Closed Wound Suction 0043600Recalls & reported problems
C. R. Bard, Inc. · CWS 400 Closed Wound Suction Kit, Small 3/32" (7 Fr., 2.3mm) O.D. PVC Drain, 12.5" (31.7cm) Hole Pattern, X-Ray Opaque Stripe, Attached Trocar, 400cc Evacuator, 3-spring, Y-connecting Tube · Class 1 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
Quick summary
This is a Bard Hemovac closed wound suction kit, a portable drain that uses negative pressure to pull fluid from a surgical incision. Most of the 35 reports describe the tubing coming apart or disconnecting from the connector, sometimes after the patient moved or was repositioned, along with cracks, leaks, small plastic fragments, and drain holes that appear to be in the wrong place along the tube. Four injuries are described and no deaths are reported. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 35
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 35 adverse event reports that list this exact device (DI 00801741049347), received between Apr 13, 2017 and Jul 16, 2026 — classified as 4 injuries, 31 malfunctions.
The most frequently reported specific device problem is “Disconnection” (15 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Disconnection15
- Break5
- Dull, Blunt3
- Material Fragmentation3
- Fluid/Blood Leak2
- Fracture2
- Incorrect Measurement2
- Contamination of Device Ingredient or Reagent1
Patient problems
- No Clinical Signs, Symptoms or Conditions23
- No Consequences Or Impact To Patient4
- No Known Impact Or Consequence To Patient3
- Foreign Body In Patient2
- No Patient Involvement2
- Device Embedded In Tissue or Plaque1
- Discomfort1
- Exposure to Body Fluids1
Recent adverse event reports
The 25 most recent of 35 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 16, 2026Incorrect Measurement
THE CUSTOMER REPORTED A BD HEMOVAC DRAINS POSSIBLE DESIGN OR MANUFACTURING CHANGE. THEY STATED THAT, IN THE PAST, THE DRAIN PERFORATIONS EXTENDED TO THE END OF THE TUBING; HOWEVER, RECENTLY RECEIVED PRODUCTS HAD PERFORATIONS LOCATED IN THE MIDDLE SECTION OF THE TUBING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25849848 ↗MalfunctionOct 1, 2025Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT WHEN WOUND EVACUATOR WAS USED, A LEAK WAS CONFIRMED. THE LOCATION OF THE LEAK WAS CURRENTLY UNKNOWN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23191104 ↗MalfunctionJul 3, 2025Disconnection; Disconnection
IT WAS REPORTED THAT THE PATIENT HAD HEMOVAC DRAIN CLOSED WOUND SUCTION EVACUATION DEVICE TO THE BACK INCISION NEAR THE SPINE. HEMOVAC DRAIN WAS TAPED AND PREVIOUSLY WORKING. WITH REPOSITIONING DEVICE HAD DISCONNECTED AT THE TUBING AND IMMEDIATELY ADDRESSED. DEVICE WAS REMOVED FROM PATIENT AS PER ORDER PROMPTLY. PATIENT HAD SURGERY LIKE T10 - S1 INSTRUMENTATION AND FUSION OF SPINAL COLUMNS. DRAIN WAS USED TO REMOVE EXCESS DRAINAGE FROM SURGERY SITE. PER ADDITIONAL INFORMATION RECEIVED VIA EMAIL ON 09JUL2025, IT WAS REPORTED THAT NURSING STAFF HAVE TRIED TO PREVENT THIS BY TAPING THE TUBING BUT WAS CLEARLY INEFFECTIVE REGARDLESS. THE SITE CLOSEST TO PATIENT WAS CLEANSED WITH CHOLORHEXADINE. DEVICE WAS REMOVED FROM PATIENT AS PER ORDER PROMPTLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22395499 ↗MalfunctionJul 1, 2025Disconnection; Disconnection
