Hickman 0600330Recalls & reported problems
Bard Access Systems, Inc. · Hickman 9 Fr Dual-Lumen CV Catheter, Cutdown Tray with SureCuff Tissue Ingrowth Cuff · Class 2 · Implant
All models of Hickman (28)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00801741051753), received between Mar 23, 2017 and Jul 26, 2021 — classified as 8 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Fluid/Blood Leak5
- Break3
- Burst Container or Vessel2
- Contamination1
- Fracture1
- Split1
Patient problems
- No Known Impact Or Consequence To Patient5
- No Consequences Or Impact To Patient3
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 26, 2021Contamination; Contamination
IT WAS REPORTED THAT PRIOR TO THE CATHETER PLACEMENT, THE DEVICE OUTER PACKAGE WAS INTACT BUT THE DEVICE INSIDE THE PACKAGE WAS EXTREMELY DIRTY AND DISCOLORED. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12224536 ↗MalfunctionOct 26, 2020Break; Fracture; Break; Fracture
THIS REPORT SUMMARIZES ONE MALFUNCTION. A REVIEW OF THE REPORTED INFORMATION INDICATED THAT MODEL 0600330 CHRONIC CATHETER ALLEGEDLY EXPERIENCED FRACTURE. THIS INFORMATION WAS RECEIVED FROM ONE SOURCE. THE ONE REPORTED MALFUNCTION INVOLVED ONE PATIENT WITH NO CONSEQUENCES. AGE, WEIGHT AND GENDER WERE NOT PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10733080 ↗MalfunctionAug 22, 2018Burst Container or Vessel; Fluid/Blood Leak; Burst Container or Vessel; Fluid/Blood Leak
IT WAS REPORTED THAT THE WHITE LUMEN OF THE CATHETER HAD A LEAK. THE DEVICE WAS REMOVED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7806975 ↗MalfunctionJun 20, 2018Burst Container or Vessel; Fluid/Blood Leak; Burst Container or Vessel; Fluid/Blood Leak
IT WAS REPORTED THAT WHEN FLUSHING THE CATHETER WITH 10ML SYRINGE, A "POP" WAS HEARD AND SALINE WAS OBSERVED TO BE SQUIRTING FROM THE LINE. EXTERNAL CATHETER FRACTURE WAS OBSERVED. THE DEVICE WAS CLAMPED AND REMOVED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 7623271 ↗MalfunctionApr 17, 2018Fluid/Blood Leak
IT WAS REPORTED THAT THE ¿PATIENT WAS DUE IV FLUCLOXACILLIN ANTIBIOTICS MADE UP INTO 20MLS. WHEN ATTEMPTING TO BOLUS THE DRUG INTO THE WHITE PORT, LIQUID APPEARED TO SQUIRT OUT OF THE SIDE OF THE ACTUAL LINE. ON EXAMINATION A SMALL FRACTURE APPROXIMATELY 2MM IN LENGTH COULD BE SEEN JUST BELOW THE CLAMP.¿ HICKMAN CATHETER. NO PATIENT INJURY REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7438564 ↗MalfunctionMar 6, 2018Fluid/Blood Leak; Split
IT WAS REPORTED THAT AFTER 3 DAYS POST INSERTION, THE NURSE OBSERVED LEAKING FLUID FROM THE LUMEN WHILE FLUSHING. A SPLIT ABOVE THE WHITE ADAPTER LEG WAS OBSERVED BY IV ACCESS TEAM. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7317800 ↗MalfunctionJun 29, 2017Break; Fluid/Blood Leak
THE HICKMAN CATHETER WAS INSERTED TO THE PATIENT ON (B)(6) 2017 WITHOUT ANY PROBLEM. THE CATHETER WAS BEEN USED UNTIL (B)(6) 2017. CATHETER LEAKAGE REPORTED AND BROKEN POINT WAS FOUND ON THE EXTENSION LEG OF THE CATHETER. THE CATHETER WAS THEN REMOVED ACCORDING TO THEIR HOSPITAL PROTOCOL. NO PATIENT INJURY REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6676985 ↗MalfunctionMar 23, 2017Break
AS REPORTED BY THE FACILITY, THE HICKMAN CATHETER WAS FOUND BROKEN AT ONE OF THE EXTENSION LEGS. THE CATHETER WAS THEN HANDLED AND REMOVED ACCORDING TO HOSPITAL PROTOCOL. NO HARM TO PATIENT REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6428453 ↗
Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices
- Groshong NXT ClearVue Catheter 7617405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,708 reports
- Groshong NXT ClearVue Catheter 7655405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,536 reports
- PowerPICC Catheter 3174118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall402 reports
- L-Cath™ S/L PICC S/L 26GA (1.9F) x 30cmArgon Medical Devices, Inc. · Catheter,Intravascular,Therapeutic,Long-Term Greater Than 30 Days2 recalls245 reports
- PowerPICC SOLO2 Catheter 1194108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 recalls216 reports
- Broviac 0600520Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall265 reports
- Groshong® Catheter 7717405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days279 reports
- PowerPICC® Catheters 1174108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall222 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page