ENCOR® ECP0110GRecalls & reported problems
Senorx, Inc. · ENCOR® Probe Stereotactic/Ultrasound 10G · Class 2
All models of ENCOR (17)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This SenoRx Encor breast biopsy probe has 392 reports in the FDA database, with 8 injuries and no deaths. The reports often describe weak or lost vacuum suction, failure to collect a tissue sample, leaks, and contamination with other material. Some reports describe bleeding or a needle that could not be removed. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 390
- Deaths reported
- 0
- Injuries reported
- 8
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 390 adverse event reports that list this exact device (DI 00801741086243), received between Jan 13, 2017 and Jun 3, 2026 — classified as 8 injuries, 382 malfunctions.
The most frequently reported specific device problem is “Suction Problem” (236 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Suction Problem236
- Failure to Obtain Sample105
- Leak/Splash47
- Device Contamination with Chemical or Other Material45
- Filling Problem27
- Difficult to Remove9
- Failure to Cycle6
- Adverse Event Without Identified Device or Use Problem4
Patient problems
- No Consequences Or Impact To Patient259
- No Patient Involvement73
- No Clinical Signs, Symptoms or Conditions50
- Hematoma5
- Pain4
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Needle Stick/Puncture1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 25 most recent of 390 reports, in the reporter's own words as published by the FDA.
InjuryJun 3, 2026Suction Problem; Device-Device Incompatibility
ON (B)(6) 2026, A PATIENT UNDERWENT A STEREOTACTIC GUIDED VACUUM ASSISTED BREAST BIOPSY THROUGH NORMAL DENSITY TISSUE USING THE ENCOR PROBE. DURING THE PROCEDURE, THE PROBE SHOWED A VACUUM ERROR EVEN THOUGH ALL CONSUMABLES WERE TIGHTLY SEALED, AND A CLICKING SOUND WAS HEARD FROM THE ENCOR DRIVER. ADDITIONALLY, IT WAS REPORTED THAT THE PATIENT DEVELOPED A HEMATOMA, AND THE BLEEDING COULD NOT BE STOPPED. THE CURRENT STATUS OF THE PATIENT IS UNKNOWN.
Patient problems: Hematoma
MAUDE report 25369790 ↗MalfunctionSep 23, 2025Failure to Cycle; Difficult to Remove; Suction Problem; Failure to Cycle; Difficult to Remove; Suction Problem
ON (B)(6) 2025, A PATIENT UNDERWENT A STEREOTACTIC GUIDED VACUUM ASSISTED BREAST BIOPSY PROCEDURE USING THE ENCOR PROBE. DURING THE PROCEDURE, THE DEVICE JAMMED WITH THE NEEDLE STILL INSIDE THE PATIENT. ALTHOUGH THE EQUIPMENT APPEARED TO BE PERFORMING THE VACUUM FUNCTION, IT WAS NOT. WHEN THE PEDALS WERE PRESSED, THE EQUIPMENT DID NOT RESPOND, AND THE NEEDLE COULD NOT BE REMOVED FROM THE PATIENT. IT WAS FURTHER REPORTED THAT THE NEEDLE FAILED TO CLOSE DUE TO A VACUUM ISSUE. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23119922 ↗InjurySep 2, 2025Failure to Cycle; Device-Device Incompatibility; Failure to Cycle; Device-Device Incompatibility
ON (B)(6) 2025, A PATIENT UNDERWENT A BREAST BIOPSY PROCEDURE WITH THE ENCOR PROBE AND MARKER PLACEMENT WITH A SENORMARK ULTRA. DURING THE PROCEDURE, A CLICK SOUND WAS HEARD DURING VACUUMING AND THE PATIENT FELT PAIN. THE DEVICE WAS NOT FULLY CLOSED, AND AN OBSTRUCTION WAS FELT. THE MARKER WAS SUCCESSFULLY DEPLOYED. HOWEVER, THE PROVED WAS NOT FULLY CLOSED AFTER REMOVAL. IT WAS FURTHER REPORTED THAT THE PATIENT WAS EXPERIENCING SIGNIFICANT BLEEDING, AND A LARGE HEMATOMA WAS PRESENT. HOWEVER, THE CURRENT STATUS OF THE PATIENT IS UNKNOWN.
