VascuTrak® V186040Recalls & reported problems
Bard Peripheral Vascular, Inc. · VascuTrak® PTA Dilatation Catheter 6 mm x 40 mm, 140 cm catheter · Class 2
All models of VascuTrak (20)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00801741061752), received between Dec 7, 2018 and Nov 30, 2021 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Material Deformation” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Material Deformation3
- Deflation Problem1
- Device-Device Incompatibility1
- Difficult to Insert1
- Failure to Advance1
- Retraction Problem1
Patient problems
- No Consequences Or Impact To Patient3
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 30, 2021Deflation Problem; Deflation Problem
IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE PTA BALLOON ALLEGEDLY FAILED TO DEFLATE. IT WAS FURTHER REPORTED THAT THE BALLOON AND SHEATH WERE REMOVED TOGETHER AS ONE UNIT. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12906185 ↗MalfunctionJul 16, 2020Failure to Advance; Material Deformation; Failure to Advance; Material Deformation
THIS REPORT SUMMARIZES ONE MALFUNCTION. A REVIEW OF THE REPORTED INFORMATION INDICATED THAT MODEL V186040 PTA DILATATION CATHETER ALLEGEDLY EXPERIENCED FAILURE TO CROSS LESION AND HAD A KINKED CATHETER SHAFT. THIS INFORMATION WAS RECEIVED FROM ONE SOURCE. OF THE ONE MALFUNCTION, ONE PATIENT WAS INVOLVED WITH NO PATIENT CONSEQUENCE OR IMPACT. THE AGE, WEIGHT AND GENDER OF THE PATIENT WAS NOT PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10286927 ↗MalfunctionJan 15, 2020Difficult to Insert; Material Deformation; Difficult to Insert; Material Deformation
THIS REPORT SUMMARIZES ONE MALFUNCTION. A REVIEW OF THE REPORTED INFORMATION INDICATED THAT MODEL V186040 VASCUTRACK PTA DILATATION CATHETER ALLEGEDLY EXPERIENCED MATERIAL DEFORMATION (KINK) AND DIFFICULT TO INSERT ISSUES. THIS INFORMATION WAS RECEIVED FROM ONE SOURCE. OF THE ONE MALFUNCTION, ONE PATIENT WAS INVOLVED WITH NO PATIENT CONSEQUENCE OR IMPACT. PATIENT'S AGE, WEIGHT, AND GENDER WAS NOT PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9594171 ↗MalfunctionDec 7, 2018Retraction Problem; Device-Device Incompatibility; Material Deformation; Retraction Problem; Device-Device Incompatibility; Material Deformation
IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN THE SFA THE SCORING WIRE OF THE PTA BALLOON ALLEGEDLY KINKED DURING THE SECOND INFLATION ATTEMPT. IT WAS FURTHER REPORTED THAT THERE WAS RESISTANCE REMOVING THE GUIDEWIRE FROM THE BALLOON. REPORTEDLY, THERE WAS DIFFICULTY RETRACTING THE BALLOON THROUGH THE SHEATH. THE PROCEDURE WAS CONCLUDED, AND NO FURTHER TREATMENT WAS PROVIDED. THERE WAS NO REPORTED PATIENT INJURY
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8144319 ↗
Other Catheter, percutaneous, cutting/scoring devices
- AngioSculpt OTW PTA Scoring Balloon Catheter 2092-6040The Spectranetics Corporation · Catheter, Percutaneous, Cutting/Scoring41 reports
- AngioSculpt OTW PTA Scoring Balloon Catheter 2076-5040The Spectranetics Corporation · Catheter, Percutaneous, Cutting/Scoring31 reports
- AngioSculpt OTW PTA Scoring Balloon Catheter 2092-5040The Spectranetics Corporation · Catheter, Percutaneous, Cutting/Scoring28 reports
- 2cm Peripheral Cutting Balloon® M001PCB8020900Boston Scientific Corporation · Catheter, percutaneous, cutting/scoring18 reports
- AngioSculpt OTW PTA Scoring Balloon Catheter 2105-6040The Spectranetics Corporation · Catheter, Percutaneous, Cutting/Scoring17 reports
- AngioSculpt OTW PTA Scoring Balloon Catheter 2076-6040The Spectranetics Corporation · Catheter, Percutaneous, Cutting/Scoring17 reports
- ULTRASCORE US14150530Bard Peripheral Vascular, Inc. · Catheter, percutaneous, cutting/scoring15 reports
- ULTRASCORE US3513058Bard Peripheral Vascular, Inc. · Catheter, percutaneous, cutting/scoring14 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page