DRAINAGE CATHETER. IT WAS REPORTED THAT DURING PREPARATION OF A NAVARRE OPTI DRAIN PLACEMENT PROCEDURE, THE DRAINAGE CATHETERS WAS ALLEGEDLY FOUND DIFFERENT LENGTH IN THE SAME BATCH. IT WAS FURTHER REPORTED THAT THE DRAINAGE CATHETERS WERE ALLEGEDLY TOO SHORT CANNOT BE USED FOR PATIENT OPERATION. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22838258 ↗MalfunctionAug 19, 2025Nonstandard Device; Defective Component; Nonstandard Device; Defective Component
A PATIENT UNDERWENT A DRAINAGE CATHETER PLACEMENT PROCEDURE BY USING NAVARRE DRAINAGE CATHETER. IT WAS REPORTED THAT DURING PREPARATION OF A NAVARRE OPTI DRAIN PLACEMENT PROCEDURE, THE DRAINAGE CATHETERS WAS ALLEGEDLY FOUND DIFFERENT LENGTH IN THE SAME BATCH. IT WAS FURTHER REPORTED THAT THE DRAINAGE CATHETERS WERE ALLEGEDLY TOO SHORT CANNOT BE USED FOR PATIENT OPERATION. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22838241 ↗MalfunctionAug 19, 2025Nonstandard Device; Defective Component; Nonstandard Device; Defective Component
A PATIENT UNDERWENT A DRAINAGE CATHETER PLACEMENT PROCEDURE BY USING NAVARRE DRAINAGE CATHETER. IT WAS REPORTED THAT DURING PREPARATION OF A NAVARRE OPTI DRAIN PLACEMENT PROCEDURE, THE DRAINAGE CATHETERS WAS ALLEGEDLY FOUND DIFFERENT LENGTH IN THE SAME BATCH. IT WAS FURTHER REPORTED THAT THE DRAINAGE CATHETERS WERE ALLEGEDLY TOO SHORT CANNOT BE USED FOR PATIENT OPERATION. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22838238 ↗MalfunctionOct 11, 2023Fracture; Fracture
IT WAS REPORTED THAT SOMETIME POST AN ULTRASOUND GUIDED PERCUTANEOUS TRANSHEPATIC CHOLANGIOGRAPHY DRAINAGE CATHETER PLACEMENT, THE DRAINAGE CATHETER CONNECTOR THAT CONNECTED WITH DRAINAGE BAG ALLEGEDLY FRACTURED. REPORTEDLY, THE CATHETER WAS REMOVED AND REPLACED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17910557 ↗MalfunctionMar 10, 2023Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT TWO DAYS POST DIALYSIS CATHETER IMPLANT, THE DEVICE ALLEGEDLY HAD LEAK. REPORTEDLY, THE CATHTER WAS REMOVED AND REPLACED WITH A NEW CATHETER. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16520346 ↗MalfunctionJan 17, 2022Fluid/Blood Leak; Fracture; Fluid/Blood Leak; Fracture
IT WAS REPORTED THAT SOME TIME POST CATHETER PLACEMENT UNDER ULTRASOUND-GUIDED PERCUTANEOUS RIGHT RENAL PELVIS PUNCTURE AND DRAINAGE, THE DRAINAGE TUBE INTERFACE WAS ALLEGEDLY CRACKED AND THE DRAINAGE FLUID WAS ALLEGEDLY LEAKED. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13266400 ↗MalfunctionJan 12, 2022Fluid/Blood Leak; Fracture; Fluid/Blood Leak; Fracture
IT WAS REPORTED THAT POST CATHETER PLACEMENT PROCEDURE, CATHETER WAS ALLEGEDLY CRACKED AND THE DRAINAGE FLUID WAS ALLEGEDLY LEAKED THROUGH THE CRACK. IT WAS FURTHER REPORTED THAT THE TEMPORARY TAPE WAS WRAPPED AROUND IT. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13234626 ↗MalfunctionJan 15, 2020Fluid/Blood Leak; Fracture; Fluid/Blood Leak; Fracture
THIS REPORT SUMMARIZES ONE (1) MALFUNCTION. A REVIEW OF THE REPORTED MALFUNCTION INDICATED THAT MODEL NOD8PT CHRONIC CATHETER ALLEGEDLY HAD A FLUID LEAK. THIS REPORT WAS RECEIVED FROM ONE SOURCE. THIS MALFUNCTION INVOLVED A PATIENT WITH NO PATIENT CONSEQUENCES. AGE, WEIGHT, AND GENDER WAS NOT PROVIDED FOR THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9593275 ↗
Other CATHETER, IRRIGATION devices
- Navarre Opti-Drain NOD8LPTBard Access Systems, Inc. · CATHETER, IRRIGATION901 reports
- Navarre NNU8LPTBard Access Systems, Inc. · CATHETER, IRRIGATION736 reports
- BLAKE 2227Ethicon Inc. · CATHETER, IRRIGATION331 reports
- Navarre NNU6LPTBard Access Systems, Inc. · CATHETER, IRRIGATION211 reports
- Navarre Opti-Drain NOD10LPTBard Access Systems, Inc. · CATHETER, IRRIGATION178 reports
- Navarre NNU10LPTBard Access Systems, Inc. · CATHETER, IRRIGATION127 reports
- BLAKE 2233Ethicon Inc. · CATHETER, IRRIGATION114 reports
- Vasoshield VH-5001Maquet Cardiovascular LLC · Catheter, Irrigation1 recall27 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page