Skip to content

Port Access Needle LH-0037YNRecalls & reported problems

Bard Access Systems, Inc. · SafeStep 20G x 0.5 inch with y-site · Class 2

All models of Port Access Needle (122)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00801741066337), received between Mar 19, 2021 and Mar 2, 2023 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Jun 19, 2020Class IITerminated

    Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that have developed breaks or cracks can cause leakage resulting in exposure to chemicals such a chemotherapeutics or biohazard material (e.g., blood). A crack or break could result in potential blood loss, catheter occlusion, air embolism, under or interrupted infusion, or site contamination which could lead to infection.

    Z-2669-2020 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Detachment of Device or Device Component1
  • Fluid/Blood Leak1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 2, 2023Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THE ONCOLOGY DEPARTMENT ENCOUNTERED A LEAK THIS MORNING ON A HUBER NEEDLE. THE LEAK OCCURRED BETWEEN THE WHITE PART AND THE TRANSPARENT PART OF THE CONNECTION VALVE. THE IDE HAD TO REMOVE THE NEEDLE AND USE ANOTHER ONE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16475486 ↗
  • MalfunctionMar 19, 2021Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT "THE NEEDLE DOES NOT GET FIXED TO ITS HUB, IT'S LOOSE, LEADING TO AN INSTABILITY AND JEOPARDIZING ITS FUNCTIONALITY." NO OTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11529106 ↗

Other Set, administration, intravascular devices

All Set, administration, intravascular devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page