Port Access Needle 0652034Recalls & reported problems
Bard Access Systems, Inc. · PowerLoc 20G x 0.75 inch · Class 2
All models of Port Access Needle (122)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This Bard PowerLoc needle is used to access implanted ports for chemotherapy and other infusions. Most of the 67 reports describe leaking at the hub during flushing, sometimes forcing staff to re-poke the patient, along with restricted flow, needles not activating their safety lock, and a few needlestick injuries to staff. Reports include 3 injuries and no deaths. There are no recalls naming this device, and reports fell from 7 in the prior 12 months to 4 in the last 12 months.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 67
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 67 adverse event reports that list this exact device (DI 00801741047763), received between Aug 22, 2018 and Sep 30, 2026 — classified as 3 injuries, 64 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (33 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Fluid/Blood Leak33
- Activation Problem10
- Restricted Flow rate8
- Contamination /Decontamination Problem4
- Fail-Safe Did Not Operate3
- Material Twisted/Bent3
- Break2
- Pressure Problem2
Patient problems
- No Clinical Signs, Symptoms or Conditions47
- Needle Stick/Puncture7
- No Known Impact Or Consequence To Patient5
- Exposure to Body Fluids2
- No Consequences Or Impact To Patient2
- No Patient Involvement2
- Chemical Exposure1
- Erythema1
Recent adverse event reports
The 25 most recent of 67 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026Activation Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50954904 ↗MalfunctionSep 30, 2026Activation Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50954899 ↗MalfunctionJun 23, 2026Component Missing
IT WAS REPORTED NURSING DISCOVERED NEEDLE WITH NO PROTECTIVE PLASTIC COVER ON NEEDLE WITHIN PACKAGE WHEN OPENING. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25602657 ↗InjuryMar 31, 2026Activation Problem
IT WAS REPORTED AFTER AN ASSOCIATE'S NEEDLE STICK, THE ASSOCIATE REPORTED SOME CONCERNS REGARDING THE DEVICE 20G HUBER NEEDLE IN THE PROCESS OF REMOVING FROM PORT-A-CATH. THE SAFETY FEATURE WAS PARTIALLY ACTIVATED, THE INJURY OCCURRED AFTER ACTIVATING THE SAFETY FEATURE. ASSOCIATE BELIEVES THAT ACTIVATING THE SAFETY FEATURE WAS MORE DIFFICULT THAN IT USUALLY IS. NEEDLE USUALLY SLIDES INTO THE CATHETER SLEEVE UPON REMOVING FROM PORT-A-CATH BUT IT DID NOT SLIDE ALL THE WAY LIKE IT WAS SUPPOSE TO. THE CLINICIAN SUSTAINED A NEEDLE STICK INJURY AFTER NEEDLE WAS REMOVED FROM IMPLANTED PORT. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: Exposure to Body Fluids; Needle Stick/Puncture
MAUDE report 24749950 ↗MalfunctionAug 1, 2025Activation Problem; Activation Problem
IT WAS REPORTED BY CUSTOMER THAT THE NEEDLES NOT ENGAGING THE SAFETY LOCK. IT WAS REPORTED THIS OCCURRED WITH FOUR DEVICES. THIS REPORT ADDRESSES ALL FOUR DEVICES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22690128 ↗MalfunctionAug 1, 2025Activation Problem; Activation Problem
IT WAS REPORTED BY CUSTOMER THAT THE NEEDLES NOT ENGAGING THE SAFETY LOCK.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22690116 ↗MalfunctionJul 16, 2025Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED "WE HAD A CONTRAST EXTRAVASATION WITH ONE NEEDLE THAT WAS NOTED TO BE BENT UPON REMOVAL FROM THE PORT." NO OTHER INFORMATION WAS PROVIDED.
