Skip to content

MAX-CORE MC1825Recalls & reported problems

Bard Peripheral Vascular, Inc. · MAX-CORE® Disposable Core Biopsy Instrument, 18g x 25cm · Class 2

Quick summary

This Bard MAX-CORE disposable biopsy instrument is used to take tissue samples. Many of the 290 reports describe the instrument not firing, firing on its own, or failing to collect samples. Reports rose from 28 in the prior 12 months to 62 in the last 12 months. The reports include no deaths and 8 injuries. Its single recall started in October 2019, is now closed, and was about a rise in complaints of firing failures.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
284
Deaths reported
0
Injuries reported
4
Recalls naming it
1

Summary

The FDA's MAUDE database contains 284 adverse event reports that list this exact device (DI 00801741084492), received between Nov 4, 2016 and Sep 30, 2026 — classified as 4 injuries, 280 malfunctions.

The most frequently reported specific device problem is “Self-Activation or Keying” (126 reports).

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 3 reportsDec 2023: 2 reportsJan 2024: 0 reportsFeb 2024: 3 reportsMar 2024: 7 reportsApr 2024: 5 reportsMay 2024: 5 reportsJun 2024: 4 reportsJul 2024: 2 reportsAug 2024: 5 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 7 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 5 reportsSep 2025: 12 reportsOct 2025: 2 reportsNov 2025: 0 reportsDec 2025: 3 reportsJan 2026: 6 reportsFeb 2026: 15 reportsMar 2026: 21 reportsApr 2026: 2 reportsMay 2026: 2 reportsJun 2026: 4 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 2 reportsOct 2026: 0 reports20242025202621
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 25, 2019Terminated

    During the firm's review of May 2019 complaint data, a significant increase in Disposable Core Biopsy Instrument complaints were observed for Failure to Fire, Firing Problem, and Failure to Obtain Samples.

    Z-0724-2020 · Root cause: Equipment maintenance · FDA record ↗

Most reported problems

Device problems

  • Self-Activation or Keying126
  • Failure to Fire109
  • Failure to Prime62
  • Failure to Obtain Sample38
  • Break21
  • Difficult to Remove18
  • Material Twisted/Bent10
  • Device Contamination with Chemical or Other Material8

Patient problems

  • No Clinical Signs, Symptoms or Conditions149
  • No Consequences Or Impact To Patient116
  • No Patient Involvement15
  • Insufficient Information2
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Bacterial Infection1
  • Injury1
  • Pain1

Recent adverse event reports

The 25 most recent of 284 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Failure to Fire; Activation Problem

    BARD MAX-CORE Disposable Core Biopsy Instrument did not fire currently when trigger was pressed, only 1/2 mechanisms activated.LOT: RELQ4188.REF: MC1825.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50944753 ↗
  • MalfunctionSep 23, 2026Failure to Fire

    On (b)(6) 2026, a patient underwent a kidney biopsy procedure using a Maxcore instrument. During the procedure, the cannula failed to retract automatically, making it impossible to obtain a tissue sample. There was no reported patient injury.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50868110 ↗
  • MalfunctionJul 29, 2026Misfire; Failure to Obtain Sample; Therapeutic or Diagnostic Output Failure

    During case biopsy gun was misfiring and failing to collect needed biopsies. Device was removed and replaced with a new one, case resumed without further complications

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50068668 ↗
  • MalfunctionJun 25, 2026Break

    ON (B)(6) 2026, A PATIENT UNDERWENT A BIOPSY PROCEDURE USING THE MAXCORE INSTRUMENT. DURING THE PROCEDURE, WHEN THE DEVICE WAS FIRED FOR TISSUE EXTRACTION, THE DEVICE TIP BROKE OFF AND THE DEVICE WAS SUBSEQUENTLY REMOVED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25617401 ↗
  • MalfunctionJun 18, 2026Failure to Fire

    ON (B)(6) 2026, A PATIENT UNDERWENT AN ULTRASOUND GUIDED KIDNEY BIOPSY USING A MAXCORE INSTRUMENT. DURING THE PROCEDURE, AFTER PRESSING THE BUTTON TO INSERT THE NEEDLE, THE INTRODUCER NEEDLE PRODUCED AN ABNORMAL SOUND AND FAILED TO PUNCTURE THE TISSUE. THE USER ATTEMPTED THE PROCEDURE THREE TIMES, BUT NO SPECIMEN COULD BE OBTAINED. THE INTRODUCER NEEDLE WAS IMMEDIATELY EXCHANGED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25547013 ↗
  • MalfunctionJun 4, 2026Break; Failure to Obtain Sample

    ON (B)(6) 2026, A PATIENT UNDERWENT AN ECHO GUIDED PROSTATE BIOPSY THROUGH NORMAL DENSITY TISSUE USING THE MAXCORE DEVICE. DURING THE PROCEDURE, THE DEVICE ALLEGEDLY MADE AN ABNORMAL SOUND OR MINIMAL TO NO SAMPLE WAS OBTAINED. THE PROCEDURE WAS COMPLETED BY ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25387028 ↗
  • MalfunctionJun 1, 2026Failure to Fire

