Skip to content

Bard® Hydro-Glide™ Guidewire 145FS35Recalls & reported problems

C. R. Bard, Inc. · Bard® Hydro-Glide™ Guidewire · Class 1

All models of Bard Hydro-Glide Guidewire (3)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Bard Hydro-Glide guidewire is a flexible, hydrophilic-coated wire used to reach the bladder, ureter and kidney during urology procedures. The 31 reports mostly describe the coating flaking or peeling off the wire, sometimes with material breaking off or the tip appearing damaged, plus difficulty passing the wire past stones. Two reports describe injuries. No recalls name this device, and no reports came in the last 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
31
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 31 adverse event reports that list this exact device (DI 00801741053252), received between Jul 10, 2019 and Jun 21, 2024 — classified as 2 injuries, 29 malfunctions.

The most frequently reported specific device problem is “Flaked” (22 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Flaked22
  • Difficult to Insert2
  • Accessory Incompatible1
  • Device Contamination with Chemical or Other Material1
  • Inadequacy of Device Shape and/or Size1
  • Incorrect Measurement1
  • Material Fragmentation1
  • Material Protrusion/Extrusion1

Patient problems

  • No Clinical Signs, Symptoms or Conditions26
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available3
  • Exposure to Body Fluids1
  • No Consequences Or Impact To Patient1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 25 most recent of 31 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 21, 2024Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT THE GUIDEWIRE OF PROXIMAL TIP WAS DAMAGED WHEN LOOKING AT THE PROXIMAL TIP BEFORE USING IT. PER FOLLOW UP INFORMATION RECEIVED VIA IBC ON (B)(6) 2024, THEY CONFIRMED THE PROXIMAL TIP SEEMED TO BE A LITTLE BIT BROKEN WHEN CHECKING FOR THE PRODUCT BEFORE USING IT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19583419 ↗
  • MalfunctionSep 15, 2023Difficult to Insert; Difficult to Insert

    CRITICAL CARE NURSES REPORTED IN AN ONLINE SURVEY THAT THE GUIDEWIRE WAS NOT SUCCESSFUL IN PROVIDING TRANSURETHRAL ACCESS INTO THE URETER BECAUSE URETERAL CALCULUS WAS NOT ALLOWING PASSAGE OF THE GUIDEWIRE IN RELATION TO 145FS35: GUIDEWIRE, HYDROGLIDETM, FLEXIBLE STRAIGHT TIP, FIXED CORE, STANDARD TAPER, HYDROPHILIC, 145CM LONG,.035", STERILE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 17754601 ↗
  • MalfunctionSep 15, 2023Difficult to Insert; Difficult to Insert

    CRITICAL CARE NURSES REPORTED IN AN ONLINE SURVEY THAT THE GUIDEWIRE WAS NOT SUCCESSFUL IN PROVIDING PERCUTANEOUS ACCESS INTO THE RENAL PELVIS DUE TO STONE OBSTRUCTION IN THE RENAL INFUNDIBULUM, WHICH DID NOT ALLOW ACCESS TO THE GUIDEWIRE IN RELATION TO 145FS35: GUIDEWIRE, HYDROGLIDETM, FLEXIBLE STRAIGHT TIP, FIXED CORE, STANDARD TAPER, HYDROPHILIC, 145CM LONG,.035", STERILE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 17754598 ↗
  • MalfunctionSep 12, 2023Material Protrusion/Extrusion; Material Protrusion/Extrusion

    IT WAS REPORTED THAT PURPLE COATING COLOR OF THE GUIDEWIRE WAS PEELING AND COULD BE SEEN UNDER ENDOSCOPY. STATED THAT THE USER WAS EXPOSED TO HAZARDOUS, BLOOD, OR BODILY FLUIDS. THE USER NEEDED TO REINSERT A NEW GUIDEWIRE. IT WAS UNKNOWN WHAT MEDICAL INTERVENTION WAS PROVIDED. PER SAMPLE EVALUATION NOTIFICATION RECEIVED ON 22AUG2023, STATED THAT THE GUIDEWIRE WAS FOUND TO BE EXPOSED DURING THE SAMPLE EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17729261 ↗
  • InjurySep 11, 2023Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size

