Sensor M0066703080Recalls & reported problems
Boston Scientific Corporation · Nitinol Wire with Hydrophilic Tip · Class 1
All models of Sensor (7)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Boston Scientific Sensor hydrophilic-tip guidewire, used to guide catheters, has 118 adverse event reports, with 24 injuries and no deaths. Many describe the coating peeling off or the tip detaching, and some describe the wire breaking. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 118
- Deaths reported
- 0
- Injuries reported
- 24
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 118 adverse event reports that list this exact device (DI 08714729376361), received between Jun 28, 2017 and Aug 28, 2026 — classified as 24 injuries, 94 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (63 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component63
- Adverse Event Without Identified Device or Use Problem37
- Break29
- Peeled/Delaminated28
- Nonstandard Device7
- Entrapment of Device4
- Material Deformation4
- Defective Device3
Patient problems
- No Clinical Signs, Symptoms or Conditions85
- No Consequences Or Impact To Patient20
- Foreign Body In Patient7
- Hematuria2
- No Code Available2
- No Known Impact Or Consequence To Patient2
- Pain2
- Perforation2
Recent adverse event reports
The 25 most recent of 118 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 28, 2026Unsealed Device Packaging
It was reported that a packaging seal of sensor wire device package had been peeled off, however, there was no damage to the cardboard box or the outer carton. There was no patient involved.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50498495 ↗MalfunctionJul 6, 2026Peeled/Delaminated; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING THE STENT PLACEMENT PROCEDURE, A SENSOR WIRE WAS USED. IT WAS FOUND THAT THE COATING WAS OBSERVED TO BE PEELED ON THE PROXIMAL HANDLE SIDE PORTION. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED. THE ANALYSIS OF THE RETURNED DEVICE IDENTIFIED SLEEVE DETACHMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25725488 ↗MalfunctionMay 26, 2026Nonstandard Device; Peeled/Delaminated
IT WAS REPORTED THAT DURING A URETERIC STONE REMOVAL PROCEDURE, A SENSOR WIRE WAS USED. WHEN THE PHYSICIAN ADVANCED THE STENT ON THE GUIDE WIRE THE DISTAL TIP COATING CAME OFF. ANOTHER NEW DEVICE WAS OPENED AND SUCCESSFULLY COMPLETED THE PROCEDURE. NO PATIENT COMPLICATIONS WERE REPORTED. THE ANALYSIS OF THE RETURNED DEVICE IDENTIFIED SLEEVE DETACHMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25292747 ↗MalfunctionMay 14, 2026Device Contaminated During Manufacture or Shipping
IT WAS REPORTED THAT AFTER UNPACKING A SENSOR WIRE, THERE WAS A FOREIGN BODY IN THE PACKAGE. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25178935 ↗MalfunctionApr 13, 2026Break
IT WAS REPORTED THAT A SENSOR WIRE WAS USED DURING SURGICAL PROCEDURE. DURING PROCEDURE, IT WAS FOUND THAT THE STRAIGHT TIP SENSOR WIRE BLACK FLEXIBLE TIP BROKE OFF. THE PROCEDURE WAS COMPLETED. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: Foreign Body In Patient
MAUDE report 24857428 ↗MalfunctionMar 31, 2026Break; Nonstandard Device; Peeled/Delaminated
IT WAS REPORTED THAT DURING A TRANSURETHRAL LASER LITHOTRIPSY PROCEDURE, A SENSOR WIRE WAS USED. WHILE THE GUIDEWIRE WAS INSIDE THE PATIENT, THE PTFE COATING WAS OBSERVED TO HAVE PEELED OFF. THE ANALYSIS OF THE RETURNED DEVICE IDENTIFIED CORE WIRE BREAK.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24749910 ↗MalfunctionMar 18, 2026Detachment of Device or Device Component
