Quick summary

This Bard EZ Huber needle is used to access implanted venous ports for infusions. The 58 reports mostly describe packaging and component problems — sealed packaging that wasn't sealed, broken parts, and needles without needle guards — along with blood leaking from the tubing or a clamp that wouldn't close, and ports that could no longer be flushed. No deaths or injuries are reported. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
58
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 58 adverse event reports that list this exact device (DI 00801741089138), received between Jun 16, 2017 and May 23, 2024 — classified as 58 malfunctions.

The most frequently reported specific device problem is “Unsealed Device Packaging” (31 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Unsealed Device Packaging31
  • Defective Component12
  • Loss of or Failure to Bond9
  • Packaging Problem9
  • Device Packaging Compromised2
  • Fluid/Blood Leak2
  • Component Missing1
  • Detachment of Device or Device Component1

Patient problems

  • No Patient Involvement36
  • No Clinical Signs, Symptoms or Conditions15
  • No Known Impact Or Consequence To Patient7

Recent adverse event reports

The 25 most recent of 58 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 23, 2024Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED BY THE CUSTOMER "I WANTED TO GIVE THE PATIENT ANTIBIOTICS VIA THE VENOUS PORT. ABOUT 3 HOURS BEFORE, IT WAS STILL POSSIBLE TO GIVE INFUSIONS WITHOUT ANY PROBLEMS. NOW, HOWEVER, IT WAS NO LONGER POSSIBLE TO EITHER FLUSH OR ASPIRATE BLOOD. BLOOD HAD COLLECTED AT THE FRONT OF THE LINE (AT THE Y-CONNECTION). IN THE END, THE NEEDLE HAD TO BE PULLED OUT. 2 NURSES HAD TRIED TO FLUSH. AFTER THE NEEDLE WAS PULLED OUT, I TRIED TO FLUSH IT AGAIN, IT WAS VERY DIFFICULT AT FIRST, BUT THEN IT WORKED. THERE WAS A RISK OF AN AIR EMBOLISM DUE TO THE LEAK IN THE TUBE. HOWEVER, THE PATIENT SHOWED NO SYMPTOMS OF AN AIR EMBOLISM AND NEVER EXPRESSED ANY COMPLAINTS." NO OTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19381105 ↗
  • MalfunctionNov 30, 2023Component Missing; Component Missing

    IT WAS REPORTED BY THE CUSTOMER "I WANTED TO PIERCE A PATIENT'S PORT AND PREPARED THE MATERIAL. I UNPACKED THE NEEDLE AND DROPPED IT ON THE STERILE SURFACE FROM THE PACKAGING. WHEN I WENT TO TAKE THE NEEDLE, I SAW THAT THERE WAS NO NEEDLE GUARD. THE PRODUCT WAS NOT USED. SO NO CONSEQUENCES FOR THE PATIENT. UNFORTUNATELY, A SAMPLE WILL NOT BE AVAILABLE AS THE CUSTOMER HAS DISPOSED OF IT ON SITE."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18243023 ↗
  • MalfunctionApr 4, 2023Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED, DURING PREPARATION THE NEEDLE IS DRIPPING AT THE HOSE ATTACHMENT. THERE IS NO IMPACT ON PATIENT, USER, OR THIRD PARTY. NO OTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16678842 ↗
  • MalfunctionJun 2, 2022Defective Component; Defective Component

    IT WAS REPORTED "WITH PORT NEEDLES, THE CLAMP DOESN'T CLOSE PROPERLY, BLOOD FLOW CONTINUES. PATIENT CAN BLEED OUT IF UNNOTICED." IT WAS REPORTED THIS OCCURRED WITH EIGHT DEVICES. THIS REPORT ADDRESSES THE SEVENTH DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14582214 ↗
  • MalfunctionJun 2, 2022Defective Component; Defective Component

    IT WAS REPORTED "WITH PORT NEEDLES, THE CLAMP DOESN'T CLOSE PROPERLY, BLOOD FLOW CONTINUES. PATIENT CAN BLEED OUT IF UNNOTICED." IT WAS REPORTED THIS OCCURRED WITH EIGHT DEVICES. THIS REPORT ADDRESSES THE SECOND DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14582211 ↗
  • MalfunctionJun 2, 2022Defective Component; Defective Component

