Skip to content

Port Access Kit PA-0032Recalls & reported problems

Bard Access Systems, Inc. · Port Access Kit SafeStep 20G x 1 inch · Class 2

All models of Port Access Kit (49)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00801741085543), received between Mar 17, 2023 and Mar 17, 2023 — classified as 1 malfunction.

The most frequently reported specific device problem is “Fluid/Blood Leak” (1 report).

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Jun 19, 2020Class IITerminated

    Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that have developed breaks or cracks can cause leakage resulting in exposure to chemicals such a chemotherapeutics or biohazard material (e.g., blood). A crack or break could result in potential blood loss, catheter occlusion, air embolism, under or interrupted infusion, or site contamination which could lead to infection.

    Z-2666-2020 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Fluid/Blood Leak1

Patient problems

  • Anxiety1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 17, 2023Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED BY THE CUSTOMER THAT WHEN ACCESSING THE PORT AND DRAWING BACK BLOOD UPON ACCESS AND THEN PULSATILE FLUSH BACK AND GET SALINE MIXED WITH THE BLOOD COMING FROM THE HUB, LEAKED FORCEFULLY, AND IT CAME OUT FORCEFULLY AND HIT THE PATIENT IN THE NECK. IT WAS STATED THERE WAS NO INJURY BUT THE DEVICE LEAKED ON THE PATIENT AND THE NURSE.

    Patient problems: Anxiety; No Clinical Signs, Symptoms or Conditions

    MAUDE report 16565293 ↗

Other Set, administration, intravascular devices

All Set, administration, intravascular devices →

Sources and updates