Adverse event reports
10
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00801741047404), received between Oct 22, 2019 and Sep 16, 2024 — classified as 10 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (7 reports).

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 19, 2020Terminated

    Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that have developed breaks or cracks can cause leakage resulting in exposure to chemicals such a chemotherapeutics or biohazard material (e.g., blood). A crack or break could result in potential blood loss, catheter occlusion, air embolism, under or interrupted infusion, or site contamination which could lead to infection.

    Z-2670-2020 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Fluid/Blood Leak7
  • Break2
  • Crack2
  • Disconnection2
  • Detachment of Device or Device Component1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • No Consequences Or Impact To Patient6

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 16, 2024Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT THE PORT ACCESS NEEDLE POWERLOC MAX 22G X 0.75INCH (REFERENCE NUMBER (B)(4) ) HAD TWO LEAKS BELOW THE Y-SITE ON THE EXTENSION SET OF THE NEEDLE. BOTH WERE ACCESSED FOR 3 DAYS BEFORE THE Y-SITE BEGAN TO BREAK AND LEAK. IT WAS REPORTED THIS OCCURRED WITH TWO DEVICES. THIS REPORT ADDRESSES BOTH DEVICES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20224724 ↗
  • MalfunctionJun 3, 2021Crack; Fluid/Blood Leak; Crack; Fluid/Blood Leak

    IT WAS REPORTED "THREE PATIENTS HAVE RETURNED TO HAVE THEIR BOTTLE OF CHEMO AND PORT DE-ACCESSED; DRESSINGS APPEARED SLIGHTLY DAMP AND THE CRACKED AREA WAS THEN NOTED. JUST PRIOR TO, THE NURSE TRIED TO FLUSH AND THERE IS A SMALL HOLE THAT ALSO LEAKS."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11932017 ↗
  • MalfunctionMay 26, 2021Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED "THE PATIENT RETURNED TO HAVE THE BOTTLE OF CHEMO AND PORT DE-ACCESSED; DRESSINGS APPEARED SLIGHTLY DAMP AND THE CRACKED AREA WAS THEN NOTED. JUST PRIOR TO, THE NURSE TRIED TO FLUSH AND THERE IS A SMALL HOLE THAT ALSO LEAKS. ADD INFO RCVD 05/17/2021: PORT ACCESS APRIL 28. PORT DE-ACCESSED APRIL 30 WITH DAMP DRESSING. PORT - RIGHT CHEST. CATHETER SECUREMENT: DRESSING/SECURE WITH TAPE AS PER ROUTINE PRACTICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11886896 ↗
  • MalfunctionMay 26, 2021Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED "THE PATIENT RETURNED TO HAVE THE BOTTLE OF CHEMO AND PORT DE-ACCESSED; DRESSINGS APPEARED SLIGHTLY DAMP AND THE CRACKED AREA WAS THEN NOTED. JUST PRIOR TO, THE NURSE TRIED TO FLUSH AND THERE IS A SMALL HOLE THAT ALSO LEAKS." ADD INFO RCVD: 05/17/2021: PORT ACCESS APRIL 21 PORT DE-ACCESSED APRIL 23 WITH DAMP DRESSING. PORT - RIGHT CHEST. CATHETER SECUREMENT: DRESSING/SECURE WITH TAPE AS PER ROUTINE PRACTICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11886894 ↗
  • MalfunctionJan 29, 2021Break; Break; Disconnection

    IT WAS REPORTED THE TUBING BECAME UNATTACHED AT THE HUB CONNECTION LEAVING AN OPEN PATHWAY. NO OTHER INFORMATION WAS PROVIDED. ADDITIONAL INFORMATION RECEIVED 01/07/2021: WAS THERE PATIENT HARM REPORTED?: CAUSED RE-ACCESS OF A PEDIATRIC PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient

    MAUDE report 11251682 ↗
  • MalfunctionJan 20, 2021Break; Detachment of Device or Device Component; Break

    IT WAS REPORTED THE TUBING BECAME UNATTACHED AT THE HUB CONNECTION LEAVING AN OPEN PATHWAY. ADDITIONAL INFORMATION RECEIVED 1/7/2021: IT WAS REPORTED THE PEDIATRIC PATIENT HAD TO BE RE-ACCESSED BUT NO HARM WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient

    MAUDE report 11205585 ↗
  • MalfunctionJan 20, 2021Disconnection; Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THE TUBING BECAME UNATTACHED AT THE HUB CONNECTION LEAVING AN OPEN PATHWAY. ADDITIONAL INFORMATION RECEIVED 1/7/2021: IT WAS REPORTED THE PEDIATRIC PATIENT HAD TO BE RE-ACCESSED, BUT NO HARM WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient

    MAUDE report 11205584 ↗
  • MalfunctionApr 21, 2020Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED LEAKING AT THE Y SITE. NO OTHER INFORMATION WAS PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9983790 ↗
  • MalfunctionApr 21, 2020Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED LEAKING AT THE Y SITE. NO OTHER INFORMATION WAS PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9983788 ↗
  • MalfunctionOct 22, 2019Crack; Crack

    IT WAS REPORTED THAT THE FACILITY IS HAVING A LEAKING ISSUE WITH THE HUBER SAFESTEP 22 X ¾¿ NEEDLE. IT WAS STATED THE LEAKING SEEMS TO BE OCCURRING INSIDE THE HUB NOT AT THE CONNECTION POINT. (B)(6) 2019 ADDITIONAL INFORMATION RECEIVED FROM FACILITY STATED, THIS OCCURRED POST CHEMO. THE FRACTURE WAS STATED TO BE RUNNING LENGTH WISE ON HUB AT THE CLOSEST PORT TO THE PATIENT WHERE THEY DRAW LABS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9219656 ↗

Other Set, administration, intravascular devices

All Set, administration, intravascular devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page