PowerPICC SV Catheter 1274108DRecalls & reported problems
Bard Access Systems, Inc. · PowerPICC SV Catheter with Sherlock 3CG Tip Positioning System (TPS) Stylet Maximal Barrier Kit, 4F Dual-Lumen 55cm · Class 2 · Implant
All models of PowerPICC SV Catheter (15)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Bard PowerPICC SV catheter is a long-term central line placed through a vein in the arm. The 30 reports mostly describe the catheter kinking, cracking or deforming, and leaking blood or fluid, sometimes forcing removal and replacement; 2 injuries and no deaths are reported. One open recall, started in May 2026, concerns lidocaine ampoules in the kit made under deficient manufacturing practices.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 30
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 30 adverse event reports that list this exact device (DI 00801741087110), received between Apr 11, 2017 and Feb 3, 2026 — classified as 2 injuries, 28 malfunctions.
The most frequently reported specific device problem is “Material Deformation” (10 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- May 21, 2026Class IIOpen, Classified
Due to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled by supplier.
Z-2659-2026 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
Most reported problems
Device problems
- Material Deformation10
- Fluid/Blood Leak9
- Leak/Splash5
- Break2
- Crack2
- Physical Resistance/Sticking2
- Adverse Event Without Identified Device or Use Problem1
- Appropriate Device Problem Term/Code Not Available1
Patient problems
- No Clinical Signs, Symptoms or Conditions15
- No Consequences Or Impact To Patient9
- No Known Impact Or Consequence To Patient7
- Device Embedded In Tissue or Plaque1
Recent adverse event reports
The 25 most recent of 30 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 3, 2026Contamination /Decontamination Problem
IT WAS REPORTED THAT A PEN WAS FOUND INSIDE THE KIT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24251889 ↗MalfunctionJan 5, 2026Fluid/Blood Leak
IT WAS REPORTED THAT THE CATHETER WAS NOTED TO BE LEAKING FROM THE RED LUMEN AND THE TEAM IDENTIFIED A HOLE IN THE CATHETER. PICC EXCHANGE WAS PERFORMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23958392 ↗MalfunctionFeb 5, 2025Material Deformation; Material Deformation
IT WAS REPORTED, KINKED PICC. CATHETER PLACED (B)(6) 2024 IN BASILIC VEIN. CATHETER LENGTH: 40CM, 0CM EXIT SITE MARKINGS. STATLOCK USED. KINK NOTED AT 30CM MARK. DWELL TIME: 5 DAYS. GENERAL LOCATION: SUBCLAVIAN REGION. CATHETER EXCHANGED. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21317144 ↗MalfunctionFeb 5, 2025Material Deformation; Material Deformation
IT WAS REPORTED, KINKED PICC. CATHETER PLACED (B)(6)2024 IN BRACHIAL VEIN. CATHETER LENGTH: 34CM, 3CM EXIT SITE MARKINGS. STATLOCK USED. KINK NOTED AT 12CM MARK. GENERAL LOCATION: AXILLARY REGION. CATHETER PULLED BACK. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21317095 ↗MalfunctionJan 20, 2025Material Deformation; Material Deformation
PICC WAS INSERTED ON (B)(6) 2024 IN THE RIGHT BRACHIAL VEIN WITH A TOTAL CATHETER LENGTH OF 37 CM AND 0 CM LEFT EXTERNAL. CATHETER SECURED WITH STATLOCK. KINK WAS DISCOVERED IN THE INNOMINATE REGION 6 DAYS AFTER INSERTION. CATHETER WAS EXCHANGED. CATHETER HAD BEEN TPA'D 1X DURING THE 6-DAY INDWELL PERIOD.