Bardex® All-Silicone Temperature-Sensing Pediatric Foley Catheter 119110Recalls & reported problems
C. R. Bard, Inc. · Bardex® All-Silicone Temperature-Sensing Pediatric Foley Catheter · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Bardex all-silicone temperature-sensing pediatric Foley catheter is a urine drainage catheter with a balloon and a temperature-monitoring cord for children. The 26 reports mostly describe the balloon not holding water or not deflating, urine flow blocked or slow, leaks, holes in the material, and the catheter coming out on its own; the temperature cords are also mentioned as not working with some monitors. Two reports involve injuries. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 26
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 26 adverse event reports that list this exact device (DI 00801741046001), received between Feb 9, 2018 and Aug 27, 2026 — classified as 2 injuries, 24 malfunctions.
The most frequently reported specific device problem is “Deflation Problem” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Deflation Problem5
- Inaccurate Flow Rate5
- Material Puncture/Hole4
- Fluid/Blood Leak3
- Leak/Splash3
- Material Invagination3
- Partial Blockage3
- Difficult to Remove2
Patient problems
- No Clinical Signs, Symptoms or Conditions11
- No Consequences Or Impact To Patient10
- No Known Impact Or Consequence To Patient9
- No Patient Involvement5
- Pain3
- Bacterial Infection1
- No Code Available1
Recent adverse event reports
The 25 most recent of 26 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 27, 2026Material Puncture/Hole
It was reported that after placement on a pediatric cardiology patient, the Foley catheter was found to have spontaneously dislodged. There was no fixation fluid, and although it had been secured with ample slack, it still dislodged spontaneously. The catheter was not placed for placement, and the patient was placed under observation.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50485336 ↗MalfunctionDec 22, 2025Complete Blockage; Complete Blockage
IT WAS REPORTED THAT PRIOR TO SURGERY, NO URINE FLOW WAS OBSERVED WHEN THE FOLEY CATHETER WAS INSERTED. DURING THE SURGERY, THEY CONTINUED TO MONITOR BUT STILL DID NOT OBSERVE ANY URINE FLOW. UPON REMOVING THE CATHETER AND EXAMINING IT, THEY CONFIRMED AN OBSTRUCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23872351 ↗MalfunctionAug 27, 2025Inflation Problem; Material Puncture/Hole; Inflation Problem; Material Puncture/Hole
IT WAS REPORTED THAT DURING A CUFF CHECK, ONLY SMALL AMOUNT OF WATER CAN BE PUMPED INTO THE FOLEY BALLOON, BUT ALMOST NO WATER CAN BE PUMPED IN, CAUSING THE WATER TO OVERFLOW FROM THE VALVE. THERE WAS A DIFFICULTY IN WATER INJECTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22904074 ↗MalfunctionNov 24, 2023Accessory Incompatible; Accessory Incompatible
IT WAS REPORTED THAT THE BLACK CORDS THAT WERE WITHIN THE 8F AND 10F SURESTEP FOLEY TRAY SYSTEM WERE NOT COMPATIBLE WITH THE MONITORS FOR TEMPERATURE MONITORING. THEY HAD CONFIRMED WITH CLINICAL ENGINEERING THAT THESE CORDS WERE NOT COMPATIBLE AND WOULD NOT WORK WITH THE DEVICES IN THE PEDS ED. THEY HAD ATTACHED PICTURES TO SHOW WHAT THE CONNECTOR NEEDED TO LOOK LIKE (WHITE CONNECTOR) THE OTHER WAS THE BLACK CORD THAT CAME WITH THE FOLEY KIT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18204751 ↗MalfunctionAug 10, 2023Material Invagination; Material Invagination
IT WAS REPORTED THAT THE FOLEY CATHETER BALLOON WAS DEFORMED DURING THE PRETEST. PER PHOTO SAMPLE RECEIVED FROM IBC ON 01AUG2023, IT WAS NOTED THAT THAT TWO LOT NUMBERS HAD THIS ISSUE (BATCH# NGGP3275 AND BATCH# NGGV5075) PER FOLLOW-UP INFORMATION RECEIVED FROM IBC ON 02AUG2023, CLARIFIED THAT THE BALLOON WRINKLED WHEN THE BALLOON WAS DEFLATED DURING PRETEST.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17499371 ↗MalfunctionApr 25, 2023Partial Blockage; Inaccurate Flow Rate; Partial Blockage; Inaccurate Flow Rate
