Port Access Needle LH-0035YNRecalls & reported problems
Bard Access Systems, Inc. · SafeStep 22G x 0.5 inch with y-site · Class 2
All models of Port Access Needle (122)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 14
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 00801741066290), received between Jan 2, 2020 and Jan 9, 2025 — classified as 14 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (8 reports).
1 FDA recall record name this device's identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Jun 19, 2020Class IITerminated
Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that have developed breaks or cracks can cause leakage resulting in exposure to chemicals such a chemotherapeutics or biohazard material (e.g., blood). A crack or break could result in potential blood loss, catheter occlusion, air embolism, under or interrupted infusion, or site contamination which could lead to infection.
Z-2669-2020 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Fluid/Blood Leak8
- Defective Component3
- Break1
- Failure to Infuse1
- Obstruction of Flow1
- Positioning Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions10
- No Consequences Or Impact To Patient3
- No Patient Involvement2
Recent adverse event reports
The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 9, 2025Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT THE PORT NEEDLE HAD BEEN INSITU FOR ABOUT 3 DAYS WHEN THE NURSE NOTICED IT STARTED LEAKING WHEN THEY WERE FLUSHING THROUGH ONE SIDE OF THE Y SET. THE WAS NO INCREASE OF RESISTANCE DURING FLUSHING AND THE PORT IS NOT BLOCKED. THE PORT NEEDLE SET HAS ONE BONDED NEEDLELESS DEVICE AND ONE WHERE WE HAVE TO ATTACH A SINGLE NEEDLELESS DEVICE. IT WAS THE BONDED NEEDLELESS DEVICE THAT WAS LEAKING. NO HARM WAS CAUSED TO THE PATIENT, THE PORT NEEDLE WAS REMOVED EARLY DUE TO THE ISSUE AND A DIFFERENT NEEDLE INSERTED. NO OTHER INFORMATION PROVIDED, AT THIS TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21115053 ↗MalfunctionDec 29, 2023Defective Component; Defective Component
IT WAS REPORTED THAT WHEN BLOOD RETURN WAS ATTEMPTED TO BE CONFIRMED, AIR WAS ASPIRATED. THERE WAS NO REPORTED PATIENT INJURY. (B)(6) 2023 - A FUNCTIONAL TEST REVEALED THAT AIR COULD ENTER TWO INFUSION SETS THROUGH THE Y-SITE DURING ASPIRATION. THIS REPORT ADDRESSES THE SECOND DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18416957 ↗MalfunctionDec 28, 2023Defective Component; Defective Component
IT WAS REPORTED THAT WHEN BLOOD RETURN WAS ATTEMPTED TO BE CONFIRMED, AIR WAS ASPIRATED. THERE WAS NO REPORTED PATIENT INJURY. (B)(6) 2023 - A FUNCTIONAL TEST REVEALED THAT AIR COULD ENTER TWO INFUSION SETS THROUGH THE Y-SITE DURING ASPIRATION. THIS REPORT ADDRESSES THE FIRST DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18414231 ↗MalfunctionSep 5, 2023Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT WHEN THE DEVICE WAS PRIMED PRIOR TO USE, AIR WAS ASPIRATED WHEN THE PLUNGER OF THE SYRINGE WAS PULLED. THERE WAS NO REPORTED PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17684615 ↗MalfunctionJun 10, 2023Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED BY THE CUSTOMER WHEN THE DEVICE WAS PRIMED WITH SALINE SOLUTION PRIOR TO USE, AIR WAS ASPIRATED AND AIR CONTAMINATION WAS FOUND. THERE WAS NO REPORTED PATIENT INVOLVEMENT.IT WAS REPORTED THIS OCCURRED WITH TWO DEVICES. THIS REPORT ADDRESSES THE SECOND DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17102123 ↗MalfunctionJun 10, 2023Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED BY THE CUSTOMER WHEN THE DEVICE WAS PRIMED WITH SALINE SOLUTION PRIOR TO USE, AIR WAS ASPIRATED AND AIR CONTAMINATION WAS FOUND. THERE WAS NO REPORTED PATIENT INVOLVEMENT. IT WAS REPORTED THIS OCCURRED WITH TWO DEVICES. THIS REPORT ADDRESSES THE FIRST DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17102121 ↗MalfunctionNov 29, 2022Failure to Infuse; Failure to Infuse
