Skip to content

Port Access Needle 0652010Recalls & reported problems

Bard Access Systems, Inc. · PowerLoc 20G x 1 inch · Class 2

All models of Port Access Needle (122)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
14
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 00801741047749), received between May 21, 2019 and Aug 20, 2024 — classified as 14 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (7 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fluid/Blood Leak7
  • Restricted Flow rate3
  • Crack2
  • Activation Problem1
  • Detachment of Device or Device Component1
  • Labelling, Instructions for Use or Training Problem1
  • Material Puncture/Hole1

Patient problems

  • No Clinical Signs, Symptoms or Conditions8
  • No Consequences Or Impact To Patient4
  • No Known Impact Or Consequence To Patient1
  • No Patient Involvement1

Recent adverse event reports

The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 20, 2024Labelling, Instructions for Use or Training Problem; Labelling, Instructions for Use or Training Problem

    IT WAS REPORTED THAT THEY NEED TO VERIFY A PRODUCT'S MRI STATUS AND MRI CONDITIONS. THE PRODUCT IS POWER LOC SAFETY INFUSION SET. THE LABELING INDICATED THE PRODUCT IS MRI CONDITIONAL BUT THE PACKET INSERT DOES NOT SPECIFY THE MRI CONDITIONS. CURRENTLY, THIS PRODUCT INDICATES IT IS MRI CONDITIONAL WITH THE YELLOW MRI TRIANGLE. WITHOUT TO DOCUMENTATIONS OF THE MRI CONDITIONS OUR FACILITY CAN NOT PROCEED WITH A PATIENT'S MRI WITHOUT THIS INFORMATION. NO OTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20002708 ↗
  • MalfunctionNov 3, 2023Activation Problem; Activation Problem

    IT WAS REPORTED BY CUSTOMER THAT SHOWED US THAT WHEN SHE GOES TO ACTIVATE THE SAFETY ON THE POWERLOC HUBER NEEDLE, IT DOES NOT COMPLETELY CLICK INTO PLACE UNLESS EXTREME FORCE IS USED. IT SEEMS THAT IT HAS BEEN ACTIVATED BUT IF ANY PRESSURE IS APPLIED THE NEEDLE COMES BACK THROUGH THE SAFETY AND COULD STICK SOMEONE. NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18067614 ↗
  • MalfunctionDec 23, 2022Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED BY THE CUSTOMER "FLUSH SITE WHEN PULLED BACK SHE GOT 5CC BLOOD AND THE REST WAS AIR. TO DO WASTE. THEN USE STERILE SYRINGE PULLED BACK AND THEN GOT MORE BLOOD BUT AIR AS WELL. DID THIS THREE TIMES AND HAPPEN EVERY TIME. THIRD TIME WAS WITHOUT CLAVE SO THEY RULED OUT THAT THE CLAVE WAS PROBLEM."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16044660 ↗
  • MalfunctionDec 9, 2022Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED BY THE CUSTOMER THAT "NURSE ACCESSED PORT, WHEN FLUSHING FOR THE FIRST TIME, SALINE CAME OUT OF SIDE OF THE DEVICE." NO OTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15949308 ↗
  • MalfunctionMay 7, 2022Restricted Flow rate; Restricted Flow rate

    IT WAS REPORTED VIA MS&S "CALLER STATES WHILE PERFORMING A CT OR CTA WITH THE PRESSURE SET AT 250 OR 300 THE MED RAD IS REACHING THE MAX PRESSURE WITH THE POWERLOC WHICH IS CAUSING THE RATE TO DECREASE. THE MED RAD IS SET TO AUTOMATICALLY DECREASE IF THE MAX PRESSURE IS MET FOR SAFETY REASONS." ADDITIONAL INFO. RECEIVED 04/13/2022: HIGH PRESSURE LIMITS ON MEDRAD INJECTOR (STELLANT) WHILE PERFORMING A ANGIO EXAM AT 4.5 ML A SEC AT 325 PSI LIMIT. THIS HAS BEEN CONSISTENTLY HAPPENING AT LOW AND HIGH RATES OF INJECTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14316468 ↗
  • MalfunctionApr 29, 2022Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED "I HAVE A HOME HEALTH NURSE WHO COMES TO MY HOME TO ACCESS ME & SHE ACCESSED ME TUESDAY & WHEN I HOOKED UP TO MY PUMP IS STARTED LEAKING OUT MY BANDAGE & SOAKED MY SHIRT."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14251636 ↗
  • MalfunctionApr 14, 2022Restricted Flow rate; Restricted Flow rate