    IT WAS REPORTED "WITH PORT NEEDLES, THE CLAMP DOESN'T CLOSE PROPERLY, BLOOD FLOW CONTINUES. PATIENT CAN BLEED OUT IF UNNOTICED." IT WAS REPORTED THIS OCCURRED WITH FOUR DEVICES. THIS REPORT ADDRESSES THE SECOND DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14582208 ↗
  • MalfunctionJun 2, 2022Defective Component; Defective Component

    IT WAS REPORTED "WITH PORT NEEDLES, THE CLAMP DOESN'T CLOSE PROPERLY, BLOOD FLOW CONTINUES. PATIENT CAN BLEED OUT IF UNNOTICED." IT WAS REPORTED THIS OCCURRED WITH EIGHT DEVICES. THIS REPORT ADDRESSES THE FOURTH DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14582135 ↗
  • MalfunctionJun 2, 2022Defective Component; Defective Component

    IT WAS REPORTED "WITH PORT NEEDLES, THE CLAMP DOESN'T CLOSE PROPERLY, BLOOD FLOW CONTINUES. PATIENT CAN BLEED OUT IF UNNOTICED." IT WAS REPORTED THIS OCCURRED WITH EIGHT DEVICES. THIS REPORT ADDRESSES THE THIRD DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14582127 ↗
  • MalfunctionJun 2, 2022Defective Component; Defective Component

    IT WAS REPORTED "WITH PORT NEEDLES, THE CLAMP DOESN'T CLOSE PROPERLY, BLOOD FLOW CONTINUES. PATIENT CAN BLEED OUT IF UNNOTICED." IT WAS REPORTED THIS OCCURRED WITH FOUR DEVICES. THIS REPORT ADDRESSES THE THIRD DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14582120 ↗
  • MalfunctionJun 2, 2022Defective Component; Defective Component

    IT WAS REPORTED "WITH PORT NEEDLES, THE CLAMP DOESN'T CLOSE PROPERLY, BLOOD FLOW CONTINUES. PATIENT CAN BLEED OUT IF UNNOTICED." IT WAS REPORTED THIS OCCURRED WITH EIGHT DEVICES. THIS REPORT ADDRESSES THE FIFTH DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14582079 ↗
  • MalfunctionJun 2, 2022Defective Component; Defective Component

    IT WAS REPORTED "WITH PORT NEEDLES, THE CLAMP DOESN'T CLOSE PROPERLY, BLOOD FLOW CONTINUES. PATIENT CAN BLEED OUT IF UNNOTICED." IT WAS REPORTED THIS OCCURRED WITH FOUR DEVICES. THIS REPORT ADDRESSES THE FOURTH DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14582072 ↗
  • MalfunctionJun 2, 2022Defective Component; Defective Component

    IT WAS REPORTED "WITH PORT NEEDLES, THE CLAMP DOESN'T CLOSE PROPERLY, BLOOD FLOW CONTINUES. PATIENT CAN BLEED OUT IF UNNOTICED." IT WAS REPORTED THIS OCCURRED WITH EIGHT DEVICES. THIS REPORT ADDRESSES THE SIXTH DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14582036 ↗
  • MalfunctionJun 2, 2022Defective Component; Defective Component

    IT WAS REPORTED "WITH PORT NEEDLES, THE CLAMP DOESN'T CLOSE PROPERLY, BLOOD FLOW CONTINUES. PATIENT CAN BLEED OUT IF UNNOTICED." IT WAS REPORTED THIS OCCURRED WITH FOUR DEVICES. THIS REPORT ADDRESSES THE FIRST DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14581979 ↗
  • MalfunctionJun 2, 2022Defective Component; Defective Component

    IT WAS REPORTED "WITH PORT NEEDLES, THE CLAMP DOESN'T CLOSE PROPERLY, BLOOD FLOW CONTINUES. PATIENT CAN BLEED OUT IF UNNOTICED." IT WAS REPORTED THIS OCCURRED WITH EIGHT DEVICES. THIS REPORT ADDRESSES THE EIGHTH DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14581856 ↗
  • MalfunctionJun 2, 2022Defective Component; Defective Component

