Skip to content

SureStep™ Foley Tray System Bardex® All- Silicone Foley Catheter Tray A902416Recalls & reported problems

C. R. Bard, Inc. · SureStep® Foley Tray System Bard Silicone Foley Catheter Tray 350 ml Urine Meter · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
9
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00801741074042), received between Feb 21, 2018 and Jan 3, 2025 — classified as 9 malfunctions.

The most frequently reported specific device problem is “Component Misassembled” (2 reports).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 25, 2026Class IIOpen, Classified

    Due to complaints received indicating presence of contamination on the surface of pericare wipes provided with various Foley and Intermittent Catheter Trays.

    Z-3115-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  2. Aug 5, 2021Class IIOpen, Classified

    There is potential for packaging defects that may impact the sterile barrier.

    Z-2539-2021 · Root cause: Packaging change control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Component Misassembled2
  • Component Missing2
  • Burst Container or Vessel1
  • Deflation Problem1
  • Device Fell1
  • Difficult to Insert1
  • Inflation Problem1
  • Leak/Splash1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • No Known Impact Or Consequence To Patient4
  • No Consequences Or Impact To Patient2
  • No Patient Involvement1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 3, 2025Component Misassembled; Component Misassembled

    IT WAS REPORTED THAT THE PLACE ON THE FOLEY METER BAG WHERE THE DRAIN TUBING WAS ATTACHED HAD MANUFACTURING ISSUES. WHEN THE DRAIN TUBE WAS HEAT PRESSED ON THE PROCESS WAS ALSO MELTING THE BACK PORTION OF THE BAG TO THE FRONT AND TUBING AREA ESSENTIALLY SEALING THE BACK OF THE BAG TO THE TUBING HOLE ON THE FRONT AND MAKING IT UNABLE TO EMPTY THE BAGS OF URINE. IF THE CUSTOMER ATTEMPTED TO SEPARATE THE AREA, THE EDGE OF THE BAG OR THE AREA AROUND THE TUBING WOULD TEAR AND SPILL OR LEAK URINE OUT ONTO THE FLOOR OR ON THE HANDS OF THE CUSTOMER. CUSTOMER STATED THAT THIS HAD BEEN OCCURRING OFTEN SINCE THE BAGS WERE REDESIGNED AND THE DRAIN TUBE WITH PLASTIC CLAMP WAS ADDED TO REPLACE THE DRAIN TUBE WITH METAL CLAMP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21066316 ↗
  • MalfunctionSep 15, 2023Component Missing; Component Missing

    IT WAS REPORTED THAT FOLEY TRAY DID NOT CONTAIN A LABEL AND ONLY HAD 1CC OF STERILE WATER VERSUS 10CC.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17754829 ↗
  • MalfunctionSep 7, 2023Short Fill; Short Fill

    IT WAS REPORTED THAT FOLEY TRAY DID NOT CONTAIN A LABEL AND ONLY HAD 1CC OF STERILE WATER VERSUS 10CC .

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17697792 ↗
  • MalfunctionMar 31, 2022Component Misassembled; Component Misassembled

    IT WAS REPORTED THAT THE DRAINAGE PORT CONNECTED TO THE GREEN PORT TO DRAIN URINE OUT OF THE FOLEY DRAIN BAG APPEARED TO BE FUSED TOGETHER WHERE THERE WAS JUST A TRICKLE OF URINE ALLOWED TO DRAIN. HOWEVER, THE BAG COULD BE DRAINED IF FLOWED BACK INTO THE HARD PLASTIC SQUARE PART IN THE FRONT OF THE BAG WHERE THE NOZZLE COULD BE SCREWED OFF TO EMPTY THE URINE. STATED THAT THE FOLEY CATHETER WAS REMOVED FROM PATIENT AND FURTHER INVESTIGATION SHOWED THE BAG WAS DEFINITELY FUSED TOGETHER AT THE SITE OF EMPTYING INTO THE GREEN PORT TO EMPTY. WHEN THE BAG WAS PULLED ON IN THAT AREA TO SEE IF IT WOULD RELEASE, IT TORE. CUSTOMER STATED THAT NEW FOLEY PLACED IN PATIENT WITH NO ISSUES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13965906 ↗
  • MalfunctionApr 14, 2021Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging

    IT WAS REPORTED THAT THE FOLEY TRAY WAS PUNCTURED AND GOT DAMAGED, AND SO THE MEDICAL SUPPLIES WERE REFUSED BY THE DISTRIBUTOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11665942 ↗
  • MalfunctionApr 9, 2019Component Missing

    IT WAS REPORTED THAT THE SURESTEP FOLEY TRAY ARRIVED WITHOUT THE FOLEY CATHETER INCLUDED.

    Patient problems: No Known Impact Or Consequence To Patient; No Patient Involvement

    MAUDE report 8494966 ↗
  • MalfunctionSep 26, 2018Inflation Problem; Difficult to Insert; Leak/Splash; Material Puncture/Hole; Leak/Splash

    IT WAS REPORTED THAT ONCE THE CATHETER WAS INSERTED INTO THE PATIENT; IT FELL OUT ONCE THE BALLOON WAS INFLATED. A SECOND CATHETER WAS THEN INFLATED OUTSIDE OF THE PATIENT AND IT WAS DETERMINED THAT THERE WAS A PIN HOLE LEAK ON THE BALLOON. NO MEDICAL INTERVENTION WAS REQUIRED.

    Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient

    MAUDE report 7909426 ↗
  • MalfunctionSep 25, 2018Deflation Problem; Device Fell; Deflation Problem

    IT WAS REPORTED THAT ONCE THE CATHETER WAS INSERTED INTO THE PATIENT; IT FELL OUT ONCE THE BALLOON WAS INFLATED. NO MEDICAL INTERVENTION WAS REQUIRED.

    Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient

    MAUDE report 7907718 ↗
  • MalfunctionFeb 21, 2018Burst Container or Vessel

    IT WAS REPORTED THAT THE IODINE PACKET HAD BURST IN THE FOLEY TRAY.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7287900 ↗

Other CATHETER, RETENTION TYPE, BALLOON devices

All CATHETER, RETENTION TYPE, BALLOON devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page