MARQUEE MQ1410Recalls & reported problems

Bard Peripheral Vascular, Inc. · MARQUEE® Disposable Core Biopsy Instrument, 14g x 10cm · Class 2

Quick summary

This Bard MARQUEE disposable core biopsy instrument is a hand-held needle gun used to take tissue samples, mainly from the breast. The 51 reports mostly describe the device misfiring on its own, sticking so it can't be removed from the patient, breaking, or parts detaching and small plastic pieces breaking off. No deaths or injuries were reported. One recall, started in January 2021 and since terminated, involved the penetration depth switch detaching during use.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
51
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 51 adverse event reports that list this exact device (DI 00801741097133), received between Jan 6, 2021 and Mar 5, 2026 — classified as 51 malfunctions.

The most frequently reported specific device problem is “Self-Activation or Keying” (18 reports).

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 8 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 1 reportsAug 2024: 3 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 2 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 2 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520268
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jan 14, 2021Terminated

    Due to the Penetration Depth Switch used to select the desired penetration depth of the biopsy needle for soft tissue biopsies, detaches from the device during use.

    Z-1289-2021 · Root cause: Process control · FDA record ↗

Most reported problems

Device problems

  • Self-Activation or Keying18
  • Difficult to Remove11
  • Break9
  • Device Contamination with Chemical or Other Material9
  • Detachment of Device or Device Component7
  • Packaging Problem7
  • Tear, Rip or Hole in Device Packaging7
  • Unsealed Device Packaging7

Patient problems

  • No Clinical Signs, Symptoms or Conditions51

Recent adverse event reports

The 25 most recent of 51 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 5, 2026Nonstandard Device; Detachment of Device or Device Component

    ON (B)(6) 2026, A PATIENT UNDERWENT A BIOPSY PROCEDURE USING THE MARQUEE INSTRUMENT. IT WAS REPORTED THAT DURING THE SECOND BIOPSY ATTEMPT, THE CANNULA (GRADUATED PORTION) REMAINED INSIDE THE BREAST. A SUBSEQUENT ATTEMPT WAS MADE OUTSIDE THE BREAST, DURING WHICH THE TROCAR ALLEGEDLY RELEASED UNEXPECTEDLY AND SHOT TOWARD THE WALL. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24517271 ↗
  • MalfunctionOct 29, 2025Self-Activation or Keying; Failure to Obtain Sample; Self-Activation or Keying; Failure to Obtain Sample

    ON (B)(6) 2025, A PATIENT UNDERWENT A BREAST BIOPSY PROCEDURE USING THE MARQUEE INSTRUMENT WITH IMAGING OF ECHOGRAPHY WITH NORMAL DENSITY OF LESION. DURING THE PROCEDURE, IT WAS REPORTED THAT DEVICE DOES NOT OBTAIN GOOD SAMPLES. IT WAS FURTHER REPORTED THAT THE PROBE WAS FIRED AUTOMATICALLY WHEN THE PROBE IS IN SEMIAUTOMATIC MODE. NO COAXIAL NEEDLES WERE USED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23412742 ↗
  • MalfunctionOct 29, 2025Self-Activation or Keying; Failure to Obtain Sample; Self-Activation or Keying; Failure to Obtain Sample

    ON (B)(6) 2025, A PATIENT UNDERWENT A BREAST BIOPSY PROCEDURE USING THE MARQUEE INSTRUMENT WITH IMAGING OF ECHOGRAPHY WITH NORMAL DENSITY OF LESION. DURING THE PROCEDURE, IT WAS REPORTED THAT DEVICE DOES NOT OBTAIN GOOD SAMPLES. IT WAS FURTHER REPORTED THAT THE PROBE WAS FIRED AUTOMATICALLY WHEN THE PROBE IS IN SEMIAUTOMATIC MODE. NO COAXIAL NEEDLES WERE USED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23411896 ↗
  • MalfunctionApr 22, 2025Break; Crack; Break; Crack

    IT WAS IDENTIFIED BASED ON SAMPLE EVALUATION; THE PENETRATION DEPTH MARKER APPEARED TO BE DETACHED FROM THE DEVICE. IT WAS FURTHER IDENTIFIED THAT A CRACK WAS NOTED TO THE DEPTH ADJUSTMENT PLATE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21883876 ↗
  • MalfunctionJan 29, 2025Break; Crack; Difficult to Insert; Break; Crack; Difficult to Insert