Patient problems: Extravasation
MAUDE report 22522366 ↗MalfunctionJun 23, 2025Break; Break
IT WAS REPORTED BY THE CUSTOMER PATIENT'S PORT TUBING SNAPPED WHILE RECEIVING HOME INFUSIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22309656 ↗MalfunctionJun 23, 2025Break; Break
IT WAS REPORTED BY THE CUSTOMER PATIENT'S PORT TUBING SNAPPED WHILE RECEIVING HOME INFUSIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22309653 ↗MalfunctionMar 6, 2025Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED BY A CUSTOMER THAT A PARENT CALLED NURSE TO REPORT BLOOD LEAKING FROM PATIENT'S MEDIPORT. BLINATUMOMAB INFUSING. NURSE SAW THAT TUBING FROM MEDIPORT (ABOVE CAP) WAS CRACKED AND LEAKING BLOOD. UNABLE TO DRAW BACK BLOOD. MEDICATION PAUSED. PATIENT DE-ASSESSED AND REASSESSED. MEDICATION RESTARTED. NO HARM TO PATIENT. NO FURTHER INFORMATION PROVIDED.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 21536187 ↗MalfunctionNov 21, 2024Fluid/Blood Leak; Material Deformation; Fluid/Blood Leak; Material Deformation
IT WAS REPORTED, THAT PORT ACCESSED AND HUBER NEEDLE LEAKING AT CONNECTION SITE TO CAP. ATTEMPTED TO CHANGE CAP, BUT STILL LEAKING. PORT DE-ACCESSED AND NEW NEEDLER USE. NO PATIENT HARM. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20753116 ↗MalfunctionJul 9, 2024Contamination /Decontamination Problem; Contamination /Decontamination Problem
IT WAS REPORTED BY CUSTOMER THAT THE PORT NEEDLE WAS NOTED TO HAVE A HAIR/FIBER ON THE TIP OF THE NEEDLE UPON OPENING. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19707287 ↗InjuryJun 26, 2024Activation Problem; Activation Problem
IT WAS REPORTED RN RECOMMEND PATIENT PORT NEEDLE AT THE END OF IMMUNOTHERAPY TREATMENT. PORT WAS A POWERLOC BARD 20G 0.75INCH NEEDLE. RN PULLED BACK, ATTEMPTING TO HAVE NEEDLE LOCK VIA SAFETY FEATURE. THE NEEDLE DID NOT LOCK INTO THE SAFETY AND RN STUCK WITH NEEDLE. NEEDLES WERE CHANGED 4-6 MONTHS AGO. THEY SEEM TO BE A LITTLE MORE FLIMSY THAN OTHERS WE USED TO USE. NO HARM REPORTED. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: Needle Stick/Puncture
MAUDE report 19618020 ↗MalfunctionJun 6, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT A NURSE ON MY TEAM DISCONNECTED A PATIENT FROM THEIR 5FU PUMP AND NOTED THAT THERE WAS WHITE POWDER ON THE PATIENTS SKIN AROUND THE PORT. WHEN HE FLUSHED THE LINE, FLUID CAME OUT AT THE TOP OF THE NEEDLE (NEAR THE BUTTERFLY WINGS). HE THEN REMOVED THE DRESSING AND DEACCESSED THE PORT-A-CATH TO CLEAN THE SITE. UNFORTUNATELY, THE PATIENT DID HAVE SOME REDNESS AT THE PORT SITE THAT WE EXPECT WITH EXPOSURE TO 5FU. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: Chemical Exposure; Erythema
