Bardia® All-Silicone Foley Catheter 806514Recalls & reported problems

C. R. Bard, Inc. · Bardia® All-Silicone Foley Catheter, 5cc Balloon · Class 2

All models of Bardia All-Silicone Foley Catheter (13)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
10
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00801741039164), received between Oct 30, 2019 and Mar 20, 2025 — classified as 10 malfunctions.

The most frequently reported specific device problem is “Difficult to Remove” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 3 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult to Remove3
  • Failure to Infuse2
  • Component Misassembled1
  • Inflation Problem1
  • Material Puncture/Hole1
  • Material Rupture1
  • Material Split, Cut or Torn1
  • Obstruction of Flow1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • No Consequences Or Impact To Patient3
  • No Known Impact Or Consequence To Patient2
  • Pain1

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 20, 2025Material Rupture; Material Rupture

    IT WAS REPORTED THAT THE FOLEY CATHETER BALLOON BURSTS SHORTLY AFTER INSERTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21652972 ↗
  • MalfunctionFeb 7, 2025Material Puncture/Hole; Material Puncture/Hole

    IT WAS REPORTED THAT FOLEY CATHETER HAD A PIN HOLE AT THE Y CONNECTOR. SALINE WAS COMING OUT. LOT#NGJW3118. FOLEY CATHETER BALLOON WOULD NOT DEFLATE LOT#NGJT2607. BALLOON WOULD NOT DEFLATE LOT#NGJS5070.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21333374 ↗
  • MalfunctionFeb 7, 2025Difficult to Remove; Difficult to Remove

    IT WAS REPORTED THAT FOLEY CATHETER HAD A PIN HOLE AT THE Y CONNECTOR. SALINE WAS COMING OUT. LOT # NGJW3118. FOLEY CATHETER BALLOON WOULD NOT DEFLATE. LOT # NGJT2607. BALLOON WOULD NOT DEFLATE LOT# NGJS5070. PER CUSTOMER VIA EMAIL ON 04MAR2025, IT WAS REPORTED THAT LOT # NGJW3118 ITEM # 806514 CUSTOMER STATED THAT THE BARD FOLEY HAD A PINHOLE IN IT AT THE Y CONNECTOR. SALINE WAS CAME OUT. THEY WENT TO TEST THE FOLEY SECUREMENT, AND IT CAME OUT. BALLOON WAS DEFLATED. LOT# FOR THE AFFECTED 2, 1 WITH BALLOON WOULD NOT DEFLATE. LOT# NGJS5070. OR LOT# NGJT2607. NOT SURE WHAT PACKAGE EITHER FOLEY WAS PULLED OUT OF. THE OTHER ONE THAT CUSTOMER DID NOT KNOW WHAT WAS WRONG WITH IT WAS THE 5070 LOT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21333373 ↗
  • MalfunctionFeb 7, 2025Difficult to Remove; Difficult to Remove

    IT WAS REPORTED THAT FOLEY CATHETER HAD A PIN HOLE AT THE Y CONNECTOR. SALINE WAS COMING OUT. LOT#NGJW3118. FOLEY CATHETER BALLOON WOULD NOT DEFLATE. LOT#NGJT2607. BALLOON WOULD NOT DEFLATE LOT#NGJS5070. PER CUSTOMER VIA EMAIL ON (B)(6) 2025, IT WAS REPORTED THAT LOT# NGJW3118 ITEM# 806514 CUSTOMER STATED THAT THE BARD FOLEY HAD A PINHOLE IN IT AT THE Y CONNECTOR. SALINE WAS CAME OUT. THEY WENT TO TEST THE FOLEY SECUREMENT, AND IT CAME OUT. BALLOON WAS DEFLATED. LOT# FOR THE AFFECTED 2, 1 WITH BALLOON WOULD NOT DEFLATE. LOT# NGJS5070. OR LOT# NGJT2607. NOT SURE WHAT PACKAGE EITHER FOLEY WAS PULLED OUT OF. THE OTHER ONE THAT CUSTOMER DID NOT KNOW WHAT WAS WRONG WITH IT WAS THE 5070 LOT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21333372 ↗
  • MalfunctionJan 31, 2025Component Misassembled; Component Misassembled

    IT WAS REPORTED THAT THE FOLEYS CATHETER THAT SHIPPED ON JAN2024 WERE PAINFUL UPON INSERTION, TIPS NOT AS ROUNDED. NO MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: Pain

    MAUDE report 21282286 ↗
  • MalfunctionApr 30, 2024Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED THAT PATIENT RECEIVED THREE CATHETERS ON MOST RECENT ORDER, THE FIRST ONE WAS USED SUNDAY NIGHT AND ON MONDAY MORNING THE CATHETER HAD COME OUT, THEY STATED THERE WAS A PINHOLE AT THE TOP OF THE Y WHERE THE FOLEY MEETS THE Y. THE SECOND FOLEY INFLATED BUT WILL NOT RELEASE THE WATER. CAREGIVER WAS NOW ON THIRD CATHETER AND THIRD INSERTION TRAY DUE TO THESE ISSUES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19213781 ↗
  • MalfunctionSep 5, 2023Difficult to Remove; Difficult to Remove

    IT WAS REPORTED THAT THE FOLEY CATHETER BALLOON WOULD NOT DEFLATE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17682098 ↗
  • MalfunctionFeb 18, 2021Failure to Infuse; Failure to Infuse

    IT WAS REPORTED THAT THE CATHETER BALLOON DID NOT INFLATE.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient

    MAUDE report 11345913 ↗
  • MalfunctionMay 27, 2020Failure to Infuse; Inflation Problem; Failure to Infuse

    IT WAS REPORTED THAT THE BALLOON OF THE CATHETER WOULD NOT HOLD WATER.

    Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient

    MAUDE report 10092513 ↗
  • MalfunctionOct 30, 2019Partial Blockage; Obstruction of Flow; Partial Blockage

    IT WAS REPORTED THAT THE CATHETER BECAME CLOGGED AFTER 2 WEEKS OF USE, AND THE CUSTOMER WAS UNABLE TO FLUSH IT SO IT WAS REMOVED AND REPLACED. THE CUSTOMER NOTED THAT THE CATHETER BECAME CLOGGED FROM A BUILD OF THE INTERIOR COATING OF THE CATHETER.

    Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient

    MAUDE report 9257269 ↗

FDA clearance and classification

The FDA regulates product code GBM (catheter, urethral) under 21 CFR 876.5130 as a Class II device. 21 CFR 876.5130 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Bardia® All-Silicone Foley Catheter 806514

Has the Bardia® All-Silicone Foley Catheter 806514 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00801741039164), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Bardia® All-Silicone Foley Catheter 806514?

The FDA's MAUDE database lists 10 adverse event reports for this device, received between Oct 30, 2019 and Mar 20, 2025. In those reports, the most common specific device problems are “Difficult to Remove” (3), “Failure to Infuse” (2) and “Component Misassembled” (1); the most common patient problem is “Pain” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Bardia® All-Silicone Foley Catheter 806514?

None of the reports are classified as deaths, 0 as injuries and 10 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the Bardia® All-Silicone Foley Catheter 806514, and is it still sold?

The FDA's device database lists C. R. Bard, Inc. as the company responsible for it (brand Bardia® All-Silicone Foley Catheter, model 806514). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.