Patient problems: Hematoma
MAUDE report 22935671 ↗MalfunctionJul 15, 2025Missing Information; Missing Information
ON (B)(6) 2025, A PATIENT UNDERWENT A BIOPSY PROCEDURE USING ENCOR PROBE. PRIOR TO THE PROCEDURE, IT WAS NOTICED THAT BOTH THE EXTERNAL AND INTERNAL PACKAGES WERE MISSING THE MANUFACTURING DATE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22504543 ↗MalfunctionJun 13, 2025Device Markings/Labelling Problem; Device Markings/Labelling Problem
ON (B)(6) 2024, A PATIENT UNDERWENT A BIOPSY PROCEDURE USING THE ENCOR BIO PROBE. DURING THE PREPARATION OF A PROCEDURE, CUSTOMER RECEIVED THE MATERIAL WITH THE WRONG LABEL SO PROCEDURE CAN'T BE DONE. THERE WAS NO REPORTED PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22201186 ↗MalfunctionJan 6, 2025Dull, Blunt; Device Contamination with Chemical or Other Material; Dull, Blunt; Device Contamination with Chemical or Other Material
IT WAS REPORTED THAT PRIOR TO A STEREOTACTIC GUIDED VACUUM-ASSISTED BREAST BIOPSY THROUGH NORMAL TISSUE DENSITY, THE DEVICE ALLEGEDLY HAD FUZZ BALL OF SOME SORT OF PLASTIC. NO COAXIAL WAS USED AND THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21076926 ↗MalfunctionDec 1, 2024Break; Improper or Incorrect Procedure or Method; Suction Problem; Failure to Obtain Sample; Break; Improper or Incorrect Procedure or Method; Suction Problem; Failure to Obtain Sample
IT WAS REPORTED THAT DURING A BREAST BIOPSY THROUGH SOFT DENSITY TISSUE, THE SUCTION POWER OF THE DEVICE ALLEGEDLY HAD INSUFFICIENT VACUUM. IT WAS FURTHER REPORTED THAT THE TISSUE ALLEGEDLY COULD NOT BE REMOVED. THE PROCEDURE WAS COMPLETED USING ANOTHER METHOD. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20815952 ↗MalfunctionJan 16, 2024Difficult to Remove; Material Twisted/Bent; Difficult to Remove; Material Twisted/Bent
IT WAS REPORTED THAT DURING A BIOPSY PROCEDURE, THE DEVICE ALLEGEDLY TURNED OUT TO BE COMPLETELY CROOKED. IT WAS FURTHER REPORTED THAT THE NEEDLE WAS ALLEGEDLY DIFFICULT TO REMOVE FROM THE PATIENT. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18519976 ↗MalfunctionJul 13, 2023Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material
IT WAS REPORTED THAT PRIOR TO A BIOPSY PROCEDURE, JELLY-LIKE MATERIAL ALLEGEDLY ADHERED NEAR THE OPENING GATE OF THE NEEDLE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17314942 ↗MalfunctionMar 24, 2023Break; Break
IT WAS REPORTED THAT DURING A STEREOTACTIC GUIDED BREAST BIOPSY PROCEDURE, THE DEVICE WAS ALLEGEDLY BROKEN. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16608638 ↗MalfunctionMar 24, 2023Device-Device Incompatibility; Device Contamination with Chemical or Other Material; Device-Device Incompatibility; Device Contamination with Chemical or Other Material
IT WAS REPORTED THAT DURING A STEREOTACTIC BREAST BIOPSY PROCEDURE THROUGH NORMAL DENSITY TISSUE, THE MARKER WOULD NOT FIT THROUGH THE PROBE. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16608156 ↗InjuryFeb 15, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING A BREAST BIOPSY, THE PATIENT ALLEGEDLY EXPERIENCED HEMATOMA AND PAIN. REPORTEDLY LOCAL ANESTHETIC AND VACUUM WERE USED TO REDUCE THE HEMATOMA, AND PAIN KILLERS WERE PROVIDED. THE CURRENT PATIENT STATUS WAS UNKNOWN.
