Conquest® CQ5064Recalls & reported problems
Bard Peripheral Vascular, Inc. · Conquest® PTA Balloon Dilatation Catheter 6 mm x 40 mm, 50 cm catheter · Class 2
All models of Conquest (32)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Bard Conquest PTA balloon dilatation catheter is used to widen narrowed blood vessels during angioplasty. Most of the 68 reports describe the balloon bursting, fraying, or its material separating from the catheter while inflating, with blood leaking into the pressure pump; procedures were typically finished with another device. There are no deaths and one injury, plus one report of vessel perforation. No recalls name this device, and reports in the last 12 months (3) are fewer than the 8 in the prior year.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 69
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 69 adverse event reports that list this exact device (DI 00801741062872), received between Feb 27, 2017 and Sep 28, 2026 — classified as 1 injury, 68 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (60 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Material Rupture60
- Leak/Splash10
- Material Frayed10
- Peeled/Delaminated10
- Unraveled Material9
- Break5
- Difficult to Remove3
- Deflation Problem2
Patient problems
- No Clinical Signs, Symptoms or Conditions60
- No Consequences Or Impact To Patient8
- Perforation of Vessels1
Recent adverse event reports
The 25 most recent of 69 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 28, 2026Material Rupture
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50921287 ↗MalfunctionApr 30, 2026Material Rupture
ON (B)(6) 2026, A PATIENT UNDERWENT A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) PROCEDURE USING THE CONQUEST BALLOON CATHETER. DURING THE PROCEDURE, THE BALLOON CATHETER ALLEGEDLY RUPTURED. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25043345 ↗MalfunctionFeb 27, 2026Break
ON (B)(6) 2025, A PATIENT UNDERWENT AN ANGIOPLASTY PROCEDURE USING THE CONQUEST PTA BALLOON. DURING THE PROCEDURE, THE BALLOON MATERIAL ALLEGEDLY SEPARATES FROM THE CATHETER DUE TO POOR ADHESION. THE PROCEDURE WAS COMPLETED USING ANOTHER BALLOON. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24466252 ↗MalfunctionNov 3, 2025Break; Material Rupture; Break; Material Rupture
ON (B)(6) 2025, A PATIENT UNDERWENT A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY PROCEDURE USING THE CONQUEST PTA BALLOON CATHETER BY ULTRASOUND-GUIDED THROUGH THE SITE OF UPPER LIMB VEIN TO ACCESS THE CEPHALIC VEIN. DURING THE PROCEDURE THE BALLOON WAS ALLEGEDLY RUPTURED AT 14 ATM. AND THERE WAS BLOOD IN THE PRESSURE PUMP DURING RETRIEVAL. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23448356 ↗MalfunctionSep 25, 2025Material Rupture; Material Rupture
ON (B)(6) 2025, A PATIENT UNDERWENT AN ANGIOPLASTY (PTA) PROCEDURE USING THE CONQUEST BALLOON CATHETER. DURING THE PROCEDURE, UPON INFLATION TO 26 ATMOSPHERES FOR THE THIRD TIME, AIR LEAKAGE OCCURRED. THE PROCEDURE WAS COMPLETED USING ANOTHER BALLOON. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23140755 ↗MalfunctionMay 16, 2025Material Rupture; Material Rupture
ON (B)(6) 2025 DURING AN ANGIOPLASTY PROCEDURE IN LEFT FOREARM AVF SURGERY, STENOSIS IN THE AVF PUNCTURE AREA AND 2 YEARS POST PROCEDURE SAME INCISION SITE WAS SELECTED FOR A FISTULA VEIN ACCESS, THE PTA BALLOON ALLEGEDLY STARTS TO LEAK, THE RUPTURE OF THE BALLOON IS CONSIDERED. THE PROCEDURE WAS COMPLETED USING ANOTHER BALLOON. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22033298 ↗MalfunctionApr 30, 2025Material Rupture; Difficult to Open or Remove Packaging Material; Material Rupture; Difficult to Open or Remove Packaging Material
IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE PTA BALLOON ALLEGEDLY HAD A LEAKAGE AFTER DILATATION AND PRESSURIZATION TO 27 ATM. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21932199 ↗MalfunctionApr 1, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT PRIOR TO AN ANGIOPLASTY PROCEDURE, THE PTA BALLOON ALLEGEDLY DETACHED AT THE ANTERIOR SEGMENT OF THE DEVELOPER. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21736439 ↗MalfunctionMar 7, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE PTA BALLOON ALLEGEDLY RUPTURED AT 25 ATM. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21545898 ↗MalfunctionMar 5, 2025Break; Break
IT WAS REPORTED THAT PRIOR TO AN ANGIOPLASTY PROCEDURE, THE PTA BALLOON WAS ALLEGEDLY DETACHED. IT WAS FURTHER REPORTED THAT THE DETACHED PART WAS STILL ATTACHED. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21522731 ↗MalfunctionJan 30, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE BALLOON ALLEGEDLY HAD A LEAKAGE AFTER PRESSURIZATION UP TO 25 ATM. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21270915 ↗MalfunctionDec 27, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE PTA BALLOON ALLEGEDLY WAS LEAKING. THE PROCEDURE WAS COMPLETED USING ANOTHER BALLOON. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21024646 ↗MalfunctionSep 12, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT DURING AN ULTRASOUND-GUIDED ANGIOPLASTY PROCEDURE VIA RADIAL ARTERY PUNCTURE, THE PTA BALLOON ALLEGEDLY RUPTURED AT 25 ATM. THE PROCEDURE WAS COMPLETED USING ANOTHER BALLOON. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20206370 ↗MalfunctionSep 12, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN THE LEFT INTERNAL FOREARM FISTULA, THE PTA BALLOON ALLEGEDLY LEAKED AT 28 ATM. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20198944 ↗MalfunctionJul 25, 2024Break; Break
IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE VASCULAR SHEATH WAS ALLEGEDLY BROKEN. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19824551 ↗MalfunctionMay 15, 2024Leak/Splash; Detachment of Device or Device Component; Leak/Splash; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE TAIL END ALLEGEDLY DISCONNECTED WHEN THE DEVICE WAS WITHDRAWN. IT WAS FURTHER REPORTED THAT THE DEVICE ALLEGEDLY LEAKED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19317980 ↗MalfunctionMar 26, 2024Break; Leak/Splash; Difficult to Remove; Material Rupture; Break; Leak/Splash; Difficult to Remove; Material Rupture
IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, WHEN THE PRESSURE WAS PUMPED TO 8 ATM DURING REPEAT DILATATION, PRESSURE PUMP ABNORMALITY WAS NOTICED AND BLOOD WAS ALLEGEDLY SEEN ON GENTLE RETRACTION AND THE PTA BALLOON ALLEGEDLY COULD NOT BE FULLY RETRACTED. REPORTEDLY, THE BALLOON ALLEGEDLY CRACKED AND LEAKED. THE PROCEDURE WAS COMPLETED BY USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18976712 ↗MalfunctionMar 22, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE UNDER THE RIGHT FOREARM ENDOVASCULAR FISTULA, THE PTA BALLOON ALLEGEDLY RUPTURED AT LESS THAN 15 ATM. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18957535 ↗MalfunctionMar 20, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN THE LEFT FOREARM ENDOVASCULAR FISTULA, DURING THE PRESSURIZATION PROCESS, AIR LEAKAGE WAS ALLEGEDLY FOUND, AND THE PTA BALLOON WAS ALLEGEDLY FOUND TO BE LEAKING WHEN REMOVED. THE PROCEDURE WAS COMPLETED BY USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18939148 ↗MalfunctionMar 13, 2024Material Frayed; Material Rupture; Material Frayed; Material Rupture
IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN THE RIGHT UPPER LIMB, THE PTA BALLOON ALLEGEDLY RUPTURED AT 26 ATM. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18892540 ↗MalfunctionFeb 26, 2024Peeled/Delaminated; Material Rupture; Unraveled Material; Peeled/Delaminated; Material Rupture; Unraveled Material
IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE PTA BALLOON WAS ALLEGEDLY RUPTURED AT 18ATM. THE PROCEDURE WAS COMPLETED BY USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18775214 ↗MalfunctionFeb 23, 2024Material Frayed; Material Rupture; Unraveled Material; Material Frayed; Material Rupture; Unraveled Material
IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN THE VENOUS OUTFLOW TRACT THROUGH THE DISTAL RADIAL ARTERY APPROACH, THE PTA BALLOON ALLEGEDLY RUPTURED WHEN THE FIRST PRESSURE WAS EXPANDED AND THE AIR PRESSURE PUMP WAS PRESSURIZED TO ABOUT 28 ATM. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18773888 ↗MalfunctionFeb 23, 2024Material Frayed; Material Rupture; Unraveled Material; Material Frayed; Material Rupture; Unraveled Material
IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN THE RIGHT FOREARM, AFTER THE FIRST EXPANSION WHEN THE PRESSURE PUMP WAS PRESSURIZED TO 30 ATM, THE PTA BALLOON CONTRAST ALLEGEDLY LEAKED AND CANNOT BE EXPANDED NORMALLY. IT WAS FURTHER REPORTED THAT THE HEAD OF THE PTA BALLOON WAS ALLEGEDLY FOUND TO BE RUPTURED AFTER THE BALLOON WAS REMOVED. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18769133 ↗MalfunctionDec 21, 2023Material Rupture; Material Deformation; Material Rupture; Material Deformation
IT WAS REPORTED THAT DURING AN ULTRASOUND GUIDED ANGIOPLASTY PROCEDURE, THE PTA BALLOON ALLEGEDLY RUPTURED AT 26 ATM. IT WAS FURTHER REPORTED THAT THE SHEATH WAS REMOVED AT THE SAME TIME AND THE PROCEDURE WAS COMPLETED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18371681 ↗MalfunctionDec 11, 2023Material Rupture; Material Rupture
IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE PTA BALLOON WAS ALLEGEDLY RUPTURED AT 10 ATM. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18301088 ↗
FDA clearance and classification
The FDA regulates product code DQY (catheter, percutaneous) under 21 CFR 870.1250 as a Class II device. 21 CFR 870.1250 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Conquest® CQ5064
Has the Conquest® CQ5064 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00801741062872), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Conquest® CQ5064?
The FDA's MAUDE database lists 69 adverse event reports for this device, received between Feb 27, 2017 and Sep 28, 2026. In those reports, the most common specific device problems are “Material Rupture” (60), “Leak/Splash” (10) and “Material Frayed” (10); the most common patient problem is “Perforation of Vessels” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Conquest® CQ5064?
None of the reports are classified as deaths, 1 as an injury and 68 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the Conquest® CQ5064, and is it still sold?
The FDA's device database lists Bard Peripheral Vascular, Inc. as the company responsible for it (brand Conquest®, model CQ5064). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other CATHETER, PERCUTANEOUS devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths651 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall362 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall31 deaths315 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths276 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS293 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS219 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page