IT WAS REPORTED THAT WHEN THE NEEDLE WAS CONNECTED TO THE SYRINGE AND ATTEMPTED TO BE PRIMED, AIR COULD NOT BE REPLACED AND AIR BUBBLES COULD NOT BE COME OUT. THERE WAS NO REPORTED PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15880336 ↗MalfunctionOct 18, 2022Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED BY THE CUSTOMER "ON 9/28 OUR TEAM REPORTED THAT A SAFESTEP HUBER NEEDLE WAS FOUND TO BE CRACKED AND LEAKING WHILE IN USE. THERE WAS NO HARM TO THE PATIENT. "
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15629265 ↗MalfunctionJul 8, 2021Break; Break
IT WAS REPORTED "THE ISSUE WAS WITH THE PT.'S IVAD. THE IVAD WAS ACCESSED TO THE PT. AND WAS HEP LOCKED AND MOM NOTICED THAT THE CAP AND THE CAP CONNECTOR AT THE END OF THE LUMEN COMPLETELY CAME OFF ON IT OWN. THERE IS A CLEAN BREAK FROM THE FROM THE LUMEN AND CAP CONNECTOR."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12140096 ↗MalfunctionJan 8, 2021Obstruction of Flow; Positioning Problem
IT WAS REPORTED THAT THE BASE OF THE SAFESTEP COULD NOT BE MOVED WHEN THE NURSE CHECKED BEFORE USE, THEN IT WAS LOCKED BY SPD. THERE WAS NO REPORTED PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Patient Involvement
MAUDE report 11142777 ↗MalfunctionJul 15, 2020Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT THE LEAKAGE WAS FOUND AROUND THE BASE AFTER STAYED FLUSHING ALTHOUGH RESISTANCE WAS FELT, AFTER PRIMING WITH HEPARIN SALINE SOLUTION, PUNCTURING INTO THE PORT AND BLOOD RETURN CONFIRMED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10279950 ↗MalfunctionJun 19, 2020Defective Component; Defective Component
IT WAS REPORTED THAT IT WAS FOUND THAT THE NEEDLE COVER WAS DETACHED FROM THE NEEDLE, SO THAT THE NEEDLE TIP WAS UNCOVERED UPON OPENING THE PACKAGE. THERE WAS NO REPORTED PATIENT CONTACT.
Patient problems: No Patient Involvement
MAUDE report 10176651 ↗MalfunctionJan 2, 2020Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT BY MEDICAL ORDER AND WITH STERILE TECHNIQUE, AN IMPLANTABLE CATHETER CONTROL LABORATORY IS TAKEN AND LEAKAGE IS OBSERVED BETWEEN THE CONNECTION OF NEEDLE EQUIPMENT. THERE IS NO INPUT DATA (IMPLANTABLE NEEDLE NUMBER 22, PLACED ON (B)(6) 2019."
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9539753 ↗MalfunctionJan 2, 2020Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT SAFE STEP NEEDLE WITH LEAK.PARENTERAL NUTRITION WITHDRAWAL IS PERFORMED, CATHETER WASHING IS PERFORMED WITH PUSH STOP TECHNIQUE, DAMAGE IS OBSERVED (IN BLACK CLACE) OF THE 22X15 NEEDLE EXTENSION. 12/18/2019: THE CUSTOMER INFORMED THAT THE PRODUCT WAS USED IN THE PATIENT AS THE DEFECT WAS NOTICED DURING THE USE. THERE WAS NO DAMAGE TO THE PATIENT NOR THE HEALTH PROFESSIONAL.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9539681 ↗
Other Set, administration, intravascular devices
- Autosoft XC 1001680Unomedical A/S · Set, Administration, Intravascular1 death16,105 reports
- Autosoft90™ 1002817Unomedical A/S · Set, Administration, Intravascular1 recall11,907 reports
- Minimed™ Quick-set™ MMT-399AUnomedical A/S · Set, Administration, Intravascular9 deaths8,669 reports
- Medtronic Extended MMT-431AUnomedical A/S · Set, Administration, Intravascular3 deaths8,574 reports
- MiniMed™ Mio™ Advance MMT-242AUnomedical A/S · Set, Administration, Intravascular15 deaths8,164 reports
- CADD 21-7302-24Icu Medical, Inc. · Set, administration, intravascular4 recalls1 death4,657 reports
- Autosoft XC 1001681Unomedical A/S · Set, Administration, Intravascular1 death4,216 reports
- Minimed™ Quick-set™ MMT-397AUnomedical A/S · Set, Administration, Intravascular19 deaths4,134 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page