Silicone Flat Drain 0070360Recalls & reported problems
C. R. Bard, Inc. · Hubless Silicone Flat Drain, 7mm Wide, 20cm Long, 3/4 Length Perforations, X-Ray Opaque Stripe · Class 1
All models of Silicone Flat Drain (5)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00801741090738), received between Dec 6, 2022 and Oct 20, 2025 — classified as 10 malfunctions.
The most frequently reported specific device problem is “Device Contamination with Chemical or Other Material” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Device Contamination with Chemical or Other Material5
- Material Fragmentation3
- Component Misassembled1
- Component Missing1
- Device Contaminated During Manufacture or Shipping1
- Sharp Edges1
Patient problems
- No Clinical Signs, Symptoms or Conditions10
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 20, 2025Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material
IT WAS REPORTED THAT THE CUSTOMER STATED FOREIGN MATTER (INSIDE THE STERILIZED PACKAGE), FRACTIONAL GOOD PRODUCT, BIOLOGICAL MATTER (INSIDE THE STERILIZED PACKAGE) FOREIGN MATTER (OUTSIDE THE STERILIZED PACKAGE) BIOLOGICAL MATTER (INSIDE THE STERILIZED PACKAGE) FRACTIONAL GOOD PRODUCT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23335348 ↗MalfunctionOct 17, 2025Device Contamination with Chemical or Other Material; Device Contaminated During Manufacture or Shipping; Device Contamination with Chemical or Other Material; Device Contaminated During Manufacture or Shipping
IT WAS REPORTED THAT THE CUSTOMER STATED FOREIGN MATTER (INSIDE THE STERILIZED PACKAGE) FRACTIONAL GOOD PRODUCT BIOLOGICAL MATTER (INSIDE THE STERILIZED PACKAGE) FOREIGN MATTER (OUTSIDE THE STERILIZED PACKAGE) BIOLOGICAL MATTER (INSIDE THE STERILIZED PACKAGE) FRACTIONAL GOOD PRODUCT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23328401 ↗MalfunctionSep 23, 2025Sharp Edges; Sharp Edges
IT WAS REPORTED THAT ON 16SEP2025, WHEN THE WOUND DRAIN PRODUCT WAS OPENED IN THE OPERATING ROOM, A BURR WAS FOUND, AND IT WAS DISCONTINUED. WHEN A PRODUCT FROM THE SAME LOT WAS OPENED, A BURR WAS ALSO FOUND, AND IT WAS DISCONTINUED FROM USE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23123078 ↗MalfunctionDec 9, 2024Material Fragmentation; Material Fragmentation
IT WAS REPORTED THAT THERE WAS A FRAYED EDGE ON THE TIP OF THE WOUND DRAIN, THE SIDE THAT WAS LEFT INSIDE THE BODY. THERE WAS DAMAGE TO THE TIP OF THE DRAIN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20878931 ↗MalfunctionOct 1, 2024Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material
IT WAS REPORTED THAT WHEN THE WOUND DRAIN WAS OPENED IN THE OPERATING ROOM FOR USE, A FOREIGN OBJECT WAS FOUND IN IT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20346085 ↗MalfunctionFeb 14, 2024Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material
IT WAS REPORTED THAT HAIR MIXED IN STERILE BAG OF WOUND DRAIN DEVICE. PER INTERNAL COMMENTS RECEIVED ON 30JAN2024, IT WAS CONFIRMED THAT IT WAS UNOPENED. THEY CONFIRMED THAT THERE WAS A FOREIGN OBJECT IN THE STERILIZED BAG. INSIDE THE BAG CONTAINING THE PRODUCT
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18702179 ↗MalfunctionSep 27, 2023Material Fragmentation; Component Misassembled; Material Fragmentation; Component Misassembled
IT WAS REPORTED THAT WHEN CHECKING THE WOUND DRAINAGE BEFORE USE, THE END OF THE TUBE WAS FOUND TO BE BROKEN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17825534 ↗MalfunctionAug 18, 2023Material Fragmentation; Material Fragmentation
IT WAS REPORTED THAT THE THIN SIDE OF THE WOUND DRAINAGE BLUE CONNECTOR WAS BROKEN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17571996 ↗MalfunctionApr 13, 2023Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material
IT WAS REPORTED THAT A FOREIGN MATERIAL WAS MIXED IN THE WOUND DRAIN DEVICE. THE CUSTOMER REQUESTED TO CONFIRM WHETHER THAT FOREIGN MATERIAL WAS A PIECE OF DRAIN TUBE OR NOT. PER FOLLOW-UP INFORMATION RECEIVED FROM IBC ON 30MAR2023, STATED THAT THE DEFECT WAS FOUND AFTER USE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16731619 ↗MalfunctionDec 6, 2022Component Missing; Component Missing
IT WAS REPORTED THAT THE CONNECTOR WAS MISSING IN THE PACKAGE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15923927 ↗
Other CATHETER, IRRIGATION devices
- Navarre Opti-Drain NOD8LPTBard Access Systems, Inc. · CATHETER, IRRIGATION901 reports
- Navarre NNU8LPTBard Access Systems, Inc. · CATHETER, IRRIGATION736 reports
- BLAKE 2227Ethicon Inc. · CATHETER, IRRIGATION331 reports
- Navarre NNU6LPTBard Access Systems, Inc. · CATHETER, IRRIGATION211 reports
- Navarre Opti-Drain NOD10LPTBard Access Systems, Inc. · CATHETER, IRRIGATION178 reports
- Navarre NNU10LPTBard Access Systems, Inc. · CATHETER, IRRIGATION127 reports
- BLAKE 2233Ethicon Inc. · CATHETER, IRRIGATION114 reports
- Vasoshield VH-5001Maquet Cardiovascular LLC · Catheter, Irrigation1 recall27 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page