Patient problems: Hematoma; Pain
MAUDE report 16377656 ↗InjuryFeb 15, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING A BREAST BIOPSY, THE PATIENT ALLEGEDLY EXPERIENCED HEMATOMA AND PAIN. REPORTEDLY LOCAL ANESTHETIC AND VACUUM WERE USED TO REDUCE THE HEMATOMA, AND PAIN KILLERS WERE PROVIDED. THE CURRENT PATIENT STATUS WAS UNKNOWN.
Patient problems: Hematoma; Pain
MAUDE report 16377629 ↗MalfunctionFeb 8, 2023Failure to Cycle; Suction Problem; Failure to Cycle; Suction Problem
IT WAS REPORTED THAT DURING AN ULTRASOUND GUIDED BREAST BIOPSY PROCEDURE, THE DEVICE ALLEGEDLY HAD SUCTION ISSUE. IT WAS FURTHER REPORTED THAT DEVICE ALLEGEDLY HAD DIFFICULTIES WITH CLOSING THE NEEDLE APERTURE. REPORTEDLY THE APERTURE WAS OPEN WHICH RESULTED IN THE PATIENT HAVING A HEMATOMA. HEMATOMA WAS TREATED WITH COMPRESSION UNTIL NO MORE BLEEDING WAS OBSERVED AND ALSO APPLIED AN ICE PACK TO THE AREA. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16325996 ↗InjuryJan 27, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING BIOPSY PROCEDURE, THE PATIENT ALLEGEDLY EXPERIENCED SEVERE PAIN IN SHOULDER AND SPINE REGION. IT WAS FURTHER REPORTED THAT ANTI- INFLAMMATORY DRUGS WERE GIVEN AND IN ONE OF THEM, OPIOIDS WERE USED. THE CURRENT STATUS OF THE PATIENT WAS UNKNOWN.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Pain
MAUDE report 16251820 ↗MalfunctionJul 19, 2022Difficult to Remove; Material Deformation; Material Twisted/Bent; Difficult to Remove; Material Deformation; Material Twisted/Bent
IT WAS REPORTED THAT DURING A BREAST BIOPSY PROCEDURE, NEEDLE ALLEGEDLY BENT INSIDE THE PATIENT'S BREAST. IT WAS FURTHER REPORTED THAT NEEDLE ALLEGEDLY DIFFICULT TO REMOVE FROM THE PATIENT. REPORTEDLY THE PROCEDURE WAS COMPLETED AND THE NEEDLE WAS REMOVED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15050987 ↗MalfunctionJul 5, 2022Difficult to Remove; Material Twisted/Bent; Difficult to Remove; Material Twisted/Bent
IT WAS REPORTED THAT DURING AN ULTRASOUND GUIDED BREAST BIOPSY THROUGH FATTY DENSITY TISSUE, THE NEEDLE WAS ALLEGEDLY BENT. IT WAS FURTHER REPORTED THE DEVICE WAS ALLEGEDLY DIFFICULT TO REMOVE FROM THE PATIENT. THE PROCEDURE WAS COMPLETED BY USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14913369 ↗MalfunctionApr 19, 2022Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material
IT WAS REPORTED THAT PRIOR TO A BIOPSY PROCEDURE, THE FOREIGN MATERIAL WAS PRESENT INSIDE THE PACKAGE. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14139743 ↗MalfunctionNov 16, 2021Material Fragmentation; Device Contamination with Chemical or Other Material; Material Fragmentation; Device Contamination with Chemical or Other Material
IT WAS REPORTED THAT PRIOR TO A BREAST BIOPSY PROCEDURE, THE DEVICE ALLEGEDLY HAD BLACK VISCOUS SUBSTANCE ON THE NEEDLE TIP. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12825210 ↗MalfunctionSep 27, 2021Device Contamination with Chemical or Other Material; Noise, Audible; Device Contamination with Chemical or Other Material; Noise, Audible
IT WAS REPORTED THAT PRIOR TO A BREAST BIOPSY PROCEDURE, THE DEVICE NEEDLE ALLEGEDLY HAD FOREIGN MATERIAL. IT WAS FURTHER REPORTED THAT NO ABNORMAL NOISES FROM THE PROBE WERE HEARD. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12532461 ↗MalfunctionJul 21, 2021Positioning Failure; Difficult to Remove; Positioning Failure; Difficult to Remove
