Conquest® CQ7574Recalls & reported problems
Bard Peripheral Vascular, Inc. · Conquest® PTA Balloon Dilatation Catheter 7 mm x 40 mm, 75 cm catheter · Class 2
All models of Conquest (32)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Bard Conquest balloon dilatation catheter is used to open narrowed blood vessels during angioplasty. Most reports describe the balloon breaking, cracking, kinking or fraying during the procedure, and sometimes being hard to pull back out through the sheath. There are 65 reports, with 3 injuries and no deaths, and no recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 64
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 64 adverse event reports that list this exact device (DI 00801741062995), received between Nov 16, 2016 and Sep 9, 2026 — classified as 3 injuries, 61 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (33 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Material Rupture33
- Peeled/Delaminated10
- Material Deformation9
- Deflation Problem8
- Material Frayed8
- Retraction Problem6
- Burst Container or Vessel5
- Difficult to Remove4
Patient problems
- No Clinical Signs, Symptoms or Conditions30
- No Consequences Or Impact To Patient27
- No Patient Involvement4
- Extravasation1
- Needle Stick/Puncture1
- Perforation of Vessels1
- Reaction1
Recent adverse event reports
The 25 most recent of 64 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 9, 2026Deflation Problem; Entrapment of Device
On (b)(6), 2026, a patient underwent an angioplasty procedure in left forearm using the Conquest PTA balloon. During the procedure, following balloon dilation, the device could not be withdrawn. To facilitate retrieval, the balloon was punctured with a cannula to achieve deflation, after which it was successfully removed from the patient. The procedure was completed using another balloon. There was no reported patient injury.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50687620 ↗MalfunctionSep 8, 2026Material Rupture
On (b)(6) 2026, a patient underwent a percutaneous transluminal angioplasty procedure using a Conquest balloon catheter. During the procedure, while using a Bard Ultra High Pressure Balloon, it was observed that the inflation device repeatedly lost pressure during inflation attempts. The balloon was subsequently removed, and a simulated inflation test was performed outside the body, revealing a small tear in the balloon. Upon inflation, fluid sprayed directly from the tear, resulting in insufficient pressure. The procedure was successfully completed after the balloon was replaced with another balloon. There was no reported patient injury.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50663997 ↗MalfunctionAug 31, 2026Deflation Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50516918 ↗MalfunctionMay 19, 2026Material Rupture
ON (B)(6) 2026, A PATIENT UNDERWENT A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) PROCEDURE USING THE CONQUEST BALLOON CATHETER. DURING THE PROCEDURE, THE BALLOON CATHETER ALLEGEDLY RUPTURED. THE PROCEDURE WAS COMPLETED USING ANOTHER BALLOON. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25212109 ↗MalfunctionJan 29, 2026Material Rupture
ON (B)(6) 2026, A PATIENT UNDERWENT A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) PROCEDURE BY USING THE CONQUEST BALLOON CATHETER. IT WAS REPORTED THAT DURING THE PROCEDURE, THE BALLOON CATHETER ALLEGEDLY RUPTURED. THE PROCEDURE WAS COMPLETED USING ANOTHER BALLOON. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24207398 ↗MalfunctionNov 3, 2025Material Rupture; Unraveled Material; Material Rupture; Unraveled Material
ON (B)(6) 2025, A PATIENT UNDERWENT A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) PROCEDURE USING THE CONQUEST BALLOON CATHETER. DURING THE PROCEDURE, THE BALLOON CATHETER ALLEGEDLY RUPTURED. THE PROCEDURE WAS COMPLETED USING ANOTHER BALLOON. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23449835 ↗MalfunctionOct 28, 2025Burst Container or Vessel; Material Rupture; Burst Container or Vessel; Material Rupture
