Skip to content

PowerPICC SOLO2 HF Catheter 1395118QRecalls & reported problems

Bard Access Systems, Inc. · PowerPICC SOLO2 HF Catheter with Sherlock 3CG Tip Positioning System (TPS) Stylet Basic Kit, 5F Triple-Lumen 55cm · Class 2 · Implant

All models of PowerPICC SOLO2 HF Catheter (5)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00801741085727), received between Feb 10, 2017 and Jun 24, 2025 — classified as 2 injuries, 6 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fluid/Blood Leak2
  • Kinked2
  • Material Frayed2
  • Contamination /Decontamination Problem1
  • Material Deformation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • No Consequences Or Impact To Patient2
  • Atrial Fibrillation1
  • Foreign Body In Patient1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 24, 2025Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED BY THE CUSTOMER "ON (B)(6), I RECEIVED A CONSULT TO INSERT A PICC IN A CRITICAL PATIENT IN ER. SUCCESSFUL PROCEDURE, BUT WHEN COMPLETING MY FINAL FLUSHES OF THE LINE, I NOTICED NORMAL SALINE LEAKING FROM THE WHITE LUMEN AND BLOOD BACK FLOWING. UPON FURTHER INSPECTION, I DISCOVERED A SMALL CRACK IN THE LINE RIGHT AT THE HUB OF THE PICC. I IMMEDIATELY PULLED THE LINE AND PLACED A BANDAGE. PATIENT NEEDED TO GO TO ICU URGENTLY SO I FOLLOWED PATIENT THERE AND COMPLETED A SECOND SUCCESSFUL PICC ATTEMPT IN ICU USING A DIFFERENT LOT NUMBER." ADDITIONAL INFORMATION RECEIVED (B)(6) 2025: IT WAS REPORTED THE CUSTOMER "PATIENT HAD AN ADDITIONAL PERIPHERAL IV BUT IT WASN'T GREAT. IT WAS A SMALLER GAUGE AND LEAKING. SHE WAS ALSO RUNNING NOREPINEPHRINE (WHICH IS A POTENT VASOCONSTRICTOR) THAT NEEDS TO BE RUNS CENTRALLY EX) VIA A PICC LINE. NO, IT DID NOT CONTRIBUTE TO THE DECLINE IN PATIENT CONDITION. THE PICC LINE WOULD HAVE HELPED THE STAFF CARE FOR THE PATIENT EASIER BUT SHE WAS ALREADY DETERIORATING AND GOING OVER TO ICU. ICU DID HOWEVER REQUIRE THE PICC LINE AS BY THE TIME SHE GOT TRANSFERRED TO ICU HER IV LINE HAD WENT INTERSTITIAL AND THEY NEEDED TO URGENTLY INTUBATE HER, T

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22319166 ↗
  • MalfunctionJan 28, 2025Contamination /Decontamination Problem; Contamination /Decontamination Problem

    IT WAS REPORTED BY CUSTOMER, HAIR FOUND IN STERILE PICC TRAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21256962 ↗
  • MalfunctionOct 30, 2024Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED BY CUSTOMER THAT CRACKED PICC DISCOVERED WHEN FLUSHING PRIOR TO INSERTION. NO OTHER INFORMATION WAS PROVIDED. ADDITIONAL INFORMATION: DURING FOLLOW UP WITH THE CUSTOMER, THIS WAS AN INSERTER ERROR, ONE OF THE NEW NURSES CUT THE PICC AND THEN STATED THE PICC WAS FOUND CRACKED FROM THE PACKAGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20574537 ↗
  • InjuryMay 21, 2021Material Frayed; Material Frayed

    IT WAS REPORTED THAT THERE ARE STYLET ISSUES. ADDITIONAL INFORMATION RECEIVED 5/5/2021: IT WAS REPORTED ON (B)(6) THAT THE STYLET WAS BROKEN. THE BROKEN PIECE REMAINS IN THE PATIENT'S AXILLA AND HAS BEEN DETERMINED TO BE IRRETRIEVABLE. IT WAS STATED THAT THE PICC FELT RESISTANCE WHEN THREADING AND THAT THE INSERTION HAD TO BE ABORTED. THE STYLET WIRE WAS CHECKED AND IT WAS NOTED THAT 1CM OF THE STYLET TIP WAS MISSING AND WAS UNABLE TO BE FOUND IN THE STERILE FIELD. AN X-RAY WAS PERFORMED AND THE MISSING PIECE WAS FOUND IN THE PATIENT'S HEMITHORAX.

    Patient problems: Foreign Body In Patient

    MAUDE report 11866698 ↗
  • MalfunctionMay 5, 2021Material Frayed; Material Frayed

    IT WAS REPORTED PICC NURSE INSERTED PICC IN PT. AND WHEN STYLET WAS REMOVED FROM PICC IT WAS NOTED THAT STYLET TIP WAS BENT AND ALMOST BROKEN OFF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11780483 ↗
  • MalfunctionJan 2, 2020Material Deformation; Material Deformation

    IT WAS REPORTED THAT THE PICC IS KINKING ON ITSELF UNDER THE EXISTING DRESSING. NO OTHER INFORMATION WAS PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9539032 ↗
  • MalfunctionFeb 10, 2017Kinked

    NURSE REPORTED A KINK ON THE EXTERNAL PORTION OF THE PICC. THE PICC LINE WAS INSERTED ON (B)(6) 2017. THE RIGHT CEPHALIC VEIN WAS USED, AS THE BASILICA VESSEL WAS BEING USED FOR IV MEDICATION. THE PICC LINE WAS TRIMMED AT 50 CM AND HAD AN EXTERNAL LENGTH OF 1.5 CM. CHEST X-RAY WAS COMPLETED POST INSERTION AS THE PATIENT WAS IN ATRIAL FIBRILLATION. PICC REMAINS IN THE PATIENT AND HAS BEEN SECURED. NO PATIENT INJURY REPORTED.

    Patient problems: Atrial Fibrillation

    MAUDE report 6321889 ↗
  • MalfunctionFeb 10, 2017Kinked

    NURSE REPORTED A KINK ON THE EXTERNAL PORTION OF A PICC. THE PICC LINE WAS INSERTED ON (B)(6) 2017. THE RIGHT BASILIC VEIN WAS USED. THE PICC LINE WAS TRIMMED AT 50 CM AND HAD AN EXTERNAL LENGTH OF 3 CM. NO COMPLICATIONS DURING THE INSERTION AND NO MANIPULATION OF THE PICC WAS REQUIRED. THE PICC WAS INSERTED USING 3CG TECHNOLOGY AND NO CHEST X-RAY WAS DONE. THE PICC CURRENTLY REMAINS IN THE PATIENT AND HAS BEEN SECURED. NO PATIENT INJURY REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6321795 ↗

Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices

All Catheter,intravascular,therapeutic,long-term greater than 30 days devices →