    ON (B)(6) 2026, A PATIENT UNDERWENT MICROWAVE ABLATION OF UTERINE FIBROIDS IN THE HYSTEROSCOPIC OPERATING ROOM. DURING THE INTRAOPERATIVE BIOPSY OF THE MASS, WHEN THE BIOPSY GUN WAS ACTIVATED, ONLY ONE OF THE TWO SEQUENTIAL FIRING BUTTONS PROPERLY REBOUNDED, WHICH RESULTED IN TWO CONSECUTIVE FAILURES TO OBTAIN ADEQUATE PATHOLOGICAL TISSUE SAMPLES, NECESSITATING MULTIPLE ADDITIONAL PUNCTURE BIOPSY ATTEMPTS FOR THE PATIENT TO ACHIEVE A SUCCESSFUL SAMPLE COLLECTION. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25353105 ↗
  • MalfunctionMay 27, 2026Break

    ON (B)(6) 2026, A 74-YEAR-OD MALE PATIENT UNDERWENT A PROSTATE BIOPSY PROCEDURE USING THE MAXCORE DEVICE. DURING THE PROCEDURE, THE NEEDLE WAS ALLEGEDLY BROKEN. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25296301 ↗
  • MalfunctionMay 4, 2026

    DURING PROSTATE BIOPSY THE BARD MAX CORE BIOPSY INSTRUMENT WOULD NOT ENGAGE THE BIOPSY NEEDLE TO RETRIEVE PROSTATE TISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25067531 ↗
  • MalfunctionApr 28, 2026

    BARD MAX CORE DISPOSABLE CORE BIOPSY INSTRUMENT DID NOT RETRIEVE AN ADEQUATE AMOUNT OF PROSTATE TISSUE.

    Patient problems: Insufficient Information

    MAUDE report 25014740 ↗
  • MalfunctionApr 7, 2026Failure to Fire

    ON (B)(6) 2026, A PATIENT UNDERWENT A SONOGRAPHIC GUIDED KIDNEY BIOPSY PROCEDURE BY USING MAXCORE INSTRUMENT. IT WAS REPORTED THAT DURING THE PROCEDURE THE DEVICE WAS ALLEGEDLY FAILED TO FIRE. THERE WAS NO PATIENT REPORTED INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24801202 ↗
  • MalfunctionMar 31, 2026

    PREPPED THE BIOPSY NEEDLE WE DISCOVERED IT WAS MISFIRING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24741575 ↗
  • MalfunctionMar 24, 2026Self-Activation or Keying

    A PATIENT UNDERWENT A BIOPSY PROCEDURE USING THE MAXCORE INSTRUMENT. DURING THE PROCEDURE, IT WAS REPORTED THAT THE DEVICE ALLEGEDLY AUTO FIRED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24679725 ↗
  • MalfunctionMar 24, 2026Self-Activation or Keying

    A PATIENT UNDERWENT A BIOPSY PROCEDURE USING THE MAXCORE INSTRUMENT. DURING THE PROCEDURE, THE DEVICE ALLEGEDLY AUTO FIRED SINCE THE MECHANISM THAT PRESS THE BUTTON, DOES NOT WORKED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24679680 ↗
  • MalfunctionMar 24, 2026Self-Activation or Keying

    A PATIENT UNDERWENT A BIOPSY PROCEDURE USING THE MAXCORE INSTRUMENT. DURING THE PROCEDURE, IT WAS REPORTED THAT THE DEVICE ALLEGEDLY AUTO FIRED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24672815 ↗
  • MalfunctionMar 17, 2026Failure to Fire

    ON (B)(6), A PATIENT UNDERWENT AN ULTRASOUND GUIDED KIDNEY BIOPSY THROUGH NORMAL DENSITY TISSUE BY USING A MAXCORE INSTRUMENT. IT WAS REPORTED THAT DURING THE PROCEDURE, THE OUTER CANNULA OF THE INSTRUMENT ALLEGEDLY COULD NOT BE FIRED. NO COAXIAL NEEDLE WAS USED. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE, AND NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24616056 ↗
  • MalfunctionMar 17, 2026Failure to Fire

    ON (B)(6) 2026, A PATIENT UNDERWENT AN ULTRASOUND GUIDED KIDNEY BIOPSY THROUGH NORMAL DENSITY TISSUE BY USING A MAXCORE INSTRUMENT. IT WAS REPORTED THAT DURING THE PROCEDURE, THE OUTER CANNULA OF THE INSTRUMENT ALLEGEDLY COULD NOT BE FIRED. NO COAXIAL NEEDLE WAS USED. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE, AND NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24616052 ↗
  • MalfunctionMar 16, 2026Break; Failure to Obtain Sample; Failure to Fire