    IT WAS REPORTED THAT IN THE ONLINE SURVEY A PHYSICIAN STATED THAT NO, THE GUIDEWIRE WAS NOT SUCCESSFUL IN PROVIDING TRANSURETHRAL/PERCUTANEOUS ACCESS INTO THE BLADDER/URETER/RENAL PELVIS BECAUSE IMPACTED STONE AT THE PROXIMAL URETER, ATTEMPTED MULTIPLE WIRES. THEY REQUIRED A PERCUTANEOUS NEPHROSTOMY FOR AN INFECTED STONE IN RELATION TO 145FS35 - GUIDEWIRE, HYDROGLIDE¿, FLEXIBLE STRAIGHT TIP, FIXED CORE, STANDARD TAPER, HYDROPHILIC, 145CM LONG, .035 INCH, STERILE. PER ADDITIONAL INFORMATION RECEIVED VIA EMAIL ON 23AUG2023, IT WAS REPORTED THAT THE SURVEY RESPONDENT STATED THAT NO, THE GUIDEWIRE WAS NOT SUCCESSFUL IN PROVIDING TRANSURETHRAL ACCESS INTO THE RENAL PELVIS BECAUSE IMPACTED STONE AT THE PROXIMAL URETER, ATTEMPTED MULTIPLE WIRES. THEY REQUIRED A PERCUTANEOUS NEPHROSTOMY FOR AN INFECTED STONE IN RELATION TO 145FS35 - GUIDEWIRE, HYDROGLIDE¿, FLEXIBLE STRAIGHT TIP, FIXED CORE, STANDARD TAPER, HYDROPHILIC, 145CM LONG, .035", STERILE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 17718589 ↗
  • MalfunctionAug 14, 2023Flaked; Flaked

    IT WAS REPORTED THAT THE MATERIAL FELL OFF FROM SHEATH SURFACE OF THE HYDRO-GLIDE GUIDEWIRE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17540222 ↗
  • MalfunctionJul 28, 2023Flaked; Flaked

    IT WAS REPORTED THAT DURING USE OF THE URETHRAL STENT ON (B)(6) 2023, THE GUIDE WIRE WAS FOUND UNSMOOTH WITH THE COATING FLAKING OFF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17413709 ↗
  • MalfunctionJul 11, 2023Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material

    IT WAS REPORTED THAT GUIDEWIRE HAD WHITE DROPS THAT LOOK LIKE MAGGOTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17298308 ↗
  • InjuryMay 12, 2023Flaked; Flaked

    IT WAS REPORTED THAT PURPLE COATING COLOR OF THE GUIDEWIRE WAS PEELING AND COULD BE SEEN UNDER ENDOSCOPY. STATED THAT THE USER WAS EXPOSED TO HAZARDOUS, BLOOD, OR BODILY FLUIDS. THE USER NEEDED TO REINSERT A NEW GUIDEWIRE. IT WAS UNKNOWN WHAT MEDICAL INTERVENTION WAS PROVIDED. PER SAMPLE EVALUATION NOTIFICATION RECEIVED ON 22AUG2023, STATED THAT THE GUIDEWIRE WAS FOUND TO BE EXPOSED DURING THE SAMPLE EVALUATION.

    Patient problems: Exposure to Body Fluids

    MAUDE report 16918309 ↗
  • MalfunctionMay 4, 2023Flaked; Flaked

    IT WAS REPORTED THAT, GUIDEWIRE PURPLE COATING COLOR PEELING CAN BE SEEN UNDER ENDOSCOPY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16869255 ↗
  • MalfunctionMay 4, 2023Flaked; Flaked

    IT WAS REPORTED THAT, GUIDEWIRE PURPLE COATING COLOR PEELING COULD BE SEEN UNDER ENDOSCOPY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16869249 ↗
  • MalfunctionMay 4, 2023Flaked; Flaked

