Closed Wound Suction 0072360Recalls & reported problems
C. R. Bard, Inc. · 100cc Silicone Closed Wound Suction Evacuator Kit, 100cc Silicone Evacuator with Anti-Reflux Valve, Hubless Silicone Flat Drain 7mm Wide, 20cm Long, 3/4 Perforated Drain with Trocar, X-Ray Opaque · Class 1
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00801741050008), received between Dec 19, 2022 and Dec 21, 2022 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Component Misassembled” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Component Misassembled1
- Device Contamination with Chemical or Other Material1
- Material Fragmentation1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 21, 2022Material Fragmentation; Device Contamination with Chemical or Other Material; Material Fragmentation; Device Contamination with Chemical or Other Material
IT WAS REPORTED THAT THERE WAS A PRESENCE OF A SILICONE STOPPER INSIDE THE TUBING OF SUCTION EVACUATOR WHICH WAS PROXIMAL TO THE TROCAR. THE INTERNAL WHITE PART WAS PERFORATED ONLY ON THE TWO-THIRD INSTEAD OF WHOLE. THEREFORE, IT WAS IMPOSSIBLE TO USE ON CRANIOTOMY FOR TOO LONG AS IT WAS NOT PERFORATED THE FULL LENGTH.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16022890 ↗MalfunctionDec 19, 2022Component Misassembled; Component Misassembled
IT WAS REPORTED THAT THERE WAS A PRESENCE OF A SILICONE STOPPER INSIDE THE TUBING OF SUCTION EVACUATOR WHICH WAS PROXIMAL TO THE TROCAR. THE INTERNAL WHITE PART WAS PERFORATED ONLY ON THE TWO-THIRD INSTEAD OF WHOLE. THEREFORE, IT WAS IMPOSSIBLE TO USE ON CRANIOTOMY FOR TOO LONG AS IT WAS NOT PERFORATED THE FULL LENGTH.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16003905 ↗
Other APPARATUS, SUCTION, SINGLE PATIENT USE, PORTABLE, NONPOWERED devices
- Closed Wound Suction 0043610C. R. Bard, Inc. · APPARATUS, SUCTION, SINGLE PATIENT USE, PORTABLE, NONPOWERED303 reports
- Closed Wound Suction 0070740C. R. Bard, Inc. · APPARATUS, SUCTION, SINGLE PATIENT USE, PORTABLE, NONPOWERED192 reports
- Evacuator 0070800C. R. Bard, Inc. · APPARATUS, SUCTION, SINGLE PATIENT USE, PORTABLE, NONPOWERED124 reports
- Closed Wound Suction 0034610C. R. Bard, Inc. · APPARATUS, SUCTION, SINGLE PATIENT USE, PORTABLE, NONPOWERED62 reports
- CARDINAL HEALTH SU130-402DCardinal Health 200, LLC · APPARATUS, SUCTION, SINGLE PATIENT USE, PORTABLE, NONPOWERED1 recall1 report
- NA 1A8504Baxter International Inc. · APPARATUS, SUCTION, SINGLE PATIENT USE, PORTABLE, NONPOWERED50 reports
- CARDINAL HEALTH SU130-475Cardinal Health 200, LLC · APPARATUS, SUCTION, SINGLE PATIENT USE, PORTABLE, NONPOWERED1 recall0 reports
- DLP® 10061Medtronic, Inc. · APPARATUS, SUCTION, SINGLE PATIENT USE, PORTABLE, NONPOWERED1 recall0 reports