ENCOR® ENCFINSERT10GRecalls & reported problems

Senorx, Inc. · ENCOR® Breast Biopsy System Needle Guide Insert (GE™/Siemens™/Fischer™/Giotto™), 10G · Class 2

Adverse event reports
10
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00801741086465), received between Mar 21, 2017 and Sep 19, 2025 — classified as 10 malfunctions.

The most frequently reported specific device problem is “Accessory Incompatible” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 2 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Accessory Incompatible4
  • Difficult to Remove4
  • Device Emits Odor2
  • Device Markings/Labelling Problem2
  • Difficult To Position2
  • Device Contamination with Chemical or Other Material1
  • Tear, Rip or Hole in Device Packaging1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • No Consequences Or Impact To Patient3

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 19, 2025Device Emits Odor; Device Emits Odor

    ON (B)(6) 2025, A PATIENT UNDERWENT A BIOPSY PROCEDURE PERFORMED WITH THE ENCORE INSTRUMENT. DURING THE PROCEDURE, THE ENCORE CODES WERE FOUND IN BOXES THAT EMITTED A FOUL ODOR, CAUSING ILLNESS AMONG THE LABELING STAFF, INCLUDING SYMPTOMS SUCH AS NAUSEA, GENERAL MALAISE, AND DIARRHEA. THERE WAS NO PATIENT REPORTED INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23098792 ↗
  • MalfunctionSep 19, 2025Device Emits Odor; Device Emits Odor

    ON (B)(6) 2025, A PATIENT UNDERWENT A BIOPSY PROCEDURE PERFORMED WITH THE ENCORE INSTRUMENT. DURING THE PROCEDURE, THE ENCORE CODES WERE FOUND IN BOXES THAT EMITTED A FOUL ODOR, CAUSING ILLNESS AMONG THE LABELING STAFF, INCLUDING SYMPTOMS SUCH AS NAUSEA, GENERAL MALAISE, AND DIARRHEA. THERE WAS NO PATIENT REPORTED INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23098444 ↗
  • MalfunctionOct 4, 2023Accessory Incompatible; Difficult to Remove; Accessory Incompatible; Difficult to Remove

    IT WAS REPORTED THAT DURING A STEREOTACTIC BREAST BIOPSY PROCEDURE THROUGH NORMAL DENSITY TISSUE, THE NEEDLE GUIDE WAS ALLEGEDLY TOO TIGHT AND TOOK SO LONG TO REMOVE THE NEEDLE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17866653 ↗
  • MalfunctionOct 4, 2023Accessory Incompatible; Difficult to Remove; Accessory Incompatible; Difficult to Remove

    IT WAS REPORTED THAT DURING A STEREOTACTIC BREAST BIOPSY PROCEDURE THROUGH NORMAL DENSITY TISSUE, THE NEEDLE GUIDE WAS ALLEGEDLY TOO TIGHT AND TOOK SO LONG TO REMOVE THE NEEDLE. IT WAS FURTHER REPORTED THAT THE NEEDLE GUIDE WAS TOO SMALL. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17866438 ↗
  • MalfunctionJun 1, 2023Accessory Incompatible; Difficult to Remove; Accessory Incompatible; Difficult to Remove

    IT WAS REPORTED THAT DURING A BIOPSY PROCEDURE, THE NEEDLE WAS ALLEGEDLY DIFFICULT TO BE PASSED THROUGH THE GUIDE. IT WAS FURTHER REPORTED THAT THE GUIDE WAS ALLEGEDLY DIFFICULT TO BE REMOVED. IT WAS ALSO REPORTED THAT THE THERE WAS AN ASYMMETRIC FLANGE INSIDE THE GUIDE WHICH LUGS AND SITS ALMOST FIRMLY OVER SEVERAL PARTS OF THE NEEDLE. IT WAS ALSO REPORTED THAT RESISTANCE WAS FELT WHEN PULLING THE FLANGE OFF. FURTHERMORE, AFTER THE GUIDE WAS FITTED, DEPENDING ON THE EXTENT THERE WAS TIGHTNESS AROUND THE BARBS. REPORTEDLY THE FLANGE/EDGE WAS COMPLETELY CLEAR AND SHARPLY DEFINED AND IT WAS SPREAD OVER APPROXIMATELY HALF THE CIRCLE RESULTING IN NOISE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17037026 ↗
  • MalfunctionJun 1, 2023Accessory Incompatible; Difficult to Remove; Accessory Incompatible; Difficult to Remove

    IT WAS REPORTED THAT DURING A BIOPSY PROCEDURE, THE NEEDLE WAS ALLEGEDLY DIFFICULT TO BE PASSED THROUGH THE GUIDE. IT WAS FURTHER REPORTED THAT THE GUIDE WAS ALLEGEDLY DIFFICULT TO BE REMOVED. IT WAS ALSO REPORTED THAT THE THERE WAS AN ASYMMETRIC FLANGE INSIDE THE GUIDE WHICH LUGS AND SITS ALMOST FIRMLY OVER SEVERAL PARTS OF THE NEEDLE. IT WAS ALSO REPORTED THAT RESISTANCE WAS FELT WHEN PULLING THE FLANGE OFF. FURTHERMORE, AFTER THE GUIDE WAS FITTED, DEPENDING ON THE EXTENT THERE WAS TIGHTNESS AROUND THE BARBS. REPORTEDLY THE FLANGE/EDGE WAS COMPLETELY CLEAR AND SHARPLY DEFINED AND IT WAS SPREAD OVER APPROXIMATELY HALF THE CIRCLE RESULTING IN NOISE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17036980 ↗
  • MalfunctionJul 20, 2021Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging

    IT WAS REPORTED THAT PRIOR TO BIOPSY PROCEDURE, THE STERILIZED PACKAGE WAS ALLEGEDLY FOUND TO BE DAMAGED. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12195111 ↗
  • MalfunctionNov 8, 2018Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material

    IT WAS REPORTED THAT AFTER COMPLETING A STEREOTACTIC-GUIDED BIOPSY, WHEN NEEDLE WAS REMOVED, ALLEGEDLY A PLASTIC PIECE WAS FOUND IN THE NEEDLE GUIDE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8053036 ↗
  • MalfunctionAug 29, 2017Difficult To Position; Device Markings/Labelling Problem; Difficult To Position; Device Markings/Labelling Problem

    IT WAS REPORTED THAT DURING PREPARATION FOR A STEREOTACTIC BREAST BIOPSY, THE PROBE ALLEGEDLY WAS UNABLE TO GO THROUGH THE NEEDLE GUIDE INSERT TO POSITION FOR THE PROCEDURE. REPORTEDLY, ANOTHER INSERT WAS ABLE TO PERFORM THE PROCEDURE. THERE WAS NO REPORTED PATIENT CONTACT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6827338 ↗
  • MalfunctionMar 21, 2017Difficult To Position; Device Markings/Labelling Problem; Difficult To Position; Device Markings/Labelling Problem

    IT WAS REPORTED THAT DURING A STEREOTACTIC BREAST BIOPSY, THE BIOPSY NEEDLE GUIDE WAS ALLEGEDLY TO NARROW AND THE PROBE WAS UNABLE TO INSERT FOR THE PROCEDURE. REPORTEDLY, ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6420511 ↗

Other INSTRUMENT, BIOPSY devices

All INSTRUMENT, BIOPSY devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.