IT WAS REPORTED THAT A SENSOR WIRE WAS USED DURING A URETEROSCOPY PROCEDURE. DURING PROCEDURE, THE WIRE BROKE IN THE BLADDER, AND THEY SUCCESSFULLY RETRIEVED THE PIECES LEFT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24635509 ↗MalfunctionDec 17, 2025Peeled/Delaminated; Detachment of Device or Device Component; Peeled/Delaminated; Detachment of Device or Device Component
IT WAS REPORTED THAT, THE SENSOR WIRE WAS FOUND LIFTED AND PEELED OFF AT THE DISTAL TIP. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE NOTED. THE ANALYSIS OF THE RETURNED DEVICE IDENTIFIED SLEEVE DETACHMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23829112 ↗MalfunctionDec 15, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT, DURING THE CYSTOSCOPY, URETEROSCOPY WITH LASER LITHOTRIPSY AND STENT PLACEMENT PROCEDURE, A SENSOR WIRE WAS USED, WHEN THE WIRE WAS REMOVED FROM THE PATIENT, IT WAS OBSERVED THE SLEEVE HAD DETACHED; HOWEVER, THE PROCEDURE WAS COMPLETED USING THE ORIGINAL DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23809420 ↗MalfunctionDec 4, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING A TRANSURETERAL LITHOTRIPSY PROCEDURE, WHEN THE DEVICE WAS WITHDRAWN FROM THE LOOP OF THE GUIDEWIRE AND ATTEMPTED TO BE USED, THE FLEX PORTION ON THE PROXIMAL SIDE WAS PEELED OFF OUTSIDE OF THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23720652 ↗MalfunctionDec 4, 2025Unsealed Device Packaging; Unsealed Device Packaging
IT WAS FOUND THAT, WHEN THE DEVICE WAS REMOVED FROM THE PACKAGING, THE SEAL WAS COMPROMISED. THE DEVICE WAS NOT USED IN THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23718183 ↗MalfunctionNov 28, 2025Break
DURING PROCEDURE, SURGEON INSERTED A STRAIGHT TIP SENSOR WIRE INTO LEFT URETER THE THAT BROKE OFF. THE BROKEN PIECE WAS VISUALIZED, RETRIEVED AND REMOVED FROM THE PATIENT.
Patient problems: Foreign Body In Patient
MAUDE report 23671432 ↗MalfunctionNov 24, 2025Nonstandard Device; Peeled/Delaminated; Detachment of Device or Device Component; Nonstandard Device; Peeled/Delaminated; Detachment of Device or Device Component
IT WAS REPORTED THAT, DURING THE STENT REPLACEMENT PROCEDURE USING A SENSOR WIRE, THE DEVICE COATING WAS PEELED OFF. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE AND THERE WERE NO KNOWN PATIENT COMPLICATIONS REPORTED. THE ANALYSIS OF THE RETURNED DEVICE IDENTIFIED SLEEVE DETACHMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23638982 ↗MalfunctionNov 7, 2025Break; Nonstandard Device; Break; Nonstandard Device
IT WAS REPORTED THAT, A SENSOR WIRE WILL BE USED IN A FLEXIBLE URETEROSCOPIC LITHOTRIPSY PROCEDURE. UPON UNPACKING, IT WAS FOUND THAT THE GUIDE WIRE WAS FRACTURED IN THE MIDDLE. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE, AND NO PATIENT COMPLICATIONS WERE NOTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23503661 ↗MalfunctionNov 3, 2025Break; Break
IT WAS REPORTED THAT DURING THE LASER LITHOTRIPSY PROCEDURE, THE WIRE STRIPPED INSIDE PATIENT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED USING AN ALTERNATIVE DEVICE, AND THERE WERE NO KNOWN PATIENT COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23457907 ↗MalfunctionOct 23, 2025Nonstandard Device; Detachment of Device or Device Component; Adverse Event Without Identified Device or Use Problem; Nonstandard Device; Detachment of Device or Device Component; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING THE TRANSURETHRAL LITHOTRIPSY PROCEDURE, THE TIP PART GOT SEPARATED WHEN IT WAS INSERTED INTO THE ENDOSCOPE. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE AND THERE WERE NO KNOWN PATIENT COMPLICATIONS. THE ANALYSIS OF THE RETURNED DEVICE IDENTIFIED SLEEVE DETACHMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23375622 ↗MalfunctionOct 23, 2025Break; Nonstandard Device; Adverse Event Without Identified Device or Use Problem; Break; Nonstandard Device; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING THE TRANSURETHRAL LITHOTRIPSY PROCEDURE, THE DEVICE WAS CUT ON THE PROXIMAL SIDE. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE AND THERE WERE NO KNOWN PATIENT COMPLICATIONS REPORTED. ANALYSIS OF THE RETURNED DEVICE IDENTIFIED A SLEEVE DETACHMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23375310 ↗MalfunctionOct 3, 2025Nonstandard Device; Peeled/Delaminated; Adverse Event Without Identified Device or Use Problem; Nonstandard Device; Peeled/Delaminated; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING THE PROCEDURE, THE DISTAL TIP PEELED OFF. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE WITHOUT PATIENT COMPLICATIONS. ANALYSIS OF THE RETURNED GUIDEWIRE IDENTIFIED THE SLEEVE WAS DETACHED FROM THE GUIDEWIRE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23207068 ↗MalfunctionSep 29, 2025Detachment of Device or Device Component; Adverse Event Without Identified Device or Use Problem; Detachment of Device or Device Component; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING THE PROCEDURE, UPON REMOVING THE GUIDEWIRE FROM THE PROTECTIVE HOOP, THE BLUE SECTION AT THE HAND END WAS FOUND TO HAVE DETACHED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE AND THERE WERE NO KNOWN PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23173347 ↗MalfunctionSep 10, 2025Peeled/Delaminated; Adverse Event Without Identified Device or Use Problem; Peeled/Delaminated; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING UNPACKING, THE COATING OF THE GUIDEWIRE WAS PEELING OFF. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE WITHOUT PATIENT COMPLICATIONS. ANALYSIS OF THE RETURNED GUIDEWIRE IDENTIFIED THE SLEEVE WAS FULLY DETACHED FROM THE GUIDEWIRE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23018647 ↗MalfunctionAug 1, 2025Peeled/Delaminated; Adverse Event Without Identified Device or Use Problem; Peeled/Delaminated; Adverse Event Without Identified Device or Use Problem
T WAS REPORTED THAT A PATIENT UNDERWENT A RIGHT URETERAL METAL STENT PLACEMENT. THE GUIDEWIRE WAS USED TO GUIDE THE CATHETER PLACEMENT. DURING THE PROCEDURE, IT WAS FOUND THAT THE COATING OF THE GUIDEWIRE PEELED OFF. THE USE WAS STOPPED, CAUSING NO IMPACT ON THE PATIENT. THE PROCEDURE WAS COMPLETED WITH THE ORIGINAL DEVICE AND NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22687749 ↗MalfunctionJun 9, 2025Contamination; Use of Device Problem; Contamination; Use of Device Problem
IT WAS REPORTED THAT THE SENSOR WIRE EXPIRED APRIL 2025. IT WAS USED IN THE PATIENT AFTER EXPIRATION DATE. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22171966 ↗InjuryJun 5, 2025Break; Break
IT WAS REPORTED THAT THE PATIENT UNDERWENT CYSTOSCOPY, RIGHT RETROGRADE PYELOGRAM, URETEROSCOPIC HOLMIUM LASER LITHOTRIPSY, STONE BASKETING, AND TIGHT URETERAL STENT PLACEMENT. DURING STONE CLEARANCE AND FLUOROSCOPIC EVALUATION OF STONE CLEARANCE PROGRESS, THE WORKING SENSOR WIRE WAS NOT VISUALIZED. A SEPARATE WIRE FRAGMENT WAS SEEN ON FLUOROSCOPY IN THE UPPER POLE. THE WIRE FRAGMENT WAS VISUALIZED AND ATTEMPTS WERE MADE TO RETRIEVE IT USING A ZERO TIP BASKET AND PIRANHA FORCEPS. HOWEVER, RETRIEVAL OF THE BROKEN WIRE FRAGMENT WAS UNSUCCESSFUL. THE PATIENT WAS RETURNED TO THE OPERATING ROOM. A FLEXIBLE URETEROSCOPE WAS ADVANCED INTO THE KIDNEY. PYELOSCOPY WAS PERFORMED. THE RETAINED SENSOR WIRE FRAGMENT WAS IDENTIFIED AND SUCCESSFULLY GRASPED WITH A THREE-PRONG BASKET AND REMOVED. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Foreign Body In Patient
