Bard® CritiCore® Monitor 000002NRecalls & reported problems
C. R. Bard, Inc. · Bard® CritiCore® Monitor · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This C. R. Bard CritiCore monitor measures urine output and temperature. The 42 reports mostly describe inaccurate readings—especially urine volume and temperature—along with display or sensing problems and occasional breakage. They mention no deaths or injuries. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 42
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 42 adverse event reports that list this exact device (DI 00801741080777), received between Aug 10, 2017 and Apr 24, 2020 — classified as 42 malfunctions.
The most frequently reported specific device problem is “Volume Accuracy Problem” (9 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Volume Accuracy Problem9
- Incorrect, Inadequate or Imprecise Result or Readings8
- Incorrect Measurement7
- Break6
- Improper Flow or Infusion4
- Device Displays Incorrect Message3
- Device Sensing Problem3
- Erratic or Intermittent Display3
Patient problems
- No Known Impact Or Consequence To Patient25
- No Patient Involvement13
- No Consequences Or Impact To Patient12
- No Information11
Recent adverse event reports
The 25 most recent of 42 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 24, 2020Overheating of Device; Overheating of Device
IT WAS REPORTED THAT, UPON COMPLETION OF THE INVESTIGATION ON 17JUL2019, THE CRITICORE U5 COMPONENT ON THE DISPLAY PCB WAS BURNED AND THE BACKLIGHT DID NOT GO OUT.
Patient problems: No Patient Involvement
MAUDE report 9995152 ↗MalfunctionJun 4, 2019Volume Accuracy Problem; Volume Accuracy Problem; Output Problem
IT WAS REPORTED THAT THE CRITICORE MONITOR WAS NOT COUNTING THE HOURLY OUTPUT CORRECTLY.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 8669259 ↗MalfunctionMay 16, 2019Break; Incorrect, Inadequate or Imprecise Result or Readings; Break
IT WAS REPORTED THAT THE CRITICORE MONITOR WAS NOT READING RIGHT.
Patient problems: No Known Impact Or Consequence To Patient; No Patient Involvement
MAUDE report 8617332 ↗MalfunctionMay 16, 2019Break; Incorrect, Inadequate or Imprecise Result or Readings; Break
IT WAS REPORTED THAT THE CRITICORE MONITOR WAS NOT READING CORRECTLY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8616220 ↗MalfunctionMay 14, 2019Unable to Obtain Readings; Volume Accuracy Problem; Volume Accuracy Problem
IT WAS REPORTED THAT THE CRITICORE DEVICE WAS NOT READING VOLUME CORRECTLY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8607768 ↗MalfunctionMay 1, 2019Break; Display or Visual Feedback Problem; Break
IT WAS REPORTED THAT THERE WAS INTERMITTENT DISPLAY AT THE TEMPERATURE PROBE CONNECTION OF THE CRITICORE DEVICE.
Patient problems: No Information; No Known Impact Or Consequence To Patient
MAUDE report 8569245 ↗MalfunctionMar 28, 2019Erratic or Intermittent Display; Calibration Problem; Erratic or Intermittent Display; Calibration Problem
PER COMPLETION OF THE INVESTIGATION, IT WAS FOUND THAT THE OBSERVED MEASURED TEMPERATURE WAS READING LOW WHEN THE TEST TEMPERATURE PLUG WAS USED.
Patient problems: No Patient Involvement
MAUDE report 8461640 ↗MalfunctionFeb 6, 2019Erratic or Intermittent Display; Incorrect, Inadequate or Imprecise Result or Readings; Erratic or Intermittent Display
IT WAS REPORTED THAT THE CRITICORE DEVICE WAS NOT REGISTERING THE CORRECT TEMPERATURE. THE TEMPERATURE WAS OFF BY 10 DEGREES.
Patient problems: No Known Impact Or Consequence To Patient; No Patient Involvement
MAUDE report 8311188 ↗MalfunctionFeb 6, 2019Break; Device Sensing Problem; Break
IT WAS REPORTED THAT THE CRITICORE DEVICE WAS NOT READING THE TEMPERATURE CORRECTLY. THE READING WAS 10 DEGREES OFF.
Patient problems: No Known Impact Or Consequence To Patient; No Patient Involvement
MAUDE report 8310106 ↗MalfunctionJan 30, 2019Break; Loose or Intermittent Connection; Incorrect, Inadequate or Imprecise Result or Readings; Break
IT WAS REPORTED THAT THE TEMPERATURE PORT HAD A LOSE CONNECTION CAUSING UNSTABLE READINGS.
Patient problems: No Information; No Patient Involvement
MAUDE report 8290605 ↗MalfunctionJan 10, 2019Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT THE MONITOR WAS INTERMITTENTLY SHOWING ITSELF AS NOT LEVEL, DURING THE EVALUATION OF THE DEVICE.
Patient problems: No Patient Involvement
MAUDE report 8232825 ↗MalfunctionJan 4, 2019Circuit Failure; Circuit Failure
IT WAS REPORTED THAT THE MEASURED TEMPERATURE WAS INTERMITTENT.