IT WAS REPORTED THAT HEMOVAC DRAIN DISCONNECTED FROM THE Y SITE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22372723 ↗MalfunctionJul 1, 2025Disconnection; Disconnection
IT WAS REPORTED THAT PATIENT'S HEMOVAC HAD COME APART AFTER THEY ROLLED ONTO THEIR SIDE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22372714 ↗MalfunctionJun 16, 2025Disconnection; Disconnection
IT WAS REPORTED THAT HEMOVAC DAVOL CLOSED WOUND SUCTION EVACUATOR TUBING CAME OUT AT THE DEVICE SITE. PER ADDITIONAL INFORMATION RECEIVED VIA MAIL ON 11JUN2025, STATED THAT THERE WAS NO PATIENT HARM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22211141 ↗MalfunctionJun 16, 2025Disconnection; Disconnection
IT WAS REPORTED THAT THE PATIENT HEMOVAC WOUND DRAIN DISCONNECTED FROM THE TUBE EVEN WITH TAPE ON IT. PER ADDITIONAL INFORMATION RECEIVED VIA MAIL ON 11JUN2025, STATED THAT THERE WAS NO PATIENT HARM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22211129 ↗MalfunctionJun 12, 2025Disconnection; Disconnection
IT WAS REPORTED THAT THE HEMOVAC DRAIN DISCONNECTED FROM THE Y SITE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22192417 ↗MalfunctionJun 11, 2025Disconnection; Disconnection
IT WAS REPORTED THAT HEMOVAC WOUND DRAIN WAS DISCONNECTED FROM THE Y SITE. PER ADDITIONAL INFORMATION RECEIVED VIA MAIL ON 11JUN2025, STATED THAT THERE WAS NO PATIENT HARM REPORTED, AND THE TROUBLESHOOTING WAS PERFORMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22183893 ↗MalfunctionMay 2, 2025Disconnection; Disconnection
IT WAS REPORTED THAT THE HEMOVAC DISCONNECTED FROM THE RESERVOIR. PER FOLLOW UP RESPONSE RECEIVED VIA EMAIL ON 30APR2025, CUSTOMER STATED THAT ADDED CARE TIME TO PATIENT AND EXPOSED TO OPEN SYSTEM, POTENTIAL FOR PATIENT HARM. CURRENTLY STAFF WERE TAPING THE DRAINS AS A PRECAUTION WHEN THEY WERE OUT OF THE OPERATING ROOM OR AFTER THEY HAD FALLEN APART.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21951230 ↗MalfunctionApr 23, 2025Disconnection; Disconnection
IT WAS REPORTED THAT SILICONE EVACUATOR HEMOVAC DISCONNECTED FROM THE RESERVOIR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21892860 ↗MalfunctionApr 22, 2025Disconnection; Disconnection
IT WAS REPORTED THAT WOUND DRAINAGE HEMOVAC GOT DISCONNECTED FROM THE RESERVOIR AS PATIENT WAS TRYING TO MOBILIZE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21882368 ↗MalfunctionApr 21, 2025Disconnection; Disconnection
IT WAS REPORTED THAT HEMOVAC DISCONNECTED FROM PATIENT. THERE WAS NO SERIOUS HARM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21873059 ↗MalfunctionApr 21, 2025Disconnection; Disconnection
IT WAS REPORTED THAT HEMOVAC DISCONNECTED FROM THE PATIENT. THERE WAS NO SERIOUS HARM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21872660 ↗MalfunctionApr 7, 2025Disconnection; Disconnection
IT WAS REPORTED THAT THE HEMOVAC CONTAINER OF THE WOUND DRAIN WAS DISCONNECTED FROM PATIENT AT THE Y-SITE, EVEN THOUGH CONNECTION POINT WAS TAPED. DISCONNECTED FROM PATIENT FROM UNKNOWN AMOUNT OF TIME. PER ADDITIONAL INFORMATION RECEIVED VIA EMAIL ON 02APR2025, THE DEVICE HAS BEEN RETAINED AND COULD BE SENT BACK TO BD. IT WAS CONTAMINATED WITH FLUID FROM THE PATIENT IT WAS ATTACHED TO. AS PER THE SHIPPING LABEL EMAIL THEY WOULD NEED A CANISTER TO RETURN IT TO THEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21784537 ↗MalfunctionMar 20, 2025Fracture; Fracture