    IT WAS REPORTED THAT PURPLE COATING COLOR OF THE GUIDEWIRE WAS PEELING AND COULD BE SEEN UNDER ENDOSCOPY. AMONG THEM, 3 SAMPLES WOULD BE RETURNED AND ONE WOULD BE HAVING NO SAMPLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16867389 ↗
  • MalfunctionMay 2, 2023Flaked; Flaked

    IT WAS REPORTED THAT, GUIDEWIRE PURPLE COATING COLOR PEELING CAN BE SEEN UNDER ENDOSCOPY. IT WAS STATED THAT 7 DEVICES WERE REPORTED ISSUE. AMONG THEM, 5 SAMPLES WERE RETURNED AND 2 ARE NO SAMPLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16854619 ↗
  • MalfunctionMay 2, 2023Flaked; Flaked

    IT WAS REPORTED THAT, GUIDEWIRE PURPLE COATING COLOR PEELING CAN BE SEEN UNDER ENDOSCOPY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16852631 ↗
  • MalfunctionMay 2, 2023Flaked; Flaked

    IT WAS REPORTED THAT, GUIDEWIRE PURPLE COATING COLOR PEELING CAN BE SEEN UNDER ENDOSCOPY. IT WAS STATED THAT 7 DEVICES WERE REPORTED ISSUE. AMONG THEM, 5 SAMPLES WERE RETURNED AND 2 ARE NO SAMPLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16852629 ↗
  • MalfunctionMay 2, 2023Flaked; Flaked

    IT WAS REPORTED THAT, GUIDEWIRE PURPLE COATING COLOR PEELING CAN BE SEEN UNDER ENDOSCOPY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16852627 ↗
  • MalfunctionMay 1, 2023Flaked; Flaked

    IT WAS REPORTED THAT, GUIDEWIRE PURPLE COATING COLOR PEELING CAN BE SEEN UNDER ENDOSCOPY (LOT# GFGR1887 ; LOT# GFGS1239 ).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16844482 ↗
  • MalfunctionMay 1, 2023Flaked; Flaked

    IT WAS REPORTED THAT PURPLE COATING COLOR OF THE GUIDEWIRE WAS PEELING AND COULD BE SEEN UNDER ENDOSCOPY. AMONG THEM, (B)(4) SAMPLES WOULD BE RETURNED AND ONE WOULD BE HAVING NO SAMPLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16843764 ↗
  • MalfunctionMay 1, 2023Flaked; Flaked

    IT WAS REPORTED THAT, GUIDEWIRE PURPLE COATING COLOR PEELING CAN BE SEEN UNDER ENDOSCOPY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16843278 ↗
  • MalfunctionApr 28, 2023Flaked; Flaked

    IT WAS REPORTED THAT THE HYDRO GLIDE GUIDEWIRE OUTSIDE WAS SHEDDING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16832500 ↗
  • MalfunctionApr 28, 2023Flaked; Flaked

    IT WAS REPORTED THAT GUIDEWIRE OUT SIDE SHEDDING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16832497 ↗
  • MalfunctionApr 6, 2023Flaked; Flaked

    IT WAS REPORTED THAT COATING OF THE GUIDEWIRE WAS PEELING. THE DEVICE WAS NOT USED ON THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16691267 ↗
  • MalfunctionMar 12, 2023Accessory Incompatible; Accessory Incompatible

    IT WAS REPORTED THAT THE USER FELT STUCK WHEN USE THE PUSHER OF GUIDEWIRE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16527260 ↗
  • MalfunctionDec 5, 2022Flaked; Flaked

    IT WAS REPORTED THAT GUIDEWIRE PURPLE COATING COLOR PEELING WAS OBSERVED. THE PRODUCT WAS NOT USED ON PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15915717 ↗
  • MalfunctionNov 30, 2022Flaked; Flaked

    IT WAS REPORTED THAT GUIDEWIRE COATING SEEMED TO BE SCRATCHED OFF WHEN GOING THROUGH THE URETEROSCOPE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15886826 ↗

Other STYLET FOR CATHETER, GASTRO-UROLOGY devices

All STYLET FOR CATHETER, GASTRO-UROLOGY devices →

Sources and updates