Patient problems: No Patient Involvement
MAUDE report 8216814 ↗MalfunctionDec 4, 2018Volume Accuracy Problem; Volume Accuracy Problem; Device Sensing Problem
IT WAS REPORTED THAT THE CRITICORE MONITOR WAS NOT READING THE URINE VOLUME.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 8130652 ↗MalfunctionOct 24, 2018Circuit Failure; Circuit Failure; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT THE VOLUME AND THE TEMPERATURE DO NOT READ CORRECTLY.
Patient problems: No Known Impact Or Consequence To Patient; No Patient Involvement
MAUDE report 8002231 ↗MalfunctionSep 26, 2018Display or Visual Feedback Problem
IT WAS REPORTED THAT THE DEVICE DISPLAYED AMBER LIGHTS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7909618 ↗MalfunctionSep 26, 2018Image Display Error/Artifact
IT WAS REPORTED THAT THE DEVICE WAS BLINKING AMBER LIGHTS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7909617 ↗MalfunctionSep 4, 2018Device Sensing Problem; Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT THE DEVICE WAS NOT READING THE TEMPERATURE NOR RECORDING THE AMOUNT OF OUTPUT.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 7842680 ↗MalfunctionAug 23, 2018Volume Accuracy Problem
IT WAS REPORTED THAT THE DEVICE DID NOT READ THE VOLUME.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 7809832 ↗MalfunctionJul 5, 2018Device Displays Incorrect Message; Calibration Problem; Calibration Problem
IT WAS REPORTED THAT AS THE UNIT WAS BEING USED ON A PATIENT, THE NURSE NOTICED THAT DEVICE DISPLAYED 'CHECK CONTAINER'. THE DEVICE DISPLAYED "NOT LEVEL" ALTHOUGH THE DEVICE WAS ON A LEVEL SURFACE. IT WAS ALSO NOTED THAT THE DEVICE DISPLAYED '---' WHEN THE CONTAINER VOLUME KEY WAS PRESSED AND THE TANK HALF FULL.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 7663436 ↗MalfunctionJun 29, 2018Erratic or Intermittent Display; Erratic or Intermittent Display
IT WAS REPORTED THAT DURING SAMPLE EVALUATION IT WAS FOUND THAT THE THE LCD WAS NOT DISPLAYING ALL SEGMENTS IN THE CONTAINER VOLUME/FLOW RATE QUADRANT.
Patient problems: No Patient Involvement
MAUDE report 7652920 ↗MalfunctionJun 28, 2018Inaccurate Flow Rate; No Device Output; Data Problem; Inaccurate Flow Rate
IT WAS REPORTED THAT THE DEVICE WOULD NOT RECORD URINE OUTPUT.
Patient problems: No Known Impact Or Consequence To Patient; No Patient Involvement
MAUDE report 7648160 ↗MalfunctionMay 10, 2018Volume Accuracy Problem; Incorrect, Inadequate or Imprecise Result or Readings; Volume Accuracy Problem
IT WAS REPORTED BY THE CLINICAL MANAGER THAT THE DEVICE WAS NOT ACCURATELY READING THE URINE OUTPUT.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 7502652 ↗MalfunctionApr 24, 2018Volume Accuracy Problem; Incorrect Measurement; Volume Accuracy Problem
IT WAS REPORTED THAT THE DEVICE WAS ON THE PATIENT AND NOT MEASURING URINE CORRECTLY.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 7453655 ↗MalfunctionApr 23, 2018Failure to Power Up; Application Interface Becomes Non-Functional Or Program Exits Abnormally; Failure to Power Up
IT WAS REPORTED THAT THE SCREEN FROZE UPON BOOT UP.
Patient problems: No Known Impact Or Consequence To Patient; No Patient Involvement
MAUDE report 7453098 ↗MalfunctionMar 19, 2018Volume Accuracy Problem; Volume Accuracy Problem; Device Operates Differently Than Expected
IT WAS REPORTED THAT THE DEVICE WAS NOT READING URINE VOLUME ACCURATELY.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 7349208 ↗
Other URINOMETER, ELECTRICAL devices
- Sensica UO System SCCS1002C. R. Bard, Inc. · URINOMETER, ELECTRICAL2 recalls289 reports
- Sensica UO System SCCS1002EC. R. Bard, Inc. · URINOMETER, ELECTRICAL42 reports
- Sensica UO SCCS-1001Adaptec Medical Devices LLC · Urinometer, Electrical7 reports
- Sensica UO System SCCS1001C. R. Bard, Inc. · URINOMETER, ELECTRICAL6 reports
- Sensica UO System SSPK1025C. R. Bard, Inc. · URINOMETER, ELECTRICAL4 reports
- Sensica UO System SBTC1005C. R. Bard, Inc. · URINOMETER, ELECTRICAL4 reports
- Sensica UO System SBTS1001C. R. Bard, Inc. · URINOMETER, ELECTRICAL3 reports
- Bard® Uros™ Precision Urine Output and Temperature Monitor BK10001MC. R. Bard, Inc. · URINOMETER, ELECTRICAL2 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page