    IT WAS REPORTED "WITH PORT NEEDLES, THE CLAMP DOESN'T CLOSE PROPERLY, BLOOD FLOW CONTINUES. PATIENT CAN BLEED OUT IF UNNOTICED." IT WAS REPORTED THIS OCCURRED WITH EIGHT DEVICES. THIS REPORT ADDRESSES THE FIRST DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14581819 ↗
  • MalfunctionAug 17, 2018Detachment of Device or Device Component

    IT WAS REPORTED THAT THE HUBER NEEDLE WAS REMOVED IN THE CORRECT MANNER AND THE NEEDLE CAME DETACHED FROM THE HUB, THE NEEDLE REMAINED IN PATIENTS PORT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7793400 ↗
  • MalfunctionJun 16, 2017Loss of or Failure to Bond; Packaging Problem

    IT WAS REPORTED THAT THE PACKAGING WAS NOT COMPLETELY SEALED AROUND THE EDGE. DEVICE WAS NOT USED ON PATIENT. THIS REPORTED ADDRESSES DEVICE FORTY-TWO.

    Patient problems: No Patient Involvement

    MAUDE report 6648396 ↗
  • MalfunctionJun 16, 2017Unsealed Device Packaging

    IT WAS REPORTED THAT THE PACKAGING WAS NOT COMPLETELY SEALED AROUND THE EDGE. DEVICE WAS NOT USED ON PATIENT. THIS REPORTED ADDRESSES DEVICE THIRTY-EIGHT.

    Patient problems: No Patient Involvement

    MAUDE report 6648394 ↗
  • MalfunctionJun 16, 2017Unsealed Device Packaging

    IT WAS REPORTED THAT THE PACKAGING WAS NOT COMPLETELY SEALED AROUND THE EDGE. DEVICE WAS NOT USED ON PATIENT. THIS REPORTED ADDRESSES DEVICE THIRTY-TWO.

    Patient problems: No Patient Involvement

    MAUDE report 6648388 ↗
  • MalfunctionJun 16, 2017Unsealed Device Packaging

    IT WAS REPORTED THAT THE PACKAGING WAS NOT COMPLETELY SEALED AROUND THE EDGE. DEVICE WAS NOT USED ON PATIENT. THIS REPORTED ADDRESSES DEVICE THIRTY.

    Patient problems: No Patient Involvement

    MAUDE report 6648387 ↗
  • MalfunctionJun 16, 2017Loss of or Failure to Bond; Packaging Problem

    IT WAS REPORTED THAT THE PACKAGING WAS NOT COMPLETELY SEALED AROUND THE EDGE. DEVICE WAS NOT USED ON PATIENT. THIS REPORTED ADDRESSES DEVICE FORTY-ONE.

    Patient problems: No Patient Involvement

    MAUDE report 6648384 ↗
  • MalfunctionJun 16, 2017Unsealed Device Packaging

    IT WAS REPORTED THAT THE PACKAGING WAS NOT COMPLETELY SEALED AROUND THE EDGE. DEVICE WAS NOT USED ON PATIENT. THIS REPORTED ADDRESSES DEVICE THIRTY-FIVE.

    Patient problems: No Patient Involvement

    MAUDE report 6648381 ↗
  • MalfunctionJun 16, 2017Unsealed Device Packaging

    IT WAS REPORTED THAT THE PACKAGING WAS NOT COMPLETELY SEALED AROUND THE EDGE. DEVICE WAS NOT USED ON PATIENT. THIS REPORTED ADDRESSES DEVICE THIRTY-NINE.

    Patient problems: No Patient Involvement

    MAUDE report 6648376 ↗
  • MalfunctionJun 16, 2017Unsealed Device Packaging

    IT WAS REPORTED THAT THE PACKAGING WAS NOT COMPLETELY SEALED AROUND THE EDGE. DEVICE WAS NOT USED ON PATIENT. THIS REPORTED ADDRESSES DEVICE FORTY.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6648374 ↗
  • MalfunctionJun 16, 2017Unsealed Device Packaging

    IT WAS REPORTED THAT THE PACKAGING WAS NOT COMPLETELY SEALED AROUND THE EDGE. DEVICE WAS NOT USED ON PATIENT. THIS REPORTED ADDRESSES DEVICE THIRTY-SEVEN.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6648373 ↗

Other Set, administration, intravascular devices