IT WAS REPORTED THAT THE FOLEY CATHETER WAS UNABLE TO DRAIN URINE. FOLEY CATHETER WAS INSERTED FOR THE PURPOSE OF URINARY DRAINAGE DURING SURGERY. HOWEVER, FOLEY CATHETER WAS REMOVED BECAUSE THERE WAS NO URINE DRAINAGE AT ALL. FLUSHED SALINE SOLUTION INTO THE DRAINAGE LUMEN OF THE REMOVED FOLEY CATHETER, BUT SALINE SOLUTION COULD NOT BE DRAINED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16807872 ↗MalfunctionMar 30, 2023Material Puncture/Hole; Material Puncture/Hole
IT WAS REPORTED THAT THE FOLEY CATHETER SPONTANEOUSLY FELL OUT OF THE PATIENT DUE TO WATER LEAK.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16642856 ↗MalfunctionJan 23, 2023Partial Blockage; Difficult to Remove; Partial Blockage; Difficult to Remove
IT WAS REPORTED THAT THE LUMEN WAS OBSTRUCTED, AND NO OUTFLOW OF URINE COULD BE OBSERVED AFTER FOLEY CATHETER PLACEMENT. THE PLACEMENT OF THE CATHETER WAS CONFIRMED BY ECHOGRAPHY, AND THE TIP OF THE CATHETER WAS PLACED IN THE BLADDER. THE BALLOON WAS ALSO CONFIRMED TO BE INFLATED IN THE BLADDER. WHEN WATER WAS INJECTED THROUGH THE LUMEN, NO WATER ENTERED THE CATHETER, AND THE LUMEN WAS CONFIRMED TO HAVE BEEN OBSTRUCTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16223936 ↗MalfunctionSep 10, 2021Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material
IT WAS REPORTED THAT AN EYELASH-LIKE SUBSTANCE WAS FOUND ON THE TIP OF THE FOLEY CATHETER BEFORE USE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12456630 ↗MalfunctionFeb 8, 2021Partial Blockage; Partial Blockage
IT WAS REPORTED THAT NO URINE FLOW WAS OBSERVED ON THE DAY OF USE DUE TO CATHETER BLOCKAGE.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 11292249 ↗MalfunctionSep 30, 2020Deflation Problem; Material Invagination; Material Invagination
IT WAS REPORTED THAT IT WAS DIFFICULT TO DEFLATE DUE TO THE BALLOON ROLL. AS PER ADDITIONAL INFORMATION RECEIVED VIA EMAIL ON 15 SEP 2020 FROM IBC REPRESENTATIVE, IT SEEMED THAT THERE WAS NO VISIBLE DEFECT ON THE BALLOON.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 10607186 ↗MalfunctionAug 12, 2020Leak/Splash; Material Puncture/Hole; Device Fell; Material Puncture/Hole
IT WAS REPORTED THAT THE CATHETER FELL OUT. WHEN THE USER INFLATED THE CATHETER THAT HAD FALLEN OUT, THE CATHETER LEAKED AROUND THE VALVE. PER ADDITIONAL INFORMATION RECEIVED ON (B)(6) 2020, WATER LEAKED FROM THE CATHETER AND SEEMED THAT THERE WAS NO VISIBLE DAMAGE TO THE CATHETER. THE CATHETER WAS REPLACED TO THE PATIENT DUE TO A LEAK FROM THE VALVE.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 10396803 ↗MalfunctionSep 13, 2019Difficult to Remove; Difficult to Remove
IT WAS REPORTED THAT IT WAS DIFFICULT TO DEFLATE THE BALLOON DURING PRETEST.
Patient problems: No Patient Involvement
MAUDE report 9039093 ↗MalfunctionAug 21, 2019Crack; Deflation Problem; Fluid/Blood Leak; Deflation Problem
IT WAS REPORTED THAT A WATER LEAK OCCURRED WHEN THE USER INJECTED WATER INTO THE INFLATION VALVE. THE WATER LEAKED FROM A CRACK ON THE CATHETER.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 8917698 ↗MalfunctionJul 9, 2019Inaccurate Flow Rate; Inaccurate Flow Rate; No Flow
IT WAS REPORTED THAT THERE WAS NO URINE FLOW AFTER THE CATHETER WAS INSERTED INTO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 8773412 ↗MalfunctionMay 1, 2019Deflation Problem; Fluid/Blood Leak; Inflation Problem; Deflation Problem
IT WAS REPORTED THAT IT WAS DIFFICULT TO INFLATE THE BALLOON OF THE CATHETER ALLEGEDLY DUE TO WATER LEAKAGE. IT WAS LATER REPORTED PER ADDITIONAL INFORMATION RECEIVED FROM THE IBC VIA EMAIL (B)(6)2019 ; THE EVENT OCCURRED DURING PRETEST.