IT WAS REPORTED THAT DURING A BIOPSY PROCEDURE, THE NEEDLE ALLEGEDLY STUCK. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12201027 ↗MalfunctionJul 20, 2021Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material
IT WAS REPORTED THAT DURING A BIOPSY PROCEDURE, A PIECE OF PLASTIC WAS ALLEGEDLY PROTRUDING FROM THE BIOPSY CHAMBER. IT WAS FURTHER REPORTED THAT THE MARKER COULD NOT REMOVE FROM THE BIOPSY PROBE AND IN THE END. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12194519 ↗MalfunctionJun 8, 2021Self-Activation or Keying; Improper or Incorrect Procedure or Method; Self-Activation or Keying; Improper or Incorrect Procedure or Method
IT WAS REPORTED THAT DURING A BREAST BIOPSY PROCEDURE, THE DEVICE NOT USED THE PROPER WAY DUE LACK OF GOOD CLINICAL PRACTICES. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11956197 ↗MalfunctionApr 21, 2021Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material
IT WAS REPORTED THAT PRIOR TO A BIOPSY, THE DEVICE ALLEGEDLY CONTAMINATED. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11698620 ↗MalfunctionApr 5, 2021Break; Break
IT WAS REPORTED THAT PRIOR TO A BREAST BIOPSY PROCEDURE, THE SAMPLE BOX ALLEGEDLY BROKE WHEN THE USER PLACE THE NEEDLE TO BEGIN THE PROCEDURE. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11612821 ↗
FDA clearance and classification
The FDA regulates product code KNW (instrument, biopsy) under 21 CFR 876.1075 as a Class II device. 21 CFR 876.1075 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the ENCOR® ECP0110G
Has the ENCOR® ECP0110G been recalled?
No FDA recall record in our data names this exact device identifier (DI 00801741086243), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the ENCOR® ECP0110G?
The FDA's MAUDE database lists 390 adverse event reports for this device, received between Jan 13, 2017 and Jun 3, 2026. In those reports, the most common specific device problems are “Suction Problem” (236), “Failure to Obtain Sample” (105) and “Leak/Splash” (47); the most common patient problems are “Hematoma” (5), “Pain” (4) and “Needle Stick/Puncture” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the ENCOR® ECP0110G?
None of the reports are classified as deaths, 8 as injuries and 382 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the ENCOR® ECP0110G, and is it still sold?
The FDA's device database lists Senorx, Inc. as the company responsible for it (brand ENCOR®, model ECP0110G). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other INSTRUMENT, BIOPSY devices
- MAX-CORE MC1616Bard Peripheral Vascular, Inc. · INSTRUMENT, BIOPSY1 recall4,403 reports
- ENCOR® ECP017GSenorx, Inc. · INSTRUMENT, BIOPSY1 recall4,186 reports
- MAX-CORE MC1820Bard Peripheral Vascular, Inc. · INSTRUMENT, BIOPSY1 recall2,405 reports
- MAX-CORE MC1816Bard Peripheral Vascular, Inc. · INSTRUMENT, BIOPSY1 recall1 death853 reports
- MAX-CORE MC1410Bard Peripheral Vascular, Inc. · INSTRUMENT, BIOPSY1 recall402 reports
- MAX-CORE MC1610Bard Peripheral Vascular, Inc. · INSTRUMENT, BIOPSY1 recall394 reports
- ENCOR® ECP0110GVSenorx, Inc. · INSTRUMENT, BIOPSY392 reports
- MAX-CORE MC1825Bard Peripheral Vascular, Inc. · INSTRUMENT, BIOPSY1 recall286 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page