MARQUEE MQ1210Recalls & reported problems
Bard Peripheral Vascular, Inc. · MARQUEE® Disposable Core Biopsy Instrument, 12g x 10cm · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00801741097119), received between Jul 19, 2019 and Mar 17, 2026 — classified as 1 injury, 8 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (4 reports).
2 FDA recall records name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Sep 30, 2024Open, Classified
Disposable biopsy instruments were potentially manufactured missing the backend notch retention feature on the needle that results in inadequate securing of the needle inside the device, which could lead to needle detachment and misplacement, which could lead to tissue damage and bleeding.
Z-0467-2025 · Root cause: Process change control · FDA record ↗
- Jan 14, 2021Terminated
Due to the Penetration Depth Switch used to select the desired penetration depth of the biopsy needle for soft tissue biopsies, detaches from the device during use.
Z-1289-2021 · Root cause: Process control · FDA record ↗
Most reported problems
Device problems
- Detachment of Device or Device Component4
- Dull, Blunt2
- Self-Activation or Keying2
- Adverse Event Without Identified Device or Use Problem1
- Break1
- Crack1
- Device Contamination with Chemical or Other Material1
- Difficult to Remove1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- No Consequences Or Impact To Patient1
- Pain1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 17, 2026Detachment of Device or Device Component
ON (B)(6) 2026, A PATIENT WAS PREPPED FOR A BIOPSY PROCEDURE BY USING MARQUEE DEVICE. IT WAS REPORTED THAT PRIOR TO THE PROCEDURE THE ITEM ALLEGEDLY FELL APART WHILE SETTING UP. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24615700 ↗MalfunctionJul 2, 2025Crack; Detachment of Device or Device Component; Crack; Detachment of Device or Device Component
ON (B)(6) 2025, A PATIENT UNDERWENT A BIOPSY PROCEDURE USING THE MARQUEE. PRIOR THE PROCEDURE, THE DEVICE THE OUTER CANNULA FLEW OFF THE NEEDLE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22383117 ↗MalfunctionMay 20, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING AN ULTRASOUND GUIDED BREAST BIOPSY PROCEDURE THROUGH NORMAL TISSUE DENSITY, THE NEEDLE ALLEGEDLY CAME OUT OF THE WHITE PART OF THE DEVICE. NO COAXIAL WAS USED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19349671 ↗MalfunctionMay 20, 2024Nonstandard Device; Detachment of Device or Device Component; Nonstandard Device; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING AN ULTRASOUND GUIDED BREAST BIOPSY PROCEDURE THROUGH NORMAL TISSUE DENSITY, THE NEEDLE ALLEGEDLY CAME OUT OF THE WHITE PART OF THE DEVICE. NO COAXIAL WAS USED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19349638 ↗MalfunctionJun 20, 2022Self-Activation or Keying; Dull, Blunt; Self-Activation or Keying; Dull, Blunt
IT WAS REPORTED THAT PRIOR TO BIOPSY PROCEDURE, THE DEVICE ALLEGEDLY SELF ACTIVATED. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14741711 ↗MalfunctionMay 20, 2022Break; Self-Activation or Keying; Dull, Blunt; Break; Self-Activation or Keying; Dull, Blunt
IT WAS REPORTED THAT DURING A BIOPSY PROCEDURE, THE DEVICE ALLEGEDLY EXPLODED WHEN IT WAS BEING PRIMED. IT WAS FURTHER REPORTED THAT THE TIP WAS ALLEGEDLY DO NOT FEEL AS SHARP. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14460652 ↗MalfunctionFeb 16, 2021Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material
IT WAS REPORTED THAT PRIOR TO A BIOPSY PROCEDURE, BLOOD WAS FOUND ON THE TIP OF THE NEEDLE WHEN OPENING STERILE PACKAGE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11331186 ↗InjuryFeb 4, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THROUGH A POST-MARKET CLINICAL FOLLOW-UP SURVEY ON CORE BIOPSY INSTRUMENTS, THERE WERE THREE OCCURRENCES OF PAIN DURING LYMPH NODE BIOPSY PROCEDURE(S). IT WAS FURTHER REPORTED THAT THE TREATMENT PROVIDED WAS PAINKILLERS. THE CURRENT STATUS OF THE PATIENT(S) WAS NOT PROVIDED.
Patient problems: Pain
MAUDE report 11277362 ↗MalfunctionJul 19, 2019Difficult to Remove; Failure to Obtain Sample; Difficult to Remove; Failure to Obtain Sample
THIS REPORT SUMMARIZES TWO MALFUNCTIONS. A REVIEW OF THE REPORTED INFORMATION INDICATED THAT AN MQ1210 BIOPSY INSTRUMENT WAS ALLEGEDLY DIFFICULT TO REMOVE AND FAILED TO OBTAIN SAMPLES. THIS INFORMATION WAS RECEIVED FROM VARIOUS SOURCES. THE MALFUNCTIONS INVOLVED TWO PATIENTS WITH NO PATIENT CONSEQUENCE. THE AGE, WEIGHT, AND SEX WERE NOT REPORTED FOR THE PATIENT INVOLVED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8807753 ↗
Other INSTRUMENT, BIOPSY devices
- MAX-CORE MC1616Bard Peripheral Vascular, Inc. · INSTRUMENT, BIOPSY1 recall4,258 reports
- ENCOR® ECP017GSenorx, Inc. · INSTRUMENT, BIOPSY1 recall4,186 reports
- MAX-CORE MC1820Bard Peripheral Vascular, Inc. · INSTRUMENT, BIOPSY1 recall2,404 reports
- MAX-CORE MC1816Bard Peripheral Vascular, Inc. · INSTRUMENT, BIOPSY1 recall1 death848 reports
- MAX-CORE MC1410Bard Peripheral Vascular, Inc. · INSTRUMENT, BIOPSY1 recall399 reports
- MAX-CORE MC1610Bard Peripheral Vascular, Inc. · INSTRUMENT, BIOPSY1 recall394 reports
- ENCOR® ECP0110GVSenorx, Inc. · INSTRUMENT, BIOPSY392 reports
- ENCOR® ECP0110GSenorx, Inc. · INSTRUMENT, BIOPSY390 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page