Skip to content

ENCOR® ECPMR0110GRecalls & reported problems

Senorx, Inc. · ENCOR® Probe MRI 10G Trocar Tip · Class 2

Adverse event reports
9
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00801741086304), received between Oct 9, 2017 and Dec 19, 2024 — classified as 1 injury, 8 malfunctions.

The most frequently reported specific device problem is “Device Contamination with Chemical or Other Material” (4 reports).

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jan 3, 2020Terminated

    Lot numbers and products inadvertently not included in the scope of the previous recall. Original recall was conducted due to an increase in probe failures for leaks, suction issues, and failure to obtain samples.

    Z-1433-2020 · Root cause: Component change control · FDA record ↗

Most reported problems

Device problems

  • Device Contamination with Chemical or Other Material4
  • Break2
  • Detachment of Device or Device Component2
  • Failure to Obtain Sample2
  • Crack1
  • Mechanical Problem1

Patient problems

  • No Consequences Or Impact To Patient3
  • No Patient Involvement3
  • No Clinical Signs, Symptoms or Conditions2
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 19, 2024Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component

    IT WAS REPORTED THAT POST A MAMMOGRAPHY GUIDED VACUUM ASSISTED BREAST BIOPSY PROCEDURE, A PIECE OF THE NEEDLE WAS ALLEGEDLY DETACHED. IT WAS FURTHER REPORTED THAT A PIECE OF NEEDLE FOUND INSIDE THE PATIENT BODY, AND IT WAS REMOVED THROUGH SURGERY. THE CURRENT STATUS OF THE PATIENT IS STABLE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 20974684 ↗
  • MalfunctionJun 6, 2022Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THAT DURING A MRI GUIDED BREAST BIOPSY PROCEDURE THROUGH NORMAL DENSITY TISSUE, THE DEVICE ALLEGEDLY HAD MECHANICAL PROBLEM. THE PROCEDURE WAS COMPLETED BY PRESSING VACUUM BUTTON TO FORCE STOP THE SAMPLING. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14600189 ↗
  • MalfunctionAug 24, 2021Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material

    IT WAS REPORTED THAT DURING A BIOPSY PROCEDURE, THE DEVICE ALLEGEDLY HAD FOREIGN MATERIAL ON SAMPLE CUP. IT WAS FURTHER REPORTED THAT THE FOREIGN MATERIAL WAS REMOVED AND THE NEEDLE WAS USED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12357440 ↗
  • MalfunctionJul 22, 2020Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material

    THIS REPORT SUMMARIZES FOUR MALFUNCTIONS. A REVIEW OF THE REPORTED INFORMATION INDICATES THAT MODEL ECPMR0110G BIOPSY INSTRUMENT ALLEGEDLY EXPERIENCED DEVICE CONTAMINATION WITH CHEMICAL OR OTHER MATERIAL. THE INFORMATION WAS RECEIVED FROM VARIOUS SOURCES. ALL FOUR MALFUNCTIONS DID NOT INVOLVE A PATIENTS AS THERE WAS NO PATIENT CONTACT.

    Patient problems: No Patient Involvement

    MAUDE report 10308248 ↗
  • MalfunctionJan 22, 2020Failure to Obtain Sample; Failure to Obtain Sample

    IT WAS REPORTED THAT DURING AN MRI BREAST BIOPSY, THE DEVICE ALLEGEDLY FAILED TO OBTAIN SAMPLES. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9617068 ↗
  • MalfunctionJan 17, 2020Failure to Obtain Sample; Failure to Obtain Sample

    IT WAS REPORTED THAT DURING A BREAST BIOPSY, THE DEVICE ALLEGEDLY FAILED TO OBTAIN SAMPLES. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9601742 ↗
  • MalfunctionJan 14, 2020Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material

    THIS REPORT SUMMARIZES FOUR MALFUNCTIONS. A REVIEW OF THE REPORTED INFORMATION INDICATED THAT MODEL ECPMR0110G BIOPSY INSTRUMENT ALLEGEDLY EXPERIENCED FOREIGN MATERIAL PRESENT IN DEVICE. THIS INFORMATION WAS RECEIVED FROM VARIOUS SOURCES. OF THE 4 FOREIGN MATERIAL PRESENT IN DEVICES, 4 DID NOT INVOLVE PATIENTS. THERE WAS NO PROVIDED PATIENT INFORMATION.

    Patient problems: No Patient Involvement

    MAUDE report 9588475 ↗
  • MalfunctionOct 18, 2019Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material

    THIS REPORT SUMMARIZES TWO MALFUNCTIONS. A REVIEW OF THE REPORTED INFORMATION INDICATED THAT MODEL ECPMR0110G BIOPSY INSTRUMENT ALLEGEDLY HAD A FOREIGN MATERIAL PRESENT IN THE DEVICE. THIS INFORMATION WAS RECEIVED FROM VARIOUS SOURCES. NONE OF THE TWO MALFUNCTIONS INVOLVED A PATIENT. THE AGE, WEIGHT, AND GENDER OF THE INTENDED PATIENTS WAS NOT PROVIDED.

    Patient problems: No Patient Involvement

    MAUDE report 9211458 ↗
  • MalfunctionOct 9, 2017Break; Crack; Detachment of Device or Device Component; Break; Crack; Detachment of Device or Device Component

    IT WAS REPORTED THAT POST BREAST BIOPSY UNDER MRI GUIDANCE, THE PROBE TROCAR TIP WAS ALLEGEDLY DETACHED FROM THE PROBE ONCE REMOVED FROM PATIENT. IT WAS FURTHER REPORTED THE TIP WAS NOT VISIBLE IN THE PATIENT UNDER IMAGING OR IN THE DEVICE PACKAGING. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6933264 ↗

Other INSTRUMENT, BIOPSY devices

All INSTRUMENT, BIOPSY devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.