Skip to content

VascuTrak® V1840150Recalls & reported problems

Bard Peripheral Vascular, Inc. · VascuTrak® PTA Dilatation Catheter 4 mm x 150 mm, 140 cm catheter · Class 2

All models of VascuTrak (20)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00801741061585), received between Dec 27, 2016 and Sep 23, 2024 — classified as 1 injury, 3 malfunctions.

The most frequently reported specific device problem is “Deflation Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Deflation Problem2
  • Break1
  • Difficult to Remove1
  • Material Rupture1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1
  • No Known Impact Or Consequence To Patient1
  • No Patient Involvement1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 23, 2024Deflation Problem; Difficult to Remove; Deflation Problem; Difficult to Remove

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN THE LEFT SUPERFICIAL FEMORAL ARTERY, IT WAS FOUND THAT THE PTA BALLOON ALLEGEDLY COULD NOT BE REMOVED FROM THE BODY. IT WAS FURTHER REPORTED THAT THE PTA BALLOON WAS WITHDRAWN FROM THE BODY AFTER AN ALLEGED EXTRACORPOREAL PUNCTURE OF THE BALLOON WITH A MICRO PENETRATING NEEDLE UNDER THE CONTRAST MEDIUM TO RELIEVE THE PRESSURE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20276688 ↗
  • MalfunctionJul 15, 2020Material Rupture; Material Rupture

    THIS REPORT SUMMARIZES ONE MALFUNCTION. A REVIEW OF THE REPORTED INFORMATION INDICATED THAT MODEL V1840150 PTA BALLOON DILATATION CATHETER ALLEGEDLY EXPERIENCED A MATERIAL RUPTURE. THIS INFORMATION WAS RECEIVED FROM A SINGLE SOURCE. THE MALFUNCTION INVOLVED A PATIENT WITH NO REPORTED CONSEQUENCES. THE PATIENT WAS REPORTED AS A 67-YEAR-OLD FEMALE WEIGHING 55 KGS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10279721 ↗
  • MalfunctionJul 17, 2018Break; Break

    IT WAS REPORTED THAT UPON OPENING OF THE DEVICE, THE EXTERNAL WIRE OF THE PTA BALLOON WAS ALLEGEDLY BROKEN. THERE WAS NO REPORTED PATIENT INVOLVEMENT.

    Patient problems: No Patient Involvement

    MAUDE report 7696783 ↗
  • InjuryDec 27, 2016Deflation Problem; Deflation Problem

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN THE SFA, THE PTA BALLOON CATHETER ALLEGEDLY WOULD NOT DEFLATE DURING THE FIRST ATTEMPT; THEREFORE, A NEEDLE STICK THROUGH THE SKIN WAS REQUIRED TO DEFLATE THE BALLOON. THE PTA BALLOON WAS EXCHANGED OVER THE GUIDEWIRE FOR ANOTHER THAT WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO REPORTED CLINICAL CONSEQUENCE TO THE PATIENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6205226 ↗

Other Catheter, percutaneous, cutting/scoring devices

All Catheter, percutaneous, cutting/scoring devices →

Sources and updates