MAUDE report 19479931 ↗MalfunctionMay 2, 2024Fluid/Blood Leak; Contamination /Decontamination Problem; Fluid/Blood Leak; Contamination /Decontamination Problem
IT WAS REPORTED BY CUSTOMER THAT THE TWO NEEDLES THAT HAVE FLUID IN THE TUBING, NOTED UPON OPENING PRODUCT. IT WAS REPORTED THE LEAK WAS IDENTIFIED PRIOR TO PATIENT USE. IT WAS REPORTED THIS OCCURRED WITH TWO DEVICES. THIS REPORT ADDRESSES THE SECOND DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19229401 ↗MalfunctionMay 1, 2024Fluid/Blood Leak; Contamination /Decontamination Problem; Fluid/Blood Leak; Contamination /Decontamination Problem
IT WAS REPORTED BY CUSTOMER THAT THE TWO NEEDLES THAT HAVE FLUID IN THE TUBING, NOTED UPON OPENING PRODUCT. IT WAS REPORTED THE LEAK WAS IDENTIFIED PRIOR TO PATIENT USE. IT WAS REPORTED THIS OCCURRED WITH TWO DEVICES. THIS REPORT ADDRESSES THE FIRST DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19222058 ↗MalfunctionApr 26, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED BY THE CUSTOMER THAT POWERLOC PORT ACCESS NEEDLE IS LEAKING FROM THE TUBING. ADDITIONAL INFORMATION RECEIVED 4/3/24: IT WAS REPORTED BY THE CUSTOMER "THE POWERLOC NEEDLE WAS PRIMED WITH SALINE AND NO ISSUE WAS NOTED. THE PATIENT'S PORT WAS ACCESSED WITH THIS NEEDLE AND IT WAS NOTED THAT FOR ABOUT EVERY 3-4 MLS OF SALINE THAT WAS FLUSHED THROUGH, 1 DROP WOULD FORM ON THE TOP OF NEEDLE AND DRIP OUT WHERE IT WASN'T SUPPOSED TO. THE NEEDLE WAS REMOVED FROM THE PATIENT AND A NEW PORT NEEDLE WAS USED TO ACCESS THE PATIENT WITHOUT ISSUE. THIS COULD HAVE BEEN AN ADVERSE EVENT SINCE THE PATIENT RECEIVED CHEMOTHERAPY THAT DAY, THOUGH THE NEEDLE WITH THE DEFECT WAS REMOVED AFTER ONLY BEING ACCESSED INTO THE PATIENT FOR A FEW SECONDS." NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19195637 ↗MalfunctionApr 11, 2024
WHEN ATTEMPTING TO PRIME PORT NEEDLE, UNABLE TO PUSH FLUSH THROUGH THE LINE AS RESISTANCE WAS MET.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19087214 ↗MalfunctionMar 25, 2024Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED, ¿A 3/4-INCH NEEDLE BENT WHEN ACCESSING A PATIENT. THE RN ACCESSED THE PATIENT'S IMPLANTED PORT PER USUAL PROTOCOL, AND ENTERED IN THE MIDDLE OF THE PORT. ALTHOUGH THE NEEDLE WAS IN THE MIDDLE OF THE PORT, HOWEVER, IT DID NOT FLUSH. WHEN THE RN REMOVED THE HUBERT NEEDLE, SHE NOTED THAT THE NEEDLE WAS BENT. THE PATIENT EXPRESSED THAT THE ACCESS WAS MORE PAINFUL THAN USUAL, AND HE COMPLAINED THAT IT FELT SORE AFTER, WHICH IS NOT TYPICAL.¿ NO OTHER INFORMATION WAS PROVIDED.