MAUDE report 22152028 ↗InjuryMay 9, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT A SENSOR WIRE WAS USED DURING URETEROSCOPY PROCEDURE. THE PHYSICIAN PLACED A SENSOR WIRE INTO THE URETER, LEFT IT IN PLACE, REMOVED THE OLD URETERAL STENT. IT WAS NOTED THAT A PIECE OF THE DISTAL END OF THE SENSOR WIRE HAS SHEARED OFF IN THE BLADDER. THE PIECE WAS 5 INCHES IN LENGTH. THE PHYSICIAN WAS ABLE TO REMOVE THE DETACHED PIECE FROM THE BLADDER. THE PROCEDURE WAS COMPLETED USING THE ORIGINAL DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21994985 ↗MalfunctionMar 27, 2025Material Deformation; Adverse Event Without Identified Device or Use Problem; Material Deformation; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING INSERTION, THE GUIDEWIRE WAS NOT SMOOTH ENOUGH TO USE. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE AND THE PATIENT CONDITION FOLLOWING THE PROCEDURE WAS STABLE. ANALYSIS OF THE RETURNED DEVICE REVEALED THAT THE SLEEVE WAS DETACHED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21714140 ↗
FDA clearance and classification
The FDA regulates product code EZB (stylet for catheter, gastro-urology) under 21 CFR 876.5130 as a Class I device. 21 CFR 876.5130 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Sensor M0066703080
Has the Sensor M0066703080 been recalled?
No FDA recall record in our data names this exact device identifier (DI 08714729376361), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Sensor M0066703080?
The FDA's MAUDE database lists 118 adverse event reports for this device, received between Jun 28, 2017 and Aug 28, 2026. In those reports, the most common specific device problems are “Detachment of Device or Device Component” (63), “Break” (29) and “Peeled/Delaminated” (28); the most common patient problems are “Foreign Body In Patient” (7), “Hematuria” (2) and “Pain” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Sensor M0066703080?
None of the reports are classified as deaths, 24 as injuries and 94 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Sensor M0066703080 FDA approved?
Devices of this type (product code EZB) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Boston Scientific Corporation.
Who makes the Sensor M0066703080, and is it still sold?
The FDA's device database lists Boston Scientific Corporation as the company responsible for it (brand Sensor, model M0066703080). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other STYLET FOR CATHETER, GASTRO-UROLOGY devices
- NiCore Nitinol Guidewire 150NSS35C. R. Bard, Inc. · STYLET FOR CATHETER, GASTRO-UROLOGY228 reports
- Sensor M0066703050Boston Scientific Corporation · STYLET FOR CATHETER, GASTRO-UROLOGY41 reports
- Bard® Hydro-Glide™ Guidewire 145FS35C. R. Bard, Inc. · STYLET FOR CATHETER, GASTRO-UROLOGY31 reports
- NiCore Nitinol Guidewire 150NFS35C. R. Bard, Inc. · STYLET FOR CATHETER, GASTRO-UROLOGY30 reports
- Amplatz Super Stiff M0066401090Boston Scientific Corporation · STYLET FOR CATHETER, GASTRO-UROLOGY21 reports
- Amplatz Super Stiff M0066401080Boston Scientific Corporation · STYLET FOR CATHETER, GASTRO-UROLOGY18 reports
- NiCore Nitinol Guidewire 150NSA35C. R. Bard, Inc. · STYLET FOR CATHETER, GASTRO-UROLOGY16 reports
- Sensor M0066703120Boston Scientific Corporation · STYLET FOR CATHETER, GASTRO-UROLOGY16 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page