    ON (B)(6) 2025, A PATIENT UNDERWENT AN ULTRASOUND GUIDED PROSTATE BIOPSY PROCEDURE USING MAXCORE INSTRUMENT. DURING THE PROCEDURE, THE DEVICE WAS TRIGGERED, BUT THE BIOPSY NEEDLE DID NOT TAKE A SAMPLE AND SOUNDS ALMOST LIKE THE SPRING WOULD SNAP/BREAK. THERE WAS NO PATIENT REPORTED INJURY. THIS IS REPORT 8 OF 10.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24603771 ↗
  • MalfunctionMar 16, 2026Break

    ON (B)(6) 2025, A PATIENT UNDERWENT AN ULTRASOUND GUIDED PROSTATE BIOPSY PROCEDURE USING MAXCORE INSTRUMENT. DURING THE PROCEDURE, THE DEVICE WAS TRIGGERED, BUT THE BIOPSY NEEDLE DID NOT TAKE A SAMPLE AND SOUNDS ALMOST LIKE THE SPRING WOULD SNAP/BREAK. THERE WAS NO PATIENT REPORTED INJURY. THIS IS REPORT 10 OF 10.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24603770 ↗
  • MalfunctionMar 16, 2026Break; Failure to Obtain Sample; Failure to Fire

    ON (B)(6) 2025, A PATIENT UNDERWENT AN ULTRASOUND GUIDED PROSTATE BIOPSY PROCEDURE USING MAXCORE INSTRUMENT. DURING THE PROCEDURE THE DEVICE WAS TRIGGERED, BUT THE BIOPSY NEEDLE DID NOT TAKE A SAMPLE AND SOUNDS ALMOST LIKE THE SPRING WOULD SNAP/BREAK. THERE WAS NO PATIENT REPORTED INJURY. THIS IS REPORT 7 OF 9.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24603768 ↗
  • MalfunctionMar 16, 2026Break; Failure to Obtain Sample; Failure to Fire

    ON (B)(6) 2025, A PATIENT UNDERWENT AN ULTRASOUND GUIDED PROSTATE BIOPSY PROCEDURE USING MAXCORE INSTRUMENT. DURING THE PROCEDURE, THE DEVICE WAS TRIGGERED, BUT THE BIOPSY NEEDLE DID NOT TAKE A SAMPLE AND SOUNDS ALMOST LIKE THE SPRING WOULD SNAP/BREAK. THERE WAS NO PATIENT REPORTED INJURY. THIS IS REPORT 5 OF 10.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24603767 ↗
  • MalfunctionMar 16, 2026Break; Failure to Obtain Sample; Failure to Fire

    ON (B)(6) 2025, A PATIENT UNDERWENT AN ULTRASOUND GUIDED PROSTATE BIOPSY PROCEDURE USING MAXCORE INSTRUMENT. DURING THE PROCEDURE THE DEVICE WAS TRIGGERED, BUT THE BIOPSY NEEDLE DID NOT TAKE A SAMPLE AND SOUNDS ALMOST LIKE THE SPRING WOULD SNAP/BREAK. THERE WAS NO PATIENT REPORTED INJURY. THIS IS REPORT 9 OF 9.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24603765 ↗
  • MalfunctionMar 16, 2026Break; Failure to Obtain Sample; Failure to Fire

    ON (B)(6) 2025, A PATIENT UNDERWENT AN ULTRASOUND GUIDED PROSTATE BIOPSY PROCEDURE USING MAXCORE INSTRUMENT. DURING THE PROCEDURE, THE DEVICE WAS TRIGGERED, BUT THE BIOPSY NEEDLE DID NOT TAKE A SAMPLE AND SOUNDS ALMOST LIKE THE SPRING WOULD SNAP/BREAK. THERE WAS NO PATIENT REPORTED INJURY. THIS IS REPORT 6 OF 10.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24603764 ↗
  • MalfunctionMar 16, 2026Break; Failure to Obtain Sample; Failure to Fire

    ON (B)(6) 2025, A PATIENT UNDERWENT AN ULTRASOUND GUIDED PROSTATE BIOPSY PROCEDURE USING MAXCORE INSTRUMENT. DURING THE PROCEDURE, THE DEVICE WAS TRIGGERED, BUT THE BIOPSY NEEDLE DID NOT TAKE A SAMPLE AND SOUNDS ALMOST LIKE THE SPRING WOULD SNAP/BREAK. THERE WAS NO PATIENT REPORTED INJURY. THIS IS REPORT 4 OF 10.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24603763 ↗
  • MalfunctionMar 16, 2026Break; Failure to Obtain Sample; Failure to Fire

    ON (B)(6) 2025, A PATIENT UNDERWENT AN ULTRASOUND GUIDED PROSTATE BIOPSY PROCEDURE USING MAXCORE INSTRUMENT. DURING THE PROCEDURE, THE DEVICE WAS TRIGGERED, BUT THE BIOPSY NEEDLE DID NOT TAKE A SAMPLE AND SOUNDS ALMOST LIKE THE SPRING WOULD SNAP/BREAK. THERE WAS NO PATIENT REPORTED INJURY. THIS IS REPORT 3 OF 10.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24603762 ↗

Other INSTRUMENT, BIOPSY devices

All INSTRUMENT, BIOPSY devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.