Skip to content

Hickman 0600574Recalls & reported problems

Bard Access Systems, Inc. · Hickman 7 Fr Dual-Lumen CV Catheter, Peel-Apart Introducer Kit with SureCuff · Class 2 · Implant

All models of Hickman (28)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00801741051876), received between Feb 24, 2017 and Dec 24, 2024 — classified as 3 injuries, 5 malfunctions.

The most frequently reported specific device problem is “Air/Gas in Device” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 2 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Air/Gas in Device2
  • Fracture2
  • Device Dislodged or Dislocated1
  • Failure to Infuse1
  • Hole In Material1
  • Material Integrity Problem1
  • Material Puncture/Hole1
  • Suction Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Air Embolism1
  • Device Embedded In Tissue or Plaque1
  • No Consequences Or Impact To Patient1
  • No Known Impact Or Consequence To Patient1
  • Unspecified Infection1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 24, 2024Fracture; Material Integrity Problem; Fracture; Material Integrity Problem

    IT WAS REPORTED THAT SOMETIMES POST A CHRONIC CATHETER PLACEMENT, THE OUTER SHEATH OF CATHETER WAS ALLEGEDLY DAMAGED AT BIFURCATION SITE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21009838 ↗
  • InjuryDec 24, 2024Suction Problem; Suction Problem

    IT WAS REPORTED THAT SOMETIMES POST A CHRONIC CATHETER PLACEMENT, THE DEVICE ALLEGEDLY HAD NO BLOOD RETURN. IT WAS ALSO FURTHER REPORTED THAT THE PATIENT HAD AN INFECTION. HOWEVER, THE CURRENT STATUS OF THE PATIENT WAS UNKNOWN.

    Patient problems: Unspecified Infection

    MAUDE report 21009655 ↗
  • InjuryMay 25, 2022Air/Gas in Device; Air/Gas in Device

    IT WAS REPORTED THAT DURING A CHRONIC CATHETER PLACEMENT PROCEDURE IN THE RIGHT INTERNAL JUGULAR, THE PATIENT ALLEGEDLY EXPERIENCED AIR EMBOLISM. IT WAS FURTHER REPORTED THAT MEDICAL INTERVENTIONS WAS REQUIRED TO REMOVE AIR EMBOLISM THROUGH MEDICATIONS. THE PATIENT REPORTEDLY TO BE STABLE.

    Patient problems: Air Embolism

    MAUDE report 14496965 ↗
  • MalfunctionMay 25, 2022Air/Gas in Device; Air/Gas in Device

    IT WAS REPORTED THAT DURING A CHRONIC CATHETER PLACEMENT PROCEDURE IN THE RIGHT INTERNAL JUGULAR, THE DEVICE ALLEGEDLY HAD BUBBLES IN THE LINE. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14495838 ↗
  • MalfunctionOct 5, 2021Fracture; Material Puncture/Hole; Fracture; Material Puncture/Hole

    IT WAS REPORTED THAT PRIOR TO A CHRONIC CATHETER PLACEMENT PROCEDURE, A HOLE WAS ALLEGEDLY FOUND NEAR THE DISTAL END OF THE CATHETER. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12576378 ↗
  • MalfunctionJun 17, 2019Failure to Infuse; Failure to Infuse

    IT WAS REPORTED THAT APPROXIMATELY TWO MONTHS POST CATHETER PLACEMENT, THE WHITE LUMEN ALLEGEDLY FAILED TO FLUSH. IT WAS FURTHER REPORTED THAT THE DEVICE WAS REMOVED AND REPLACED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8704478 ↗
  • InjuryApr 14, 2017Device Dislodged or Dislocated

    IT WAS REPORTED BY THE SALES REP THAT THE PATIENT HAD A HICKMAN CATHETER PLACED ON (B)(6) 2016. IN (B)(6) OF 2017, A FULL SEGMENT REPAIR WAS COMPLETED. TWO WEEKS LATER THE FACILITY TRIED TO ACCESS THE CATHETER AND IT WAS "SLUGGISH" SO THEY RAN CONTRAST WITH A 10 ML SYRINGE USING 3 ML OF CONTRAST AND ON THE THIRD TIME FLUORO NOTICED THE METAL STRUT HAD FLOATED DOWN THE CATHETER AND EXITED INTO THE PATIENT, LODGING IN HIS PULMONARY ARTERY BRANCH. THE SALES REP STATED THAT THE PATIENT IS STABLE AND THE FACILITY DID NOT TRY TO REMOVE IT. THE SALES REP STATED IT WAS THE ARTERIAL METAL STRUT THAT BROKE OFF. PATIENT REMAINS ASYMPTOMATIC. FACILITY REPORTED THAT CATHETER WAS REMOVED AND REPLACED.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 6494818 ↗
  • MalfunctionFeb 24, 2017Hole In Material

    FACILITY REPORTED THAT THE WHITE LUMEN HAD A PINHOLE AT HUB OF LINE. FACILITY ALSO REPORTED THAT THE OUTER LAYER HAD SEPARATED ABOVE THE TRIANGLE AREA WHERE THE LUMENS MEET. REQUIRED DUAL REPAIR. NO PATIENT INJURY REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6356844 ↗

Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices

All Catheter,intravascular,therapeutic,long-term greater than 30 days devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page