Conquest® CQ7554Recalls & reported problems
Bard Peripheral Vascular, Inc. · Conquest® PTA Balloon Dilatation Catheter 5 mm x 40 mm, 75 cm catheter · Class 2
All models of Conquest (32)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00801741062834), received between May 10, 2017 and May 31, 2022 — classified as 9 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Material Rupture4
- Peeled/Delaminated4
- Material Frayed3
- Unraveled Material3
- Detachment of Device or Device Component2
- Retraction Problem2
- Burst Container or Vessel1
- Deflation Problem1
Patient problems
- No Consequences Or Impact To Patient7
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 31, 2022Deflation Problem; Material Frayed; Peeled/Delaminated; Unraveled Material; Detachment of Device or Device Component; Deflation Problem; Material Frayed; Peeled/Delaminated; Unraveled Material; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN THE SUPERFICIAL FEMORAL ARTERY, THE DEVICE ALLEGEDLY FAILED TO DEFLATE. IT WAS FURTHER REPORTED THAT THE PHYSICIAN ASSISTED WITH GUIDING TO REMOVE IT. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14546144 ↗MalfunctionOct 10, 2021Material Frayed; Unraveled Material; Material Frayed; Unraveled Material
IT WAS REPORTED THAT POST ANGIOPLASTY PROCEDURE, UNRAVELED MATERIAL WAS FOUND ON THE PTA BALLOON. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12602646 ↗MalfunctionOct 27, 2020Material Rupture; Material Rupture
THIS REPORT SUMMARIZES ONE MALFUNCTION. A REVIEW OF THE REPORTED INFORMATION INDICATED THAT MODEL CQ7554 PTA BALLOON DILATATION CATHETER ALLEGEDLY EXPERIENCED BALLOON RUPTURE. THIS INFORMATION WAS RECEIVED FROM ONE SOURCE. THIS MALFUNCTION INVOLVED ONE PATIENT WITH NO CONSEQUENCES. THE PATIENT'S AGE, WEIGHT AND GENDER WERE NOT PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10742775 ↗MalfunctionNov 20, 2018Leak/Splash; Peeled/Delaminated; Leak/Splash; Peeled/Delaminated
IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE OF A FOREARM FISTULA, CONTRAST WAS ALLEGEDLY LEAKING AT THE DISTAL END OF THE PTA BALLOON UPON INFLATION. REPORTEDLY, ANOTHER BALLOON WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8091044 ↗MalfunctionJun 25, 2018Burst Container or Vessel; Retraction Problem; Material Rupture; Detachment of Device or Device Component; Burst Container or Vessel; Retraction Problem; Material Rupture; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING AN ANGIOPLASTY IN AN UPPER ARM DIALYSIS FISTULA, THE PTA BALLOON ALLEGEDLY RUPTURED AND WAS DIFFICULT TO RETRACT THROUGH THE SHEATH, WHICH LED TO A BALLOON DETACHMENT. THE PTA BALLOON AND SHEATH WERE REMOVED TOGETHER AS ONE UNIT. ANOTHER PTA BALLOON WAS USED TO SUCCESSFULLY COMPLETE THE PROCEDURE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7636304 ↗MalfunctionMay 24, 2018Peeled/Delaminated; Material Rupture; Peeled/Delaminated; Material Rupture
IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE PTA BALLOON ALLEGEDLY RUPTURED AT THE DISTAL PORTION OF THE LESION AFTER A FEW INFLATION/DEFLATION ATTEMPTS. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7543251 ↗MalfunctionMay 17, 2018Peeled/Delaminated; Unraveled Material; Peeled/Delaminated; Unraveled Material
IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE THE BALLOON FIBERS ALLEGEDLY UNRAVELED UPON REMOVING FROM THE PATIENT. ANOTHER BALLOON WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7523937 ↗MalfunctionAug 31, 2017Retraction Problem; Material Rupture; Retraction Problem; Material Rupture
IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE PTA BALLOON ALLEGEDLY RUPTURED PRIOR TO REACHING 30ATM. REPORTEDLY, THE HEALTH CARE PROVIDER ALLEGEDLY HAD DIFFICULTY REMOVING THE PTA BALLOON THROUGH SHEATH. THE PTA BALLOON WAS EXCHANGED OVER THE GUIDEWIRE FOR ANOTHER THAT WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6834962 ↗MalfunctionMay 10, 2017Material Frayed; Material Frayed
IT WAS REPORTED THAT DURING THE VASCULAR ACCESS INTERVENTIONAL THERAPY (VAIVT), FRAYED OUTER MATERIAL OF THE PTA BALLOON WAS ALLEGEDLY FOUND WHEN IT WAS REMOVED FROM THE PATIENT VIA A SHEATH. REPORTEDLY, BEFORE THE REMOVAL, THE CATHETER WAS INFLATED TO A PRESSURE OF 14 ATM ONCE IN THE LESION. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6559402 ↗
Other CATHETER, PERCUTANEOUS devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths654 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall369 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall28 deaths302 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths278 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS297 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS220 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page