Patient problems: No Known Impact Or Consequence To Patient; No Patient Involvement
MAUDE report 8567965 ↗MalfunctionFeb 11, 2019Entrapment of Device; Device Damaged Prior to Use; Device Damaged Prior to Use
IT WAS REPORTED THAT THE THERMISTOR WIRE WAS SNAKED WITHIN THE LUMEN.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 8329180 ↗MalfunctionDec 12, 2018Deflation Problem; Deflation Problem; Fluid/Blood Leak
IT WAS REPORTED THAT WATER LEAKAGE OCCURRED FROM THE BALLOON DURING A PRETEST.
Patient problems: No Known Impact Or Consequence To Patient; No Patient Involvement
MAUDE report 8155290 ↗MalfunctionDec 12, 2018Deflation Problem; Deflation Problem; Leak/Splash
IT WAS REPORTED THAT WATER LEAKED FROM THE FUNNEL AREA OF THE CATHETER DURING THE PRETEST.
Patient problems: No Known Impact Or Consequence To Patient; No Patient Involvement
MAUDE report 8155113 ↗InjuryNov 30, 2018Material Invagination; Material Invagination; Material Deformation
IT WAS REPORTED THAT THE 10FR TEMPERATURE SENSING FOLEY CATHETER BALLOON DEVELOPED A CUFF WHEN THE BALLOON WAS DEFLATED. THE PATIENT ALLEGEDLY EXPERIENCED PAIN AND WAS GIVEN LIDOCAINE BY THE NURSE. THE ROUTE OF DELIVERY WAS UNKNOWN.
Patient problems: Pain
MAUDE report 8120588 ↗MalfunctionMay 4, 2018Inaccurate Flow Rate; Inaccurate Flow Rate; Infusion or Flow Problem
IT WAS REPORTED THAT URINE WAS NOT DRAINING THROUGH THE CATHETER.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 7487756 ↗MalfunctionApr 6, 2018Mushroomed; Mushroomed; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT THERE WAS A CUFF-ROLL ON THE BALLOON OF THE CATHETER. THE PATIENT EXPERIENCED SELF LIMITED PAIN.
Patient problems: No Consequences Or Impact To Patient; Pain
MAUDE report 7404508 ↗MalfunctionApr 6, 2018Mushroomed; Material Deformation; Mushroomed
IT WAS REPORTED THAT THERE WAS A CUFF-ROLL ON THE BALLOON OF THE CATHETER. THE PATIENT EXPERIENCED SELF LIMITED PAIN.
Patient problems: No Consequences Or Impact To Patient; Pain
MAUDE report 7404507 ↗MalfunctionMar 19, 2018Restricted Flow rate; Inaccurate Flow Rate; Restricted Flow rate; Inaccurate Flow Rate; Manufacturing, Packaging or Shipping Problem
IT WAS REPORTED THAT THERE WAS NO DRAINAGE EYE ON THE CATHETER. THEN THE CATHETER WAS REMOVED AND REPLACED
Patient problems: No Code Available; No Consequences Or Impact To Patient
MAUDE report 7349097 ↗MalfunctionMar 6, 2018Inaccurate Flow Rate; Inaccurate Flow Rate; Device Operates Differently Than Expected; No Flow
IT WAS REPORTED THAT THERE WAS NO DRAINAGE EYE ON THE CATHETER PRIOR TO USE.
Patient problems: No Known Impact Or Consequence To Patient; No Patient Involvement
MAUDE report 7317455 ↗
Other CATHETER, RETENTION TYPE, BALLOON devices
- Bardex® All-Silicone Foley Catheter 165816C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON512 reports
- Bardex® Lubri-Sil® All-Silicone Foley Catheter 175816C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON476 reports
- Bardex® Lubri-Sil® I.C. All-Silicone Foley Catheter 1758SI16C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON1 death383 reports
- Bardex® Lubri-Sil® All-Silicone Foley Catheter 175812C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON234 reports
- Bardex® Lubri-Sil® All-Silicone Temp Sensing Foley Catheter 119216MC. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON167 reports
- Bardex® Lubri-Sil® All-Silicone Foley Catheter 175814C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON153 reports
- Bardex® All-Silicone Foley Catheter 165812C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON151 reports
- SureStep™ Foley Tray System Bard® Lubri-Sil® Foley Catheter Tray A942216C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON2 recalls46 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page