Skip to content

Port Access Needle LH-0034YNRecalls & reported problems

Bard Access Systems, Inc. · SafeStep 19G x 1 inch with y-site · Class 2

All models of Port Access Needle (122)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00801741066276), received between Sep 13, 2019 and Nov 18, 2025 — classified as 9 malfunctions.

The most frequently reported specific device problem is “Material Deformation” (6 reports).

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 2 reportsJan 2025: 2 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 19, 2020Class IITerminated

    Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that have developed breaks or cracks can cause leakage resulting in exposure to chemicals such a chemotherapeutics or biohazard material (e.g., blood). A crack or break could result in potential blood loss, catheter occlusion, air embolism, under or interrupted infusion, or site contamination which could lead to infection.

    Z-2669-2020 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Material Deformation6
  • Contamination /Decontamination Problem5
  • Fluid/Blood Leak2
  • Break1
  • Leak/Splash1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • Chemical Exposure2

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 18, 2025Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED NON POWER PORTS LEAKING AT THE NEEDLESS CONNECTOR. THERE HAS NOT BEEN ANY REPORTED PATIENT HARM OR BAD OUTCOME FROM THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23584400 ↗
  • MalfunctionApr 2, 2025Break; Material Deformation; Break; Material Deformation

    THE CUSTOMER RETURNED FOUR DEVICES TO THE MANUFACTURER AND ALL FOUR WERE FOUND TO HAVE PLASTIC BROKE OFF THE END CAPS. THIS REPORT ADDRESSES ALL FOUR DEVICES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21754127 ↗
  • MalfunctionFeb 5, 2025Contamination /Decontamination Problem; Material Deformation; Contamination /Decontamination Problem; Material Deformation

    IT WAS REPORTED THAT THE NEEDLE HAS A SMALL PIECE OF PLASTIC HAS BROKEN OFF THE END WHITE CAP THAT IS ATTACHED INTO THE FEMALE HUB OF THE NEEDLE RESULTING IN A FOREIGN BODY WITHIN THE HUB. THIS SHOULD NOT BE THERE. THIS CAP SHOULD BE REMOVED INTACT WITH A HUB FREE OF A FOREIGN BODY. NO OTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21316070 ↗
  • MalfunctionJan 3, 2025Contamination /Decontamination Problem; Material Deformation; Contamination /Decontamination Problem; Material Deformation

    IT WAS REPORTED THAT THE CUSTOMER IS SEEING A PIECE OF THE END CAP SHAVING OFF AND FALLING INTO THE NEEDLE LUER UPON REMOVAL. THERE WAS NO HARM, INJURY, COMPLICATIONS OR UNPLANNED MEDICAL INTERVENTION REQUIRED. THIS SITUATION DID CAUSE A MINIMAL DELAY IN PATIENT TREATMENT AS THE NURSE NEEDED TO OBTAIN ANOTHER HUBER NEEDLE. NO OTHER INFORMATION PROVIDED, AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21068768 ↗
  • MalfunctionJan 2, 2025Contamination /Decontamination Problem; Material Deformation; Contamination /Decontamination Problem; Material Deformation

    IT WAS REPORTED THAT A 1 INCH HUBER NEEDLE CONTAINS LOOSE PLASTIC AND IS PRESENTS WHEN TWISTING OFF THE CAP. THE ISSUE WAS INTERCEPTED PRIOR TO USE ON A PATIENT. NO OTHER INFORMATION PROVIDED, AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21061325 ↗
  • MalfunctionDec 31, 2024Contamination /Decontamination Problem; Material Deformation; Contamination /Decontamination Problem; Material Deformation

    IT WAS REPORTED, "AFTER LUNCH TODAY ONE OF MY STAFF FOUND 2 1INCH NEEDLES LOT# ASJQFC044 WITH SMALL SHARDS OF PLASTIC. I LOOKED AT THEM CLOSER AND IT SEEMS AS IF THE PIECE OF PLASTIC IS BREAKING OFF THE CAP. I ALSO HAVE THE CAPS ON THE NEEDLES WITH THE SHARD IN THE FEMALE END OF THE TUBING. NO ADVERSE EVENTS OR INJURY OCCURRED REGARDING THIS ISSUE." NO OTHER INFORMATION WAS PROVIDED. IT WAS REPORTED THIS OCCURRED WITH TWO DEVICES. THIS REPORT ADDRESSES BOTH DEVICES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21050570 ↗
  • MalfunctionDec 31, 2024Contamination /Decontamination Problem; Material Deformation; Contamination /Decontamination Problem; Material Deformation

    IT WAS REPORTED THAT SMALL SHARDS OF PLASTIC IS BREAKING OFF THE CAP. NO ADVERSE EVENTS OR INJURY OCCURRED REGARDING THIS ISSUE. NO OTHER INFORMATION PROVIDED, AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21047629 ↗
  • MalfunctionSep 13, 2019Fluid/Blood Leak

    IT WAS REPORTED THAT THE ANTI-REFLUX VALVE LEAKED. THE PATIENT HAD PRODUCT ON HER ARM. THE TEAM HAD TO DECONTAMINATE IT. IT WAS STATED THAT THIS EVENT OCCURRED WITH TWO PATIENTS. THIS REPORT ADDRESSES THE FIRST PATIENT.

    Patient problems: Chemical Exposure

    MAUDE report 9035905 ↗
  • MalfunctionSep 13, 2019Leak/Splash

    IT WAS REPORTED THAT THE ANTI-REFLUX VALVE LEAKED. THE PATIENT HAD PRODUCT ON HER ARM. THE TEAM HAD TO DECONTAMINATE IT. IT WAS STATED THAT THIS EVENT OCCURRED WITH TWO PATIENTS. THIS REPORT ADDRESSES THE SECOND PATIENT.

    Patient problems: Chemical Exposure

    MAUDE report 9035866 ↗

Other Set, administration, intravascular devices

All Set, administration, intravascular devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page