IT WAS REPORTED THAT THE HEMOVAC LEAKED YESTERDAY. WHEN THE STAFF WENT TO REPLACE IT, THEY DISCOVERED IT WAS CRACKED. THE PHYSICIAN ASSISTANT SWITCHED IT OUT AND THEY SAID THEY DO TRIM THE TUBING, WHICH THEY DID IN THIS CASE BECAUSE IT WAS TOO LONG AND CUT IT AT AN ANGLE BEFORE CONNECTING IT TO Y.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21653520 ↗MalfunctionMar 20, 2025Incorrect Measurement; Incorrect Measurement
IT WAS REPORTED THAT THE HEMOVAC LEAKED YESTERDAY. WHEN THE STAFF WENT TO REPLACE IT, THEY DISCOVERED IT WAS CRACKED. THE PHYSICIAN ASSISTANT SWITCHED IT OUT AND THEY SAID THEY DO TRIM THE TUBING, WHICH THEY DID IN THIS CASE BECAUSE IT WAS TOO LONG AND CUT IT AT AN ANGLE BEFORE CONNECTING IT TO Y.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21652697 ↗MalfunctionJan 17, 2025Material Fragmentation; Material Fragmentation
IT WAS REPORTED THAT A CONNECTOR ON THE HEMOVAC CRACKED AND WAS LEAKING DURING THE HEMOVAC DRAIN INSERTION PROCEDURE. THE PATIENT FELT INCONVENIENT. NO MEDICAL INTERVENTION WAS REPORTED.
Patient problems: Discomfort
MAUDE report 21180369 ↗MalfunctionJan 16, 2025Dull, Blunt; Dull, Blunt
IT WAS REPORTED THAT CUSTOMER HAD ISSUES WITH THE DULLNESS OF THE TROCAR. THE TROCAR WAS LACKING IN SHARPNESS, AND THE CUSTOMER WAS CONSIDERING SWITCHING TO COMPETITOR. CUSTOMER HAD DIFFICULTIES PASSING THE DRAIN BECAUSE OF THE TROCAR DULL. THEY PREFERRED TO USE THE MEDLINE DRAINS THAT THEY USED AT A DIFFERENT FACILITY. PER ADDITIONAL INFORMATION RECEIVED ON 27DEC2024, IT WAS REPORTED THAT THE ITEMS WERE #0043600 AND #0043610. THERE WAS NO IMPACT TO THE PATIENT AND NO MEDICAL INTERVENTION. THEY HAD BEEN HAVING PROBLEMS PASSING THE TROCAR THROUGH THE SKIN BECAUSE OF THE DULLNESS OF IT, IT HAD BEEN VERY HARD TO PASS IT ALL THE WAY WITHOUT ALMOST GETTING POKED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21169561 ↗MalfunctionDec 26, 2024Disconnection; Disconnection
IT WAS REPORTED THAT THEY WERE CONTINUING TO HAVE ISSUES WITH THEIR HEMOVAC. THERE HAS NOT BEEN A FOLLOW-UP SINCE THE INCONCLUSIVE REPORT. IT HAPPENED AT THE "Y" CONNECTION OF THE PORT, IN RECOVERY WHEN THE PATIENT WAS GETTING DRESSED OR WORKING WITH PHYSICAL THERAPY, IT WOULD DISCONNECT CAUSING BLOOD TO COME FROM THE DRAIN AND THEY WOULD HAVE TO RECONNECTED ALSO TRIED TO REINFORCE THE DRAIN PORT WITH TAPE BUT OFTEN STILL HAD PROBLEMS. NO MEDICAL INTERVENTION WAS REPORTED. PER ADDITIONAL INFORMATION RECEIVED ON 06DEC2024, IT WAS REPORTED THAT THE NURSES WERE TAPING IT DOWN BEFORE THE PATIENT LEAVES BUT WHEN THEY WERE DOING POST OP CALLS, PATIENTS WERE COMPLAINING ABOUT IT COMING APART. THE NURSES HAD ALSO SAID WHEN THEY WERE GETTING PATIENTS DRESSED, THEY HAD COME APART AND OF COURSE CAUSED BLEEDING ALL OVER THE PATIENT. THE NURSES STATED THAT THE TUBING THAT LOOKS LIKE A Y SECTION THAT CONNECTS WITH THE VAC ITSELF IS WHERE IT WAS COMING APART (LOT#: NGHT2525, LOT#: UNK, LOT#: NGEQ3519).