    IT WAS REPORTED VIA MS&S "CALLER STATES WHILE PERFORMING A CT OR CTA WITH THE PRESSURE SET AT 250 OR 300 THE MED RAD IS REACHING THE MAX PRESSURE WITH THE POWERLOC WHICH IS CAUSING THE RATE TO DECREASE. THE MED RAD IS SET TO AUTOMATICALLY DECREASE IF THE MAX PRESSURE IS MET FOR SAFETY REASONS. HE STATES THEY HAVE USED 2 SEPARATE MED RAD INJECTORS [WITH] DIFFERENT PATIENTS." THIS REPORT ADDRESSES THE SECOND EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14116557 ↗
  • MalfunctionApr 14, 2022Restricted Flow rate; Restricted Flow rate

    IT WAS REPORTED VIA MS&S "CALLER STATES WHILE PERFORMING A CT OR CTA WITH THE PRESSURE SET AT 250 OR 300 THE MED RAD IS REACHING THE MAX PRESSURE WITH THE POWERLOC WHICH IS CAUSING THE RATE TO DECREASE. THE MED RAD IS SET TO AUTOMATICALLY DECREASE IF THE MAX PRESSURE IS MET FOR SAFETY REASONS. HE STATES THEY HAVE USED 2 SEPARATE MED RAD INJECTORS [WITH] DIFFERENT PATIENTS." THIS REPORT ADDRESSES THE FIRST EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14116555 ↗
  • MalfunctionJan 20, 2021Crack; Fluid/Blood Leak

    IT WAS REPORTED "THERE WAS A SMALL CRACK IN THE PLASTIC PIECE THAT COVERS THE PORT NEEDLE AND WHEN THEY WENT TO FLUSH, THE SALINE SPRAYED OUT OF THE CRACK."ON (B)(6) 2020- THE RETURNED DEVICE LEAKED FROM THE NEEDLE.

    Patient problems: No Patient Involvement

    MAUDE report 11202568 ↗
  • MalfunctionNov 19, 2020Crack

    IT WAS REPORTED THAT THE NEEDLE CRACKED AT THE Y-SITE. NO OTHER INFORMATION WAS PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10871987 ↗
  • MalfunctionAug 13, 2020Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED CHEMO STARTED THROUGH PORT 6ML INFUSED WHEN CALLED IN AND IT HAD LEAKED OUT AT TUBING/CAP CONNECTION. PATIENT HAD TO BE REACCESSED AND PROVIDER TO DECIDE IF CHEMO CONTINUES WITH 10% LOSS OR TO MAKE A NEW DOSE. PRODUCT CONTAINS BLOOD, NO PATIENT HARM.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10405210 ↗
  • MalfunctionMar 18, 2020Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT WHEN FLUSHING, FLUID LEAKED OUT OF WHERE NEEDLE AND TUBING CONNECT. WHEN ASPIRATING, TINY BUBBLES APPEARED IN SYRINGE. NO OTHER INFORMATION WAS PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9847435 ↗
  • MalfunctionMar 12, 2020Fluid/Blood Leak; Detachment of Device or Device Component; Fluid/Blood Leak; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE PATIENT WAS BEING ASSISTED OUT OF BED WHEN THE HUBER NEEDLE SNAPPED OFF BELOW THE CAP. IT WAS STATED THERE WAS A SLIGHT TUG WHILE GETTING THE PATIENT OUT OF BED. THEY WERE ABLE TO IMMEDIATELY DEACCESS THE PATIENT AND THEN REACCESS THE PORT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9823416 ↗
  • MalfunctionMay 21, 2019Material Puncture/Hole

    IT WAS REPORTED THAT A SMALL HOLE IN THE TUBING UNDER THE NEEDLE WAS OBSERVED WHEN FLUSHING PORT. NO OTHER INFORMATION WAS PROVIDED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8629882 ↗

Other Set, administration, intravascular devices

All Set, administration, intravascular devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page