Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00801741215742), received between Feb 17, 2022 and Mar 7, 2022 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Device Sensing Problem” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Device Sensing Problem4

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 7, 2022Device Sensing Problem; Device Sensing Problem

    IT WAS REPORTED THAT THE TEMPERATURE WAS REPORTED TO BE READING HIGH ON THE SENSICA DEVICE. LATER, IT WAS RECORDED TO ONLY BE DISPLAYING ¿***¿ IN PLACE OF THE TEMPERATURE VALUE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 13684144 ↗
  • MalfunctionMar 4, 2022Device Sensing Problem; Device Sensing Problem

    IT WAS REPORTED THAT THE TEMPERATURE VALUE WAS SHOWING ASTERISKS IN THE SENSICA UNIT, AND THE PREVIOUS HOUR READINGS WERE LOW. IT WAS STATED THAT WHEN THE USER WAS SQUEEZING OR BENDING THE CABLE AT MOLEX CONNECTOR MOLDING ON THE TEMPERATURE CABLE BY SPINNING IT AROUND, THE TEMPERATURE RECORDED ON THE SENSICA WOULD FLUCTUATE BY SEVERAL DEGREES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13667516 ↗
  • MalfunctionFeb 17, 2022Device Sensing Problem; Device Sensing Problem

    IT WAS REPORTED THAT THE SENSICA UNIT WAS DISPLAYING ASTERISKS FOR TEMPERATURE OUTPUT RATHER THAN AN ACTUAL TEMPERATURE VALUE. IT WAS MENTIONED THAT THIS WAS CAUSED BY THE TEMPERATURE CABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13548571 ↗
  • MalfunctionFeb 17, 2022Device Sensing Problem; Device Sensing Problem

    IT WAS REPORTED THAT THE SENSICA SYSTEM WAS DISPLAYING ASTERISKS FOR THE TEMPERATURE OUTPUT VALUE. IT WAS MENTIONED THAT THIS WAS CAUSED BY THE TEMPERATURE CABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13548570 ↗

Other URINOMETER, ELECTRICAL devices

All URINOMETER, ELECTRICAL devices →