Skip to content

Extension Set for PowerPICC SOLO Catheter 1090000Recalls & reported problems

Bard Access Systems, Inc. · Adaptor, PowerPICC · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00801741074714), received between Jul 10, 2025 and Jul 10, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Material Frayed” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Frayed1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 10, 2025Material Frayed; Material Frayed

    IT WAS REPORTED "LAST WEEK, WE HAD A WIRE BREAK DURING A PICC INSERTION FOR A 4FR PICC (LOT #REJX3266). THIS WAS INSERTED BY ONE OF OUR EXPERIENCED STAFF, WHO REPORTED THE LINE WAS CUT TO 36 CM AND THE COPPER TIP WAS NOT VISIBLE AFTER THE PROCEDURE. A SECOND NURSE CONFIRMED THAT THE WIRE WASN¿T PRESENT IN THE TRIMMED SECTION OF THE PICC, SO THE PATIENT WAS SENT FOR AN X-RAY AS A PRECAUTIONARY MEASURE. THE PATIENT HAS A PACEMAKER IN PLACE SO NEEDED A CHEST X RAY TO CONFIRM PICC TIP PLACEMENT. THE 3CG WIRE WAS PULLED BACK, LINE WAS CUT ONLY WITH THE BLADE PROVIDED. THE CUT WAS SMOOTH WITH NO SENSATION THE WIRE WAS CUT. THE INSERTER WAS CERTAIN THEY SEEN THE COPER SECTION AFTER THEY TRIMMED THE CATHETER AND PULLED IT BACK TO BE IN LINE WITH THE TIP OF THE PICC. THE PICC WAS THEN FLUSHED TO REMOVE MICRO BUBBLES ONTO THE TRAY. SHERLOCK SHOWED LOLLYPOP GOING UP NECK, MULTIPLE ATTEMPTS MADE TO NAVIGATE IT DOWN TOWARDS THE HEART. THE 3CG SHOWED THE PICC TAKING THE TURN TO GO DOWN TOWARDS THE HEART BUT DID COULD NOT GET THE DEPTH PARAMETER TO SHOW THE PICC GOING ALL THE WAY DOWN BUT SINCE THERE WAS GOOD BLOOD RETURN AND LINE FLUSHED EASILY ORDER FOR X RAY WAS ENTERED. THEN WHEN WIRE WAS REM

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22473385 ↗

Other Set, administration, intravascular devices

All Set, administration, intravascular devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page