    IT WAS REPORTED THAT DURING AN ULTRASOUND GUIDED BREAST BIOPSY THROUGH NORMAL TISSUE, THE NEEDLE OF THE PRODUCT ALLEGEDLY DIDN'T PENETRATE THE TISSUE PROPERLY. IT WAS FURTHER REPORTED THAT NOTCH LENGTH SELECTION PARTS OF PRODUCT HAD FALLEN OFF. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21261442 ↗
  • MalfunctionJan 10, 2025Crack; Difficult to Remove; Device-Device Incompatibility; Material Deformation; Material Twisted/Bent; Crack; Difficult to Remove; Device-Device Incompatibility; Material Deformation; Material Twisted/Bent

    IT WAS REPORTED THAT DURING A BREAST BIOPSY PROCEDURE, THE GUN WAS ALLEGEDLY NOT REMOVED FROM PATIENT AFTER SHOOTING. IT WAS FURTHER REPORTED THAT THE NEEDLE WAS COMPLETELY BENT FROM THE GUIDE NEEDLE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21126880 ↗
  • MalfunctionAug 9, 2024Break; Break

    IT WAS REPORTED THAT DURING A BIOPSY PROCEDURE, THE SMALL WHITE SQUARE PLASTIC PIECE HAS ALLEGEDLY BROKEN OFF. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19946996 ↗
  • MalfunctionAug 9, 2024Break; Break

    IT WAS REPORTED THAT DURING A BIOPSY PROCEDURE, THE SMALL WHITE SQUARE PLASTIC PIECE HAS ALLEGEDLY BROKEN OFF. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19946992 ↗
  • MalfunctionAug 7, 2024Nonstandard Device; Failure to Prime; Detachment of Device or Device Component; Nonstandard Device; Failure to Prime; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING A BIOPSY PROCEDURE, THE NEEDLE APERTURE ALLEGEDLY STUCK INSIDE THE OUTER CANNULA. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19935491 ↗
  • MalfunctionJul 17, 2024Self-Activation or Keying; Firing Problem; Self-Activation or Keying; Firing Problem

    IT WAS REPORTED THAT DURING A BIOPSY PROCEDURE, THE DOCTOR PRESSED SEMI-AUTOMATIC BUTTON BUT AUTOMATIC BUTTON WAS ALLEGEDLY WORKED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19768050 ↗
  • MalfunctionJun 6, 2024Nonstandard Device; Failure to Prime; Failure to Fire; Detachment of Device or Device Component; Nonstandard Device; Failure to Prime; Failure to Fire; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING AN ULTRASOUND-GUIDED BREAST BIOPSY PROCEDURE, THE DEVICE ALLEGEDLY FAILED TO SLIDE TO OPEN THE TROUGH TO REMOVE THE SAMPLE. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19474523 ↗
  • MalfunctionApr 15, 2024Failure to Prime; Device Contamination with Chemical or Other Material; Failure to Prime; Device Contamination with Chemical or Other Material

    IT WAS REPORTED THAT DURING A BIOPSY PROCEDURE, THE DEVICE ALLEGEDLY FAILED TO PRIME. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19103296 ↗
  • MalfunctionFeb 28, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING AN ULTRASOUND GUIDED BREAST BIOPSY THROUGH NORMAL SCATTERED DENSITY TISSUE, AS THE DEVICE WAS BEING REMOVED, THE NEEDLE OF THE DEVICE ALLEGEDLY KEPT GETTING LONGER AND DID NOT COMPLETELY COME OUT OF THE COAXIAL. IT WAS FURTHER REPORTED THAT THE CUTTING NEEDLE WITH THE STRIPS THAT COVERS THE TROUGH WAS ALLEGEDLY FOUND TO DETACHED FROM THE BIOPSY UNIT. REPORTEDLY, THE WHITE PORTION OF THE BIOPSY DEVICE AND THE END OF THE CUTTING NEEDLE WERE HELD IN ORDER TO PREVENT IT FROM COMPLETELY DETACHING. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18795345 ↗
  • MalfunctionFeb 6, 2024Unsealed Device Packaging; Tear, Rip or Hole in Device Packaging; Device Contamination with Chemical or Other Material; Packaging Problem; Unsealed Device Packaging; Tear, Rip or Hole in Device Packaging; Device Contamination with Chemical or Other Material; Packaging Problem

    IT WAS REPORTED THAT PRIOR TO A BIOPSY PROCEDURE, THERE WAS ALLEGEDLY A LOT OF WATER DAMAGE AND SOME OF THE NEEDLE TRAYS WERE DAMAGED. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18649060 ↗
  • MalfