Conquest® CQ7564Recalls & reported problems
Bard Peripheral Vascular, Inc. · Conquest® PTA Balloon Dilatation Catheter 6 mm x 40 mm, 75 cm catheter · Class 2
All models of Conquest (32)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Bard Conquest PTA balloon dilatation catheter is used to open narrowed blood vessels during angioplasty. The 57 reports mostly describe the balloon or its push rod breaking, rupturing, or unraveling during the procedure, and in one case the balloon would not deflate. There are no deaths and 1 injury, and no recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 55
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 55 adverse event reports that list this exact device (DI 00801741062902), received between Jan 12, 2017 and Jan 12, 2026 — classified as 1 injury, 54 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (24 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Material Rupture24
- Peeled/Delaminated19
- Unraveled Material16
- Material Frayed12
- Material Deformation5
- Break4
- Leak/Splash4
- Retraction Problem4
Patient problems
- No Consequences Or Impact To Patient32
- No Clinical Signs, Symptoms or Conditions22
- No Patient Involvement2
- Device Embedded In Tissue or Plaque1
Recent adverse event reports
The 25 most recent of 55 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 12, 2026Material Rupture
ON (B)(6) 2025, A PATIENT UNDERWENT AN ANGIOPLASTY PROCEDURE USING THE CONQUEST PTA BALLOON. DURING THE PROCEDURE, THE BALLOON ALLEGEDLY GOT RUPTURED. THE PROCEDURE WAS COMPLETED USING ANOTHER BALLOON. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24033472 ↗MalfunctionOct 31, 2025Deflation Problem; Inflation Problem; Difficult to Remove; Deflation Problem; Inflation Problem; Difficult to Remove
ON (B)(6) 2025, A PATIENT UNDERWENT AN ANGIOPLASTY PROCEDURE USING THE CONQUEST PTA BALLOON. DURING THE PROCEDURE, THE BALLOON ALLEGEDLY REFUSED TO DEFLATE. IT WAS FURTHER REPORTED THAT THE BALLOON WAS PUNCTURED WITH A NEEDLE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23433570 ↗MalfunctionJun 3, 2025Burst Container or Vessel; Burst Container or Vessel
ON (B)(6) 2025, A PATIENT UNDERWENT A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) PROCEDURE USING THE CONQUEST PTA BALLOON CATHETER. DURING THE PROCEDURE, THE BALLOON WAS ALLEGEDLY NOT BEEN INFLATED. IT WAS FURTHER REPORTED THAT AN ULTRASOUND FOUND THAT THE PUSH ROD WAS DAMAGED, AND AFTER REMOVING IT FROM THE PATIENT'S BODY, IT WAS DETERMINED THAT THE PUSH ROD HAD BURST DURING PRESSURIZATION. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22129564 ↗MalfunctionApr 11, 2025Break; Burst Container or Vessel; Break; Burst Container or Vessel
IT WAS REPORTED THAT DURING AN UPPER LIMB VENOUS BALLOON ANGIOPLASTY PROCEDURE OF THE LEFT ANTERIOR FOREARM, THE PUSH ROD OF THE BALLOON ALLEGEDLY BROKEN. THE BALLOON WAS REMOVED UNDER ULTRASOUND GUIDANCE. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21823091 ↗MalfunctionSep 12, 2024Peeled/Delaminated; Material Rupture; Unraveled Material; Material Deformation; Peeled/Delaminated; Material Rupture; Unraveled Material; Material Deformation
IT WAS REPORTED THAT DURING AN ULTRASOUND GUIDED ANGIOPLASTY PROCEDURE IN THE PROXIMAL ANASTOMOTIC STENOSIS VIA THE RIGHT FOREARM OF RADIAL ARTERY, THE PTA BALLOON ALLEGEDLY RUPTURED AT 24 ATM. IT WAS FURTHER REPORTED THAT THE PTA BALLOON WAS REMOVED FROM THE SHEATH. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20208689 ↗MalfunctionJul 2, 2024Inflation Problem; Material Rupture; Material Deformation; Inflation Problem; Material Rupture; Material Deformation
IT WAS REPORTED THAT DURING AN ULTRASOUND GUIDED PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY PROCEDURE IN THE UPPER LIMB VEINS OF ARTERIOVENOUS FISTULA ANASTOMOSIS, THE PTA BALLOON WAS USED WITH A DILATATION PRESSURE OF 30 ATM, AND WHEN IT WAS BEING MAINTAINED, THE PRESSURE WAS ALLEGEDLY SUDDENLY RELEASED, AND THE PRESSURE COULD NOT BE GIVEN UP. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19655500 ↗MalfunctionMay 14, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT DURING AN PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY PROCEDURE, THE PTA BALLOON WAS ALLEGEDLY FOUND TO BE EITHER HAD A HOLE OR ALLEGEDLY BURST. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19308297 ↗MalfunctionFeb 23, 2024Material Rupture; Unraveled Material; Material Rupture; Unraveled Material
IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN THE VENOUS OUTFLOW TRACT, THE PTA BALLOON ALLEGEDLY RUPTURED WHEN THE FIRST PRESSURE WAS EXPANDED AND THE AIR PRESSURE PUMP WAS PRESSURIZED TO ABOUT 28 ATM. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18771033 ↗MalfunctionJan 9, 2024Material Frayed; Material Rupture; Material Deformation; Material Frayed; Material Rupture; Material Deformation
IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE BALLOON WAS FILLED WITH SALINE TO DILATE THE STENOSIS NEAR THE PATIENT'S ANASTOMOSIS AND THE PRESSURE REACHED 30ATM. AFTER THE SECOND TIME, IT BROKE AND WAS SAFELY RECOVERED WITHOUT CAUSING DAMAGE TO THE PATIENT'S BLOOD VESSELS. IT WAS FURTHER REPORTED THAT THE MIDDLE PART OF THE BALLOON WAS RUPTURED. THE OPERATION WAS COMPLETED BY CHANGING THE NEW DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18472632 ↗MalfunctionJan 3, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT DURING A LEFT FOREARM FISTULA ANGIOPLASTY PROCEDURE, THE BALLOON WAS ALLEGEDLY RUPTURED AT 26 ATM WHEN THE PRESSURE WAS APPLIED. THE PROCEDURE WAS COMPLETED BY USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18433458 ↗MalfunctionDec 29, 2023Material Rupture; Material Rupture
IT WAS REPORTED THAT DURING AN ULTRASOUND GUIDED ANGIOPLASTY PROCEDURE IN UPPER EXTREMITY VENOUS, THE PTA BALLOON ALLEGEDLY BURST AT 24 ATM. THE PROCEDURE WAS COMPLETED USING ANOTHER BALLOON. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18417712 ↗MalfunctionDec 22, 2023Material Rupture; Material Rupture
IT WAS REPORTED THAT DURING A RIGHT ANTERIOR INTERNAL FISTULA ANGIOPLASTY PROCEDURE, THE PTA BALLOON CATHETER ALLEGEDLY RUPTURED WHEN THE PRESSURE REACHED 25 ATM. THE PROCEDURE WAS COMPLETED BY USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18382717 ↗MalfunctionOct 9, 2023Difficult to Remove; Difficult to Remove
IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE TIP OF THE PTA BALLOON ALLEGEDLY WOULD NOT COME OFF WHEN RETRACTING THE BALLOON. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17894507 ↗MalfunctionNov 4, 2022Leak/Splash; Peeled/Delaminated; Material Rupture; Unraveled Material; Leak/Splash; Peeled/Delaminated; Material Rupture; Unraveled Material
IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE THROUGH INTERNAL ARTERIOVENOUS FISTULA OF CEPHALIC VEIN, THE PTA BALLOON ALLEGEDLY RUPTURED AT 28 ATM DURING THIRD INFLATION. IT WAS FURTHER REPORTED THAT THE DEVICE AND INTRODUCER SHEATH AS ONE UNIT AND THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15724108 ↗MalfunctionNov 2, 2022Detachment of Device or Device Component; Device Dislodged or Dislocated; Detachment of Device or Device Component; Device Dislodged or Dislocated
IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE PTA BALLOON HUB ALLEGEDLY HAD DISLODGED FROM THE CATHETER. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15713165 ↗MalfunctionSep 27, 2022Material Frayed; Peeled/Delaminated; Material Rupture; Unraveled Material; Material Frayed; Peeled/Delaminated; Material Rupture; Unraveled Material
IT WAS REPORTED THAT DURING AN ULTRASOUND GUIDED ANGIOPLASTY PROCEDURE, THE PTA BALLOON ALLEGEDLY RUPTURED AT 26 ATM. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15493139 ↗MalfunctionSep 27, 2022Material Frayed; Material Rupture; Unraveled Material; Material Frayed; Material Rupture; Unraveled Material
IT WAS REPORTED THAT DURING AN ULTRASOUND -GUIDED ANGIOPLASTY PROCEDURE, THE PTA BALLOON RUPTURE ALLEGEDLY RUPTURED AT 23 ATM. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15493058 ↗MalfunctionJun 16, 2021Deflation Problem; Inflation Problem; Difficult to Remove; Deflation Problem; Inflation Problem; Difficult to Remove
IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE PTA BALLOON ALLEGEDLY HAD FAILED INFLATE AND DEFLATE. IT WAS FURTHER REPORTED THAT THE DEVICE HAD DIFFICULTY IN REMOVING. THE PROCEDURE WAS COMPLETED BY USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12008608 ↗MalfunctionMay 4, 2021Break; Material Rupture; Break; Material Rupture
IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE PTA BALLOON ALLEGEDLY RUPTURED AND CATHETER WAS DAMAGED. THE PROCEDURE WAS COMPLETED BY USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11768381 ↗MalfunctionMay 3, 2021Break; Break
IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE DEVICE ALLEGEDLY BROKE. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11760710 ↗MalfunctionNov 16, 2020Material Rupture; Material Rupture
IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE PTA BALLOON WAS ALLEGEDLY RUPTURED AT 20 ATM. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10842202 ↗MalfunctionOct 28, 2020Material Rupture; Material Rupture
THIS REPORT SUMMARIZES ONE MALFUNCTION. A REVIEW OF THE REPORTED INFORMATION INDICATES THE MODEL CQ7564 PTA BALLOON DILATATION CATHETER ALLEGEDLY EXPERIENCED MATERIAL RUPTURE. THIS REPORT WAS RECEIVED FROM ONE SOURCE. THE EVENT INVOLVED A PATIENT WITH NO KNOWN IMPACT TO THE PATIENT. THE PATIENT AGE, GENDER AND WEIGHT WAS NOT PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 10752612 ↗MalfunctionOct 26, 2020Unraveled Material; Unraveled Material
THIS REPORT SUMMARIZES ONE MALFUNCTION. A REVIEW OF THE REPORTED INFORMATION INDICATES THAT MODEL CQ7564 PTA BALLOON DILATATION CATHETER ALLEGEDLY EXPERIENCED UNRAVELED MATERIAL. THE INFORMATION WAS RECEIVED FROM A SINGLE SOURCE. THIS MALFUNCTION INVOLVED ONE PATIENT WITH NO PATIENT CONSEQUENCES. THE PATIENT'S AGE, WEIGHT AND GENDER WERE NOT PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10734367 ↗MalfunctionJul 16, 2020Material Rupture; Material Rupture
THIS REPORT SUMMARIZES ONE MALFUNCTION. A REVIEW OF THE REPORTED INFORMATION INDICATES THAT MODEL CQ7564 PTA BALLOON DILATATION CATHETER ALLEGEDLY RUPTURED. THE INFORMATION WAS RECEIVED FROM A SINGLE SOURCE. ONE PATIENT WAS INVOLVED WITH NO REPORTED PATIENT INJURY. AGE, WEIGHT AND GENDER WERE NOT PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10287373 ↗MalfunctionJul 15, 2020Material Rupture; Material Deformation; Material Rupture; Material Deformation
THIS REPORT SUMMARIZES ONE MALFUNCTION. A REVIEW OF THE REPORTED INFORMATION INDICATED THAT MODEL CQ7564J PTA BALLOON DILATATION CATHETER ALLEGEDLY EXPERIENCED MATERIAL RUPTURE AND MATERIAL DEFORMATION. THIS INFORMATION WAS RECEIVED FROM ONE SOURCE. THIS MALFUNCTION INVOLVED A PATIENT WITH NO CONSEQUENCES. THE (B)(6) YEAR OLD FEMALE PATIENT WEIGHED (B)(6) KGS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10279826 ↗
FDA clearance and classification
The FDA regulates product code DQY (catheter, percutaneous) under 21 CFR 870.1250 as a Class II device. 21 CFR 870.1250 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Conquest® CQ7564
Has the Conquest® CQ7564 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00801741062902), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Conquest® CQ7564?
The FDA's MAUDE database lists 55 adverse event reports for this device, received between Jan 12, 2017 and Jan 12, 2026. In those reports, the most common specific device problems are “Material Rupture” (24), “Peeled/Delaminated” (19) and “Unraveled Material” (16); the most common patient problem is “Device Embedded In Tissue or Plaque” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Conquest® CQ7564?
None of the reports are classified as deaths, 1 as an injury and 54 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the Conquest® CQ7564, and is it still sold?
The FDA's device database lists Bard Peripheral Vascular, Inc. as the company responsible for it (brand Conquest®, model CQ7564). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other CATHETER, PERCUTANEOUS devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths651 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall362 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall31 deaths315 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths276 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS293 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS219 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page