Skip to content

Port Access Kit PA-0029YNRecalls & reported problems

Bard Access Systems, Inc. · Port Access Kit SafeStep 22G x 0.75 inch with y-site · Class 2

All models of Port Access Kit (49)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00801741085499), received between Jun 4, 2020 and Aug 26, 2020 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (4 reports).

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Jun 19, 2020Class IITerminated

    Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that have developed breaks or cracks can cause leakage resulting in exposure to chemicals such a chemotherapeutics or biohazard material (e.g., blood). A crack or break could result in potential blood loss, catheter occlusion, air embolism, under or interrupted infusion, or site contamination which could lead to infection.

    Z-2666-2020 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Fluid/Blood Leak4

Patient problems

  • No Consequences Or Impact To Patient4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 26, 2020Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED TWO BARD PORT ACCESS KITS WERE FOUND LEAKING AT THE CLAVE AND HAD TO RE-ACCESS A PATIENT. THIS REPORT ADDRESSES THE SECOND DEVICE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10457517 ↗
  • MalfunctionAug 26, 2020Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED TWO BARD PORT ACCESS KITS WERE FOUND LEAKING AT THE CLAVE AND HAD TO RE-ACCESS A PATIENT. THIS REPORT ADDRESSES THE FIRST DEVICE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10457516 ↗
  • MalfunctionJun 4, 2020Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED PORT NEEDLE DEVICE INTACT UPON PRE-FLUSHING, PATIENT PORT-A-CATH ACCESSED WITH DEVICE, ON ASPIRATION FOR BLOOD RETURN, NOTED BLOOD DROPS FORMING AT POINT WHERE TUBING CONNECTS TO PLASTIC CLEAR PORT NEEDLE DEVICE. EQUIPMENT FAILURE DUE TO HOLE/ IMPAIRMENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10121448 ↗
  • MalfunctionJun 4, 2020Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED PT ACCESSED WITH PORT NEEDLE. LINE FLUSHED AND THEN ASPIRATED AND THAT¿S WHEN BLOOD STARTED DRIPPING FROM WHERE THE LINE/ TUBE IS CONNECTED TO THE NEEDLE. ON (B)(6) 2020: THE IMPAIRMENT WAS AT THE EXACT POINT WHERE THE TUBING CONNECTS TO THE PLASTIC PORT NEEDLE DEVICE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10121440 ↗

Other Set, administration, intravascular devices