ON (B)(6) 2025, A PATIENT UNDERWENT A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) PROCEDURE USING THE CONQUEST BALLOON CATHETER. DURING THE PROCEDURE, THE BALLOON CATHETER ALLEGEDLY RUPTURED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23400910 ↗MalfunctionSep 25, 2025Burst Container or Vessel; Leak/Splash; Material Rupture; Burst Container or Vessel; Leak/Splash; Material Rupture
ON (B)(6) 2025, A PATIENT UNDERWENT AN ANGIOPLASTY PROCEDURE USING THE CONQUEST PTA BALLOON. DURING THE PROCEDURE, THE BALLOON ALLEGEDLY RUPTURED. IT WAS FURTHER REPORTED THAT THE CONTRAST AGENT LEAKED OUT. THE PROCEDURE WAS COMPLETED USING ANOTHER BALLOON. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23141142 ↗MalfunctionSep 10, 2025Difficult to Remove; Material Rupture; Difficult to Remove; Material Rupture
ON (B)(6) 2025, A PATIENT UNDERWENT AN ANGIOPLASTY PROCEDURE USING THE CONQUEST PTA BALLOON FOR THE TREATMENT OF STENOSIS. DURING THE PROCEDURE, THE BALLOON ALLEGEDLY BURST. ADDITIONALLY, IT WAS REPORTED THAT THERE WAS DIFFICULTY IN RETRACTING THE BALLOON THROUGH THE INTRODUCER SHEATH FOLLOWING THE BALLOON RUPTURE. THE PROCEDURE WAS COMPLETED USING ANOTHER BALLOON. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23012759 ↗MalfunctionAug 6, 2025Burst Container or Vessel; Burst Container or Vessel
ON (B)(6) 2025, A PATIENT UNDERWENT A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) PROCEDURE USING THE CONQUEST PTA BALLOON CATHETER. DURING THE PROCEDURE THE PRESSURE REACHED 26 ATM AND WAS BROKEN. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22730399 ↗MalfunctionJul 29, 2025Material Rupture; Material Rupture
ON (B)(6) 2025, A PATIENT UNDERWENT A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) PROCEDURE USING THE CONQUEST PTA BALLOON CATHETER TO TREAT STENOSIS. DURING THE PROCEDURE, THE BALLOON WAS RUPTURED. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22640618 ↗MalfunctionJun 10, 2025Stretched; Failure to Advance; Device-Device Incompatibility; Material Deformation; Stretched; Failure to Advance; Device-Device Incompatibility; Material Deformation
A PATIENT UNDERWENT AN ANGIOPLASTY PROCEDURE USING THE CONQUEST AND DURING THE PROCEDURE IT WAS REPORTED THAT THE GUIDEWIRE ALLEGEDLY CANNOT PASS THROUGH THE HUB. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22174118 ↗MalfunctionMay 16, 2025Material Rupture; Material Rupture
ON (B)(6) 2025, A PATIENT UNDERWENT A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) PROCEDURE USING THE CONQUEST PTA BALLOON CATHETER. DURING THE PROCEDURE, THE BALLOON WAS ALLEGEDLY RUPTURED. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22033074 ↗MalfunctionJan 13, 2025Material Rupture; Detachment of Device or Device Component; Material Deformation; Material Rupture; Detachment of Device or Device Component; Material Deformation
IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE PTA BALLOON ALLEGEDLY RUPTURED AT 18 ATM. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21134687 ↗MalfunctionAug 9, 2024Deflation Problem; Inflation Problem; Material Deformation; Deflation Problem; Inflation Problem; Material Deformation
IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE PTA BALLOON ALLEGEDLY DIFFICULT TO DEFLATE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19947726 ↗MalfunctionJul 22, 2024Break; Material Rupture; Break; Material Rupture
IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE PTA BALLOON ALLEGEDLY BLAST AT 22 ATM. FURTHER IT WAS REPORTED THAT THE BALLOON WAS REMOVED FROM THE SHEATH AND THE BALLOON WAS ALLEGEDLY FOUND TO BE LEAKING. THE PROCEDURE WAS COMPLETED USING ANOTHER BALLOON. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19792191 ↗MalfunctionApr 13, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT DURING PREPARATION OF AN ANGIOPLASTY PROCEDURE, WHEN THE PACKAGE WAS OPENED AND INSPECTED, THE PTA BALLOON WAS ALLEGEDLY FOUND TO BE LEAKING. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19102597 ↗MalfunctionFeb 27, 2024Leak/Splash; Difficult to Remove; Material Deformation; Leak/Splash; Difficult to Remove; Material Deformation
IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN THE ARTERIOVENOUS FISTULA REFLUX VEIN STENOSIS OF ARTIFICIAL BLOOD VESSELS, THE PTA BALLOON DILATATION CATHETER ALLEGEDLY LEAKED. IT WAS FURTHER REPORTED THAT THE REMOVAL OF THE CATHETER ALLEGEDLY CAUSED DEFORMATION OF THE CATHETER SHEATH. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18787305 ↗MalfunctionFeb 20, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT DURING AN ULTRASOUND GUIDED ANGIOPLASTY PROCEDURE, WHEN THE PTA BALLOON WAS REMOVED, A SMALL HOLE WAS ALLEGEDLY FOUND ON THE SIDE NEAR THE TIP OF THE BALLOON. THE PROCEDURE WAS COMPLETED BY USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18738414 ↗MalfunctionFeb 12, 2024Material Rupture; Unraveled Material; Material Rupture; Unraveled Material
IT WAS REPORTED THAT DURING THE ANGIOPLASTY PROCEDURE, MIDDLE PART OF THE PTA BALLOON ALLEGEDLY HAD RUPTURED. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18683283 ↗MalfunctionNov 2, 2022Leak/Splash; Material Split, Cut or Torn; Leak/Splash; Material Split, Cut or Torn
IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE CONNECTION BETWEEN THE BALLOON AND CATHETER ALLEGEDLY LEAKED AIR. THE PROCEDURE WAS COMPLETED BY USING ANOTHER DEVICE. THERE WAS NO REPORTED PATENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15713116 ↗MalfunctionMay 4, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT PRIOR TO AN ANGIOPLASTY PROCEDURE, THE PTA BALLOON HUB WAS ALLEGEDLY SNAPPED OFF UPON OPENING THE PACKAGE. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14274871 ↗MalfunctionApr 4, 2022Break; Leak/Splash; Break; Leak/Splash
IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE DEVICE ALLEGEDLY FOUND TO BE RUPTURED AT THE PUSH ROD NEAR THE BALLOON. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13995018 ↗MalfunctionDec 9, 2021Material Rupture; Material Rupture
IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE PTA BALLOON ALLEGEDLY RUPTURED WHEN INFLATED TO 36 ATM. THE PROCEDURE WAS COMPLETED USING ANOTHER BALLOON. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12966437 ↗MalfunctionNov 29, 2021Device Damaged Prior to Use; Packaging Problem; Device Damaged Prior to Use; Packaging Problem
IT WAS REPORTED THAT PRIOR TO AN ANGIOPLASTY PROCEDURE, THE BALLOON WAS ALLEGEDLY DAMAGED IN THE PACKAGE AND WHEN ATTEMPTED TO REMOVE THE BALLOON FROM PACKAGE IT WAS VERY DIFFICULT TO REMOVE. IT WAS FURTHER REPORTED THAT UPON REMOVAL , THE COVERING WAS NOT ON THE BALLOON PORTION AND IT GOT STUCK TO THE PLASTIC TUBING AND IT WOULD NOT ADVANCE OVER THE WIRE. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12881963 ↗
FDA clearance and classification
The FDA regulates product code DQY (catheter, percutaneous) under 21 CFR 870.1250 as a Class II device. 21 CFR 870.1250 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Conquest® CQ7574
Has the Conquest® CQ7574 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00801741062995), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Conquest® CQ7574?
The FDA's MAUDE database lists 64 adverse event reports for this device, received between Nov 16, 2016 and Sep 9, 2026. In those reports, the most common specific device problems are “Material Rupture” (33), “Peeled/Delaminated” (10) and “Material Deformation” (9); the most common patient problems are “Extravasation” (1), “Needle Stick/Puncture” (1) and “Perforation of Vessels” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Conquest® CQ7574?
None of the reports are classified as deaths, 3 as injuries and 61 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the Conquest® CQ7574, and is it still sold?
The FDA's device database lists Bard Peripheral Vascular, Inc. as the company responsible for it (brand Conquest®, model CQ7574). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other CATHETER, PERCUTANEOUS devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths651 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall362 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall31 deaths315 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths276 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS293 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS219 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record