GHIATAS® 477201Recalls & reported problems
Bard Peripheral Vascular, Inc. · MR GHIATAS® Beaded Breast Localization Wire, 20g x 7cm needle · Class 1 · Implant
All models of GHIATAS (8)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00801741079924), received between Dec 1, 2016 and Oct 23, 2020 — classified as 1 injury, 6 malfunctions.
The most frequently reported specific device problem is “Signal Artifact/Noise” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Signal Artifact/Noise6
- Device Markings/Labelling Problem3
- Break1
- Image Display Error/Artifact1
Patient problems
- No Consequences Or Impact To Patient6
- Injury1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 23, 2020Break; Image Display Error/Artifact; Break; Image Display Error/Artifact
THIS REPORT SUMMARIZES ONE MALFUNCTION. A REVIEW OF THE REPORTED INFORMATION INDICATED THAT MODEL 477201 BIOPSY INSTRUMENT EXPERIENCED BREAK AND IMAGE DISPLAY ERROR/ARTIFACT. THIS INFORMATION WAS RECEIVED FROM ONE SOURCE. THE MALFUNCTION INVOLVED A PATIENT WITH NO KNOWN IMPACT TO THE PATIENT. THE AGE, WEIGHT AND GENDER OF THE PATIENT WERE NOT PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10731360 ↗MalfunctionOct 16, 2019Signal Artifact/Noise; Device Markings/Labelling Problem; Signal Artifact/Noise; Device Markings/Labelling Problem
THIS REPORT SUMMARIZES FOUR MALFUNCTIONS. A REVIEW OF THE REPORTED INFORMATION INDICATES THAT MODEL 477201 LOCALIZATION WIRE ALLEGEDLY PRODUCED ARTIFACTS DURING IMAGING. THIS INFORMATION WAS RECEIVED FROM ONE SOURCE. THIS INVOLVED A PATIENT WITH NO CONSEQUENCES. THE PATIENT WAS REPORTED AS 86 YEARS OLD, WEIGHING 114 POUNDS, THE GENDER OF THE PATIENT WAS NOT PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9200122 ↗MalfunctionOct 16, 2019Signal Artifact/Noise; Signal Artifact/Noise
THIS REPORT SUMMARIZES SIX MALFUNCTIONS. A REVIEW OF THE REPORTED INFORMATION INDICATED THAT MODEL 477201 BIOPSY INSTRUMENT ALLEGEDLY EXPERIENCED A SIGNAL ARTIFACT. THIS INFORMATION WAS RECEIVED FROM VARIOUS SOURCES. OF THE SIX MALFUNCTIONS, ONE INVOLVED A PATIENT WITH NO KNOWN IMPACT TO THE PATIENT AND FIVE DID NOT INVOLVE A PATIENT. THE PATIENTS¿ AGE, WEIGHT, AND GENDER WERE NOT PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9198865 ↗InjuryAug 27, 2019Signal Artifact/Noise; Signal Artifact/Noise
IT WAS REPORTED THAT POST PLACEMENT OF A LOCALIZATION WIRE, AN ARTIFACT ALLEGEDLY WAS FOUND ON MRI IMAGING. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. THE CURRENT PATIENT STATUS IS UNKNOWN.
Patient problems: Injury
MAUDE report 8937675 ↗MalfunctionMay 30, 2018Signal Artifact/Noise; Device Markings/Labelling Problem; Signal Artifact/Noise; Device Markings/Labelling Problem
IT WAS REPORTED THAT POST PLACEMENT OF A LOCALIZATION WIRE, AN ARTIFACT ALLEGEDLY WAS FOUND ON MR IMAGING. THE LOCALIZATION WIRE REPORTEDLY WAS NON-MR AND INCORRECTLY PACKAGED WITH MR LABELING. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED WITH THE SAME WIRE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7554246 ↗MalfunctionMay 26, 2017Signal Artifact/Noise; Signal Artifact/Noise
IT WAS REPORTED THAT ALLEGEDLY DURING A BREAST LOCALIZATION WIRE PLACEMENT INTO NORMAL TISSUE DENSITY, THE HEALTH CARE PROVIDER WAS UNABLE TO CONFIRM THE PLACEMENT OF THE WIRE BECAUSE OF THE ALLEGED BLOOMING ARTIFACT FROM THE WIRE. THERE WAS NO REPORT OF PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6596021 ↗MalfunctionDec 1, 2016Signal Artifact/Noise; Device Markings/Labelling Problem; Signal Artifact/Noise; Device Markings/Labelling Problem
IT WAS REPORTED THAT DURING A BREAST LOCALIZATION WIRE PLACEMENT, THE DEVICE ALLEGEDLY CREATED AN ARTIFACT UNDER MRI IMAGING. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED WITH ANOTHER WIRE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6140118 ↗
FDA clearance and classification
The FDA regulates product code GDF (guide, needle, surgical) under 21 CFR 878.4800 as a Class I device. 21 CFR 878.4800 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the GHIATAS® 477201
Has the GHIATAS® 477201 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00801741079924), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the GHIATAS® 477201?
The FDA's MAUDE database lists 7 adverse event reports for this device, received between Dec 1, 2016 and Oct 23, 2020. In those reports, the most common specific device problems are “Signal Artifact/Noise” (6), “Device Markings/Labelling Problem” (3) and “Break” (1); the most common patient problem is “Injury” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the GHIATAS® 477201?
None of the reports are classified as deaths, 1 as an injury and 6 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the GHIATAS® 477201 FDA approved?
Devices of this type (product code GDF) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Bard Peripheral Vascular, Inc.
Is the GHIATAS® 477201 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.
Who makes the GHIATAS® 477201, and is it still sold?
The FDA's device database lists Bard Peripheral Vascular, Inc. as the company responsible for it (brand GHIATAS®, model 477201). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other GUIDE, NEEDLE, SURGICAL devices
- GHIATAS® 47920Bard Peripheral Vascular, Inc. · GUIDE, NEEDLE, SURGICAL20 reports
- DuaLok™ LW0107Bard Peripheral Vascular, Inc. · GUIDE, NEEDLE, SURGICAL10 reports
- PERL Localization Device R2020Focal Therapeutics, Inc. · Guide, Needle, Surgical7 reports
- GHIATAS® 475201Bard Peripheral Vascular, Inc. · GUIDE, NEEDLE, SURGICAL6 reports
- GHIATAS® 479201Bard Peripheral Vascular, Inc. · GUIDE, NEEDLE, SURGICAL5 reports
- GHIATAS® 47720Bard Peripheral Vascular, Inc. · GUIDE, NEEDLE, SURGICAL5 reports
- DuaLok™ LW0077Bard Peripheral Vascular, Inc. · GUIDE, NEEDLE, SURGICAL4 reports
- DuaLok™ LW0057Bard Peripheral Vascular, Inc. · GUIDE, NEEDLE, SURGICAL3 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page