Patient problems: Pain
MAUDE report 18975223 ↗MalfunctionFeb 14, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED ¿LOT# ASHUFC0005 THE ISSUE IS THE NEEDLE LEAKS OUT THE TOP BY THE ANGEL WINGS. OCCURS WHEN FLUSHING SALINE OR PULLING BACK BLOOD.¿ NO OTHER INFORMATION WAS PROVIDED. ADDITIONAL INFORMATION RECEIVED 01/25/2024: IT WAS REPORTED, ¿DISCOVERED WHEN PRIMING, NO HARM TO PATIENT. NO DAMAGE TO PACKAGING OR VISUALLY ABLE TO SEE, WHEN PRIMING OR IF YOU FLUSH AFTER INSERTION, LEAKS.¿ A SPECIFIC NUMBER OF AFFECTED DEVICES OR PATIENTS WAS NOT PROVIDED BY THE COMPLAINANT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18703024 ↗MalfunctionFeb 14, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED, "THE ISSUE IS THE NEEDLE LEAKS OUT THE TOP BY THE ANGEL WINGS. OCCURS WHEN FLUSHING SALINE OR PULLING BACK BLOOD." NO OTHER INFORMATION WAS PROVIDED. A SPECIFIC NUMBER OF AFFECTED DEVICES OR PATIENTS WAS NOT PROVIDED BY THE COMPLAINANT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18703023 ↗MalfunctionFeb 14, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED ¿LOT# ASHUFC0005 THE ISSUE IS THE NEEDLE LEAKS OUT THE TOP BY THE ANGEL WINGS. OCCURS WHEN FLUSHING SALINE OR PULLING BACK BLOOD.¿ NO OTHER INFORMATION WAS PROVIDED. ADDITIONAL INFORMATION RECEIVED 01/25/2024: IT WAS REPORTED, ¿DISCOVERED WHEN PRIMING, NO HARM TO PATIENT. NO DAMAGE TO PACKAGING OR VISUALLY ABLE TO SEE, WHEN PRIMING OR IF YOU FLUSH AFTER INSERTION, LEAKS.¿ A SPECIFIC NUMBER OF AFFECTED DEVICES OR PATIENTS WAS NOT PROVIDED BY THE COMPLAINANT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18703016 ↗MalfunctionFeb 14, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED, "THE ISSUE IS THE NEEDLE LEAKS OUT THE TOP BY THE ANGEL WINGS. OCCURS WHEN FLUSHING SALINE OR PULLING BACK BLOOD." NO OTHER INFORMATION WAS PROVIDED. A SPECIFIC NUMBER OF AFFECTED DEVICES OR PATIENTS WAS NOT PROVIDED BY THE COMPLAINANT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18703015 ↗MalfunctionFeb 10, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED, ¿POWERLOC SAFETY INFUSION SET HAVE LEAKED AT THE BASE OF THE NEEDLE. LUCKILY, THEY WERE CAUGHT WHILE PRIMING THE NEEDLE, PRIOR TO ACCESSING THE PATIENT.¿ NO OTHER INFORMATION WAS PROVIDED. IT WAS REPORTED THIS OCCURRED WITH TWO DEVICES. THIS REPORT ADDRESSES THE SECOND DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18682679 ↗MalfunctionFeb 10, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED, ¿POWERLOC SAFETY INFUSION SET HAVE LEAKED AT THE BASE OF THE NEEDLE. LUCKILY, THEY WERE CAUGHT WHILE PRIMING THE NEEDLE, PRIOR TO ACCESSING THE PATIENT.¿ NO OTHER INFORMATION WAS PROVIDED. IT WAS REPORTED THIS OCCURRED WITH TWO DEVICES. THIS REPORT ADDRESSES THE FIRST DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18682678 ↗
Other Set, administration, intravascular devices
- Autosoft XC 1001680Unomedical A/S · Set, Administration, Intravascular1 death16,105 reports
- Autosoft90™ 1002817Unomedical A/S · Set, Administration, Intravascular1 recall11,907 reports
- Minimed™ Quick-set™ MMT-399AUnomedical A/S · Set, Administration, Intravascular9 deaths8,669 reports
- Medtronic Extended MMT-431AUnomedical A/S · Set, Administration, Intravascular3 deaths8,574 reports
- MiniMed™ Mio™ Advance MMT-242AUnomedical A/S · Set, Administration, Intravascular15 deaths8,164 reports
- CADD 21-7302-24Icu Medical, Inc. · Set, administration, intravascular4 recalls1 death4,657 reports
- Autosoft XC 1001681Unomedical A/S · Set, Administration, Intravascular1 death4,216 reports
- Minimed™ Quick-set™ MMT-397AUnomedical A/S · Set, Administration, Intravascular19 deaths4,134 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.