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21191457 ↗MalfunctionJan 10, 2025Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT A PICC WAS PLACED ON (B)(6) 2024 MEASURING AT 45 CM WITH 1 CM VISIBLE SECUREACATH DEVICE WAS USED TO SECURE THE PICC IN PLACE. ON (B)(6) 2024 THE PICC WAS FOUND TO BE LEAKING PICC AND WAS REMOVED, A NEW PICC PLACEMENT WAS ATTEMPTED BUT FAILED. NO OTHER INFORMATION PROVIDED, AT THIS TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21125518 ↗MalfunctionJan 9, 2025Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT A PICC WAS PLACED ON (B)(6) 2024 MEASURING AT 39 CM WITH 3CM VISIBLE WITH SECUREMENT DEVICE USED TO KEEP THE PICC IN PLACE. ON (B)(6) 2024 THE PICC WAS FOUND TO BE LEAKING AT THE 2 CM MARK. PICC WAS REMOVED, WITH NEW PICC PLACED ON (B)(6) 2024. NO OTHER INFORMATION PROVIDED, AT THIS TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21120808 ↗MalfunctionJan 4, 2025Material Deformation; Appropriate Device Problem Term/Code Not Available; Material Deformation; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT 2 PICCS KINKED. ONE PATIENT HAD THEIR PICC EXCHANGED, AND THE OTHER PATIENT WAS SENT FOR IR REFERRAL. PICC DWELL TIMES WERE 4 DAYS AND 22 DAYS. THIS REPORT ADDRESSES BOTH DEVICES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21075631 ↗MalfunctionOct 19, 2023Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED BY THE CUSTOMER THAT PICC WAS LEAKING FROM HOLE IN LINE. NEW PICC HAD TO BE PLACED. IT IS A PEDIATRIC PATIENT SO THE PATIENT HAD TO BE PUT UNDER SEDATION AGAIN FOR THIS TO BE DONE. NO OTHER INFORMATION REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17965158 ↗MalfunctionMay 12, 2023Material Deformation; Material Deformation
IT WAS REPORTED BY THE CUSTOMER, "A 4FR DL POWERPICC PROVENA INSERTED IN THE RIGHT UPPER ARM BRACHIAL VEIN ON (B)(6) 2022 WAS FOUND TO BE KINKED ON (B)(6) 2022 IN THE SUBCLAVIAN REGION. PICC DWELL TIME WAS 14 DAYS. PICC WAS EXCHANGED."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16924394 ↗MalfunctionJan 5, 2023Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED BY THE CUSTOMER "EXPERIENCED LEAKAGE FROM THE WHITE PORT OF THE PICC." NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16102640 ↗MalfunctionJan 5, 2023Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED BY THE CUSTOMER, "EXPERIENCED LEAKAGE FROM THE WHITE PORT OF THE PICC." NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16102368 ↗MalfunctionDec 14, 2021Material Deformation; Material Deformation
IT WAS REPORTED "STYLET BENT." NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13002736 ↗InjuryFeb 4, 2021Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED VIA MS&S "CALLER STATES A PATIENT HAD A REACTION TO THE HICKMAN CVC, 0600580. CALLER STATES THERE WAS WEEPING DRAINAGE AT THE SITE FROM THIS DEVICE. CALLER STATES THE PATIENT HAD A SIMILAR REACTION TO PICC LINE WITH LOT NUMBER REES3172 (POWERPICC SV 1274108DNS). CALLER ASKING FOR THE MATERIALS OF THESE DEVICES AND IF THEY ARE SIMILAR." MS&S RESPONDED "THE HICKMAN CVC 0600580 IS SILICONE. THE PICC LINE 1274108DNS IS POLYURETHANE. EXPLAINED THAT OFTEN PATIENTS CAN HAVE REACTIONS TO DRESSINGS PLACED AT THE INSERTION SITE OR SITE CLEANING MATERIALS LIKE CHLORHEXIDINE, POVIDONE IODINE, ADHESIVE, ETC. DISCUSSED HAVING THE PATIENT TESTED BY ALLERGIST FOR SENSITIVITIES. THIS REPORT ADDRESSES THE REACTION TO THE POWERPICC SV.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 11281098 ↗MalfunctionDec 23, 2020Leak/Splash; Material Split, Cut or Torn
IT WAS REPORTED PICC LINE PLACED AND LEAKING NOTED NEAR TEAR WHEN FLUSHING. THE LINE WAS REMOVED AND REPLACED WITH A NEW ONE. IT WAS REPORTED THIS EVENT OCCURRED TWICE. THIS REPORT ADDRESSES THE SECOND EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 11061789 ↗MalfunctionDec 23, 2020Leak/Splash
IT WAS REPORTED PICC LINE PLACED AND LEAKING NOTED NEAR TEAR WHEN FLUSHING. THE LINE WAS REMOVED AND REPLACED WITH A NEW ONE. IT WAS REPORTED THIS EVENT OCCURRED TWICE. THIS REPORT ADDRESSES THE FIRST EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 11061785 ↗MalfunctionDec 22, 2020Material Deformation; Physical Resistance/Sticking; Material Deformation
IT WAS REPORTED DEFECTIVE GUIDE-WIRE OR 4.5F INTRODUCER, STUCK UPON INSERTION AND ATTEMPT TO WITHDRAWAL. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 11060917 ↗MalfunctionDec 22, 2020Material Deformation; Physical Resistance/Sticking; Material Deformation
IT WAS REPORTED THE LINE WAS STUCK UPON INSERTION AND WITHDRAWAL AND SURGEON HAD TO REMOVE. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 11060907 ↗MalfunctionNov 23, 2020Fluid/Blood Leak; Material Puncture/Hole; Material Twisted/Bent
IT WAS REPORTED "WRITER CALLED TO ASSESS PICC LINE THIS MORNING. RN STATED THAT WHEN DISTAL PORT OF PICC LINE WAS FLUSHED THAT BIOPATCH AND DRESSING BECAME SATURATED. WRITER ASSESSED LINE AND AGREED WITH ASSESSMENT. WRITER AND RN SPOKE WITH COVERING TEAM AND DECISION WAS MADE TO D/C PICC AND SEE HOW PATIENT WOULD DO. PIV PLACED FOR ANTIBIOTICS AND PICC LINE D/C'D. WHEN PICC REMOVED THERE WAS A TWIST IN THE CATHETER AND VISIBLE HOLE IN DISTAL PORTION OF PICC LINE. PATIENT SCHEDULED FOR SEDATION ON THURSDAY MORNING IN CASE PICC LINE NEEDS TO BE REPLACED." DID EVENT OCCUR DURING PLACEMENT, DURING REMOVAL, OR AFTER REMOVAL? "WHILE IN-DWELLING".
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10886540 ↗MalfunctionAug 21, 2020Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THE POWERPICC SV HAD A HOLE ON THE CATHETER AT THE PROXIMAL END WHERE IT CONNECTS TO THE STIFF SUTURE WING. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10437255 ↗MalfunctionAug 21, 2020Material Deformation; Material Deformation
IT WAS REPORTED THERE WAS A KINK NEAR THE ¿0¿ MARK BUT NO HOLE. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10437218 ↗MalfunctionJun 13, 2019Fracture
IT WAS REPORTED THAT AFTER REMOVING THE 3CG STYLET FROM THE PICC LINE IT WAS NOTED THAT THE STYLET WIRE WAS FRACTURED TOWARDS THE END. NO HARM TO PATIENT WAS DONE. INSERTER STATES WIRE WAS NOT CUT DURING PICC TRIMMING. COMPLAINANT STATES WAS NOT PRESENT AND CANNOT CONFIRM IF IT WAS CUT OR NOT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8696678 ↗MalfunctionJan 8, 2019Leak/Splash
IT WAS REPORTED THAT A LEAK IN 4FR RED LUMEN WAS OBSERVED. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8228941 ↗MalfunctionDec 19, 2018Leak/Splash; Tear, Rip or Hole in Device Packaging
IT WAS REPORTED THAT THE PATIENT WAS READMITTED TO THE HOSPITAL WITH A LEAKING PICC. THE STAFF CONFIRMED THERE WAS A PIN HOLE LEAK DISTAL TO THE INSERTION SITE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8180204 ↗MalfunctionMar 12, 2018Fluid/Blood Leak
IT WAS REPORTED THAT THE PICC WAS LEAKING FROM ONE OF THE ACCESS PORTS. A NEW DEVICE WAS PLACED. NO PATIENT INJURY REPORTED, ONLY A DELAY IN CARE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7333007 ↗
Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices
- Groshong NXT ClearVue Catheter 7617405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,708 reports
- Groshong NXT ClearVue Catheter 7655405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,536 reports
- PowerPICC Catheter 3174118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall402 reports
- L-Cath™ S/L PICC S/L 26GA (1.9F) x 30cmArgon Medical Devices, Inc. · Catheter,Intravascular,Therapeutic,Long-Term Greater Than 30 Days2 recalls245 reports
- PowerPICC SOLO2 Catheter 1194108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 recalls216 reports
- Broviac 0600520Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall265 reports
- Groshong® Catheter 7717405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days279 reports
- PowerPICC Catheter 3275118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days272 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page