Bard® X-Force® Balloon Dilation Catheter 997610Recalls & reported problems
C. R. Bard, Inc. · Bard® X-Force® Balloon Dilation Catheter · Class 2
All models of Bard X-Force Balloon Dilation Catheter (22)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00801741052118), received between Nov 30, 2018 and Jun 15, 2020 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Burst Container or Vessel” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Burst Container or Vessel1
- Deflation Problem1
- Display or Visual Feedback Problem1
- Fluid/Blood Leak1
- Use of Device Problem1
Patient problems
- No Known Impact Or Consequence To Patient2
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 15, 2020Fluid/Blood Leak; Burst Container or Vessel; Fluid/Blood Leak
IT WAS REPORTED THAT THE CATHETER BALLOON WAS EXPLODED DURING SURGERY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10155493 ↗MalfunctionNov 30, 2018Use of Device Problem; Deflation Problem; Display or Visual Feedback Problem; Use of Device Problem
IT WAS REPORTED THAT THE SURGEON COMPLAINED THAT THE BALLOON WAS NOT DEFLATING COMPLETELY WHEN IT WAS TIME TO REMOVE THE BALLOON FROM THE URETER FOLLOWING DILATION. UPON TESTING, THE BALLOON DEFLATED NORMALLY. THE BARD SALES REP. STATED THE ISSUE STEMMED FROM THE FACT THAT THE USER (THE SCRUB TECH) WAS NOT WAITING LONG ENOUGH FOR THE VACUUM EFFECT TO TAKE FULL EFFECT. IN ORDER TO DEFLATE THE BALLOON, THE INFLATOR SHOULD BE DEPLOYED BACK FULLY TO ALLOW THE PRESSURE TO PULL ALL OF THE FLUID OUT OF THE BALLOON AND DEFLATE IT. FINAL CONCLUSION WAS THAT THE PRODUCT WAS WORKING PROPERLY AND THIS EVENT WAS A USER ERROR.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 8120912 ↗
Other DILATOR, CATHETER, URETERAL devices
- Porges Coloplast RBD014Coloplast A/S · DILATOR, CATHETER, URETERAL1 recall0 reports
- Bard® X-Force® Balloon Dilation Catheter 998706C. R. Bard, Inc. · DILATOR, CATHETER, URETERAL38 reports
- Bard® X-Force® Balloon Dilation Catheter 998606C. R. Bard, Inc. · DILATOR, CATHETER, URETERAL27 reports
- Bard® X-Force® Balloon Dilation Catheter 998006C. R. Bard, Inc. · DILATOR, CATHETER, URETERAL25 reports
- Bard® X-Force® Balloon Dilation Catheter 998806C. R. Bard, Inc. · DILATOR, CATHETER, URETERAL23 reports
- Bard® X-Force® Balloon Dilation Catheter 998504C. R. Bard, Inc. · DILATOR, CATHETER, URETERAL16 reports
- Bard® X-Force® Balloon Dilation Catheter 998004C. R. Bard, Inc. · DILATOR, CATHETER, URETERAL15 reports
- Bard® X-Force® Balloon Dilation Catheter 998506C. R. Bard, Inc. · DILATOR, CATHETER, URETERAL15 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page