Skip to content

Conquest® CQ7568Recalls & reported problems

Bard Peripheral Vascular, Inc. · Conquest® PTA Balloon Dilatation Catheter 6 mm x 80 mm, 75 cm catheter · Class 2

All models of Conquest (32)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00801741062919), received between Jun 20, 2018 and May 14, 2024 — classified as 9 malfunctions.

The most frequently reported specific device problem is “Material Frayed” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Frayed5
  • Peeled/Delaminated5
  • Material Rupture4
  • Unraveled Material3
  • Deflation Problem1
  • Inflation Problem1
  • Retraction Problem1

Patient problems

  • No Consequences Or Impact To Patient7
  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 14, 2024Material Rupture; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE PTA BALLOON WAS ALLEGEDLY RUPTURED AT 10 ATM AT THE FIRST DILATION. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19307439 ↗
  • MalfunctionNov 26, 2020Material Rupture; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE PTA BALLOON ALLEGEDLY RUPTURED AT 12ATM. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10910273 ↗
  • MalfunctionJul 16, 2020Inflation Problem; Peeled/Delaminated; Material Rupture; Inflation Problem; Peeled/Delaminated; Material Rupture

    THIS REPORT SUMMARIZES ONE MALFUNCTION. A REVIEW OF THE REPORTED INFORMATION INDICATED THAT MODEL CQ7568 PTA DILATATION CATHETER ALLEGEDLY EXPERIENCED PINHOLE BALLOON RUPTURE AND HAD INFLATION ISSUE. THIS INFORMATION WAS RECEIVED FROM ONE SOURCE. OF THE ONE MALFUNCTION, ONE PATIENT WAS INVOLVED WITH NO PATIENT CONSEQUENCE OR IMPACT. THE PATIENT WAS REPORTED TO BE A (B)(6) YEAR OLD FEMALE WEIGHING (B)(6) LBS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10286847 ↗
  • MalfunctionJul 16, 2020Material Frayed; Material Frayed

    THIS REPORT SUMMARIZES ONE MALFUNCTION. A REVIEW OF THE REPORTED INFORMATION INDICATES THAT MODEL CQ7568 PTA BALLOON DILATATION CATHETER ALLEGEDLY EXPERIENCED MATERIAL FRAYED. THE INFORMATION WAS RECEIVED FROM ONE SOURCE. ONE PATIENT WAS INVOLVED WITH NO REPORTED PATIENT INJURY. AGE, WEIGHT AND GENDER WERE NOT PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10286349 ↗
  • MalfunctionApr 15, 2020Deflation Problem; Retraction Problem; Deflation Problem; Retraction Problem

    THIS REPORT SUMMARIZES ONE MALFUNCTION. A REVIEW OF THE REPORTED INFORMATION INDICATED THAT MODEL CQ7568 PTA BALLOON DILATATION CATHETER ALLEGEDLY EXPERIENCED DEFLATION PROBLEM AND RETRACTION PROBLEM. THIS REPORT WAS RECEIVED FROM A SINGLE SOURCE. THIS EVENT DID INVOLVE PATIENT WITH NO REPORTED PATIENT INJURY. THE PATIENTS AGE, WEIGHT AND GENDER WERE NOT PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9965563 ↗
  • MalfunctionDec 13, 2018Material Frayed; Peeled/Delaminated; Material Rupture; Material Frayed; Peeled/Delaminated; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE PTA BALLOON ALLEGEDLY RUPTURED AT 8 ATM. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8162596 ↗
  • MalfunctionJun 21, 2018Material Frayed; Peeled/Delaminated; Unraveled Material; Material Frayed; Peeled/Delaminated; Unraveled Material

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN THE RADIAL ARTERY, UPON REMOVING THE PTA BALLOON FROM THE PATIENT, UNRAVELED MATERIAL WAS ALLEGEDLY FOUND. NO FURTHER TREATMENT WAS REQUIRED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7627141 ↗
  • MalfunctionJun 20, 2018Material Frayed; Peeled/Delaminated; Unraveled Material; Material Frayed; Peeled/Delaminated; Unraveled Material

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN THE RADIAL ARTERY, UPON REMOVING THE PTA BALLOON FROM THE PATIENT, UNRAVELED MATERIAL WAS FOUND. NO FURTHER TREATMENT WAS REQUIRED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7623519 ↗
  • MalfunctionJun 20, 2018Material Frayed; Peeled/Delaminated; Unraveled Material; Material Frayed; Peeled/Delaminated; Unraveled Material

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN THE RADIAL ARTERY, UPON REMOVING THE PTA BALLOON FROM THE PATIENT, UNRAVELED MATERIAL WAS FOUND. NO FURTHER TREATMENT WAS REQUIRED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7623515 ↗

Other CATHETER, PERCUTANEOUS devices

All CATHETER, PERCUTANEOUS devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page