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 21018787 ↗InjurySep 20, 2024Disconnection; Disconnection
IT WAS REPORTED THAT HEMOVAC WOUND DRAIN WILL COME APART WHEN PATIENT WAS MOVING FOR OVERNIGHT. THE BLOOD SHOT INTO PHYSICAL THERAPIST'S EYE BECAUSE IT CAME LOOSE AND DID NOT HAVE A LOCK LIKE A JP DRAIN.
Patient problems: Exposure to Body Fluids
MAUDE report 20266525 ↗MalfunctionOct 17, 2023Fitting Problem; Fitting Problem
IT WAS REPORTED THE Y CONNECTOR OF THE SUCTION EVACUATOR WAS DAMAGED. DURING USE, SUCTION WAS NOT APPLIED, AND AIR WAS LEAKING. THE USER ASKED TO CHECK WHETHER THE CONNECTOR HAD A HOLE. AS PER INFORMATION MENTIONED IN THE INTERNAL BD COMMENTS ON (B)(6) 2023, STATED THAT THE CONNECTION BETWEEN THE MAIN UNIT AND THE TUBE WAS LOOSE AND WOBBLY AND SEEMED TO BE COMING OFF. THE Y-SHAPED PART WAS NORMAL, ONE SIDE HAD A HOLE AND THE OTHER SIDE WAS CLOSED. THEY COULD NOT SEE ANY DAMAGE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17948993 ↗MalfunctionOct 18, 2022Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT THE EVACUATOR WAS LEAKING AT THEY PORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15624514 ↗MalfunctionOct 12, 2022Leak/Splash
DEVICE FOUND LEAKING AT THE Y PORT.
Patient problems: Insufficient Information
MAUDE report 15592600 ↗MalfunctionSep 2, 2021Material Fragmentation; Material Split, Cut or Torn; Material Fragmentation; Material Split, Cut or Torn
IT WAS REPORTED THAT THE TIP OF THE NEEDLE FELL OFF DURING USE AFTER OPENING THE PACKAGE. WHEN CUSTOMER CHECKED, CUSTOMER FOUND THAT THE TUBE TIP WAS TORN AND HAD BEEN REPAIRED WITH SOME SORT OF CELLOPHANE TAPE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12414974 ↗
Other APPARATUS, SUCTION, SINGLE PATIENT USE, PORTABLE, NONPOWERED devices
- Closed Wound Suction 0043610C. R. Bard, Inc. · APPARATUS, SUCTION, SINGLE PATIENT USE, PORTABLE, NONPOWERED303 reports
- Closed Wound Suction 0070740C. R. Bard, Inc. · APPARATUS, SUCTION, SINGLE PATIENT USE, PORTABLE, NONPOWERED192 reports
- Evacuator 0070800C. R. Bard, Inc. · APPARATUS, SUCTION, SINGLE PATIENT USE, PORTABLE, NONPOWERED124 reports
- Closed Wound Suction 0034610C. R. Bard, Inc. · APPARATUS, SUCTION, SINGLE PATIENT USE, PORTABLE, NONPOWERED62 reports
- CARDINAL HEALTH SU130-402DCardinal Health 200, LLC · APPARATUS, SUCTION, SINGLE PATIENT USE, PORTABLE, NONPOWERED1 recall1 report
- NA 1A8504Baxter International Inc. · APPARATUS, SUCTION, SINGLE PATIENT USE, PORTABLE, NONPOWERED50 reports
- CARDINAL HEALTH SU130-475Cardinal Health 200, LLC · APPARATUS, SUCTION, SINGLE PATIENT USE, PORTABLE, NONPOWERED1 recall0 reports
- DLP® 10061Medtronic, Inc. · APPARATUS, SUCTION, SINGLE PATIENT USE, PORTABLE, NONPOWERED1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page