Statlock IV IV0564Recalls & reported problems

Bard Access Systems, Inc. · StatLock IV Select macro 6in SmartSite Y and valve 25 box · Class 2

All models of Statlock IV (9)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
19
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 19 adverse event reports that list this exact device (DI 00801741082047), received between Apr 26, 2018 and Jun 3, 2026 — classified as 19 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (8 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fluid/Blood Leak8
  • Leak/Splash4
  • Crack3
  • Detachment of Device or Device Component3

Patient problems

  • No Consequences Or Impact To Patient8
  • No Clinical Signs, Symptoms or Conditions5
  • No Known Impact Or Consequence To Patient4
  • Hemorrhage/Blood Loss/Bleeding1
  • No Information1

Recent adverse event reports

The 19 most recent of 19 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 3, 2026Detachment of Device or Device Component

    IT WAS REPORTED BY CUSTOMER THAT THE TUBING DISCONNECTED FROM THE IV HUB. THE WHITE TUBING BECAME DETACHED, CAUSING BLOOD AND IV FLUIDS TO SPILL OVER THE PATIENT AND REPORTEDLY ONTO THE NURSE AS WELL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25382747 ↗
  • MalfunctionApr 22, 2026Fluid/Blood Leak

    IT WAS REPORTED VIA MEDWATCH THAT BD STATLOCK CATHETER STABILIZATION DEVICE IV SELECT - TUBING BURST OPEN WHILE BEING FLUSHED. THE DEFECT OCCURRED IN THE BODY OF THE TUBING, NOT AT ANY CONNECTION POINTS. NO FURTHER DETAILS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24961284 ↗
  • MalfunctionMar 3, 2026

    BD STATLOCK CATHETER STABILIZATION DEVICE IV SELECT - TUBING BURST OPEN WHILE BEING FLUSHED. THE DEFECT OCCURRED IN THE BODY OF THE TUBING--NOT AT ANY CONNECTION POINTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24501010 ↗
  • MalfunctionFeb 2, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED BY THE CUSTOMER DISCONNECTION OF THE LUER LOCK FROM THE EXTENSION TUBING FROM STATLOCK PRODUCT. NO OTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18634694 ↗
  • MalfunctionNov 9, 2022Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED BY THE CUSTOMER VIA MEDWATCH "FOUND THAT THE CLAMP ON THE EXTENSION SET HAD MADE A CUT IN THE EXTENSION SET TUBING WHICH CAUSED THE IV TO BACK UP AND LEAK. RN REPLACED THE EXTENSION SET. THIS WAS A STATLOCK CATHETER STABILIZATION. WAS THERE ANY PATIENT HARM REPORTED? IF YES, PLEASE DESCRIBE THE HARM, HOW IT WAS TREATED AND THE OUTCOME. NO HARM REPORTED. BLOOD SPILL OCCURRED AND THE RN HAD TO REPLACE THE FAULTY EXTENSION TUBING. WHAT THEY¿RE BEING TREATED FOR: SURGERY FOR HYSTERECTOMY. *DOES THE PATIENT HAVE AN INFECTIOUS DISEASE? NO.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 15763842 ↗
  • MalfunctionNov 1, 2022Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED BY MEDWATCH "IV TUBING SEPARATED FROM LUER LOCK HUB THAT WAS ATTACHED TO A PATIENT'S IV CATHETER. CLEAN EDGES, TUBING HAD NOT BEEN CUT. PATIENT'S IV CLOTTED, SO NO BACKFLOW OF BLOOD OR BLOOD LOSS FROM PATIENT." 10/24/22 ADDITIONAL INFORMATION PROVIDED: EVENT INFORMATION ¿ DATE OF OCCURRENCE: (B)(6) 2022 ¿ PLACEMENT DATE: (B)(6) 2022 ¿ EXPLANT DATE: 1(B)(6) 2022 PATIENT INFORMATION (IF AVAILABLE) ¿ WAS THERE ANY PATIENT HARM REPORTED? IF YES, PLEASE DESCRIBE THE HARM, HOW IT WAS TREATED AND THE OUTCOME. NONE REPORTED ¿ AGE: 64 ¿ SEX: MALE ¿ WEIGHT: 210LB/ 95.3KG ¿ WHAT THEY¿RE BEING TREATED FOR: SURGERY/ARTHOPLASTY OF KNEE ¿ OTHER RELEVANT HISTORY: NONE APPLICABLE ¿ CONCOMITANT THERAPY: INTERMITTENT SALINE FLUSH OF LINE AS PER MD ORDER; NO CONTINUOUS INFUSIONS OR OTHER INTERMITTENT INFUSIONS AT TIME OF DISCOVERY ¿ *DOES THE PATIENT HAVE AN INFECTIOUS DISEASE?: NONE REPORTED

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15712340 ↗
  • MalfunctionApr 9, 2020Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THE NURSE EXPERIENCED A LEAKING IV CATHETER TUBING IN THE MIDDLE OF CARING FOR A CRITICALLY ILL ADULT. IT IS STATED THERE APPEARED TO BE A LEAK FROM A HOLE IN THE TUBING IN THE LOCATION WHERE THE CLAMP HAD BEEN ON THE TUBING. THIS REPORT ADDRESSES THE FOURTH OCCURRENCE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9945256 ↗
  • MalfunctionApr 9, 2020Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THE NURSE EXPERIENCED A LEAKING IV CATHETER TUBING IN THE MIDDLE OF CARING FOR A CRITICALLY ILL ADULT. IT IS STATED THERE APPEARED TO BE A LEAK FROM A HOLE IN THE TUBING IN THE LOCATION WHERE THE CLAMP HAD BEEN ON THE TUBING. THIS REPORT ADDRESSES THE SECOND OCCURRENCE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9945251 ↗
  • MalfunctionApr 9, 2020Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THE NURSE EXPERIENCED A LEAKING IV CATHETER TUBING IN THE MIDDLE OF CARING FOR A CRITICALLY ILL ADULT. IT IS STATED THERE APPEARED TO BE A LEAK FROM A HOLE IN THE TUBING IN THE LOCATION WHERE THE CLAMP HAD BEEN ON THE TUBING. THIS REPORT ADDRESSES THE THIRD OCCURRENCE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9945247 ↗
  • MalfunctionApr 2, 2020Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THE NURSE EXPERIENCED A LEAKING IV CATHETER TUBING IN THE MIDDLE OF CARING FOR A CRITICALLY ILL ADULT. IT IS STATED THERE APPEARED TO BE A LEAK FROM A HOLE IN THE TUBING IN THE LOCATION WHERE THE CLAMP HAD BEEN ON THE TUBING.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9916786 ↗
  • MalfunctionMar 19, 2020Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED DURING SHIFT REPORT ONCOMING RN NOTICED THAT IV CONNECTION PORT WAS BROKEN OFF OF PATIENTS Y TUBING. PER OFF-GOING RN IT HAD BEEN THAT WAY AT THE START OF HER SHIFT. PER CHARTING THE LINE WAS FLUSHED AT 0059 BUT NO IV MEDICATIONS WERE GIVEN THAT SHIFT. IV WAS REMOVED AND INFECTION PREVENTION WAS NOTIFIED. NO OTHER INFORMATION WAS PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9856812 ↗
  • MalfunctionMar 18, 2020Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT AN EXPERIENCED RN PLACED IV, DREW BLOOD FOR LABS, AND WHEN SHE WENT TO FLUSH THE INT, BLOOD SPRAYED ALL OVER HER FROM WHAT SHE DESCRIBED AS "A TINY HOLE".

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9847382 ↗
  • MalfunctionSep 13, 2019Crack; Crack

    IT WAS REPORTED THAT THE IV TUBING CRACKED AT THE HUB. NO OTHER INFORMATION WAS PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9035833 ↗
  • MalfunctionSep 12, 2019Crack; Crack

    IT WAS REPORTED THAT THE IV TUBING CRACKED AT THE HUB ON TWO DEVICES. NO OTHER INFORMATION WAS PROVIDED. THIS REPORT ADDRESSES THE FIRST CRACKED DEVICE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9010532 ↗
  • MalfunctionSep 12, 2019Crack

    IT WAS REPORTED THAT THE IV TUBING CRACKED AT THE HUB ON TWO DEVICES. NO OTHER INFORMATION WAS PROVIDED. THIS REPORT ADDRESSES THE SECOND CRACKED DEVICE.

    Patient problems: No Information

    MAUDE report 9010528 ↗
  • MalfunctionApr 26, 2018Leak/Splash

    IT WAS REPORTED BY THE FACILITY THAT THE DEVICE WAS FOUND TO BE LEAKING AROUND THE LUER LOCK. THIS REPORT ADDRESS THE THIRD OF FOUR DEVICES FOUND TO BE LEAKING.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7466875 ↗
  • MalfunctionApr 26, 2018Leak/Splash

    IT WAS REPORTED BY THE FACILITY THAT THE DEVICE WAS FOUND TO BE LEAKING AROUND THE LUER LOCK. THIS REPORT ADDRESS THE FOURTH OF FOUR DEVICES FOUND TO BE LEAKING.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7466870 ↗
  • MalfunctionApr 26, 2018Leak/Splash

    IT WAS REPORTED BY THE FACILITY THAT THE DEVICE WAS FOUND TO BE LEAKING AROUND THE LUER LOCK. THIS REPORT ADDRESS THE SECOND OF FOUR DEVICES FOUND TO BE LEAKING.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7466867 ↗
  • MalfunctionApr 26, 2018Leak/Splash

    IT WAS REPORTED BY THE FACILITY THAT THE DEVICE WAS FOUND TO BE LEAKING AROUND THE LUER LOCK. THIS REPORT ADDRESS THE FIRST OF FOUR DEVICES FOUND TO BE LEAKING.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7466863 ↗

Other Set, administration, intravascular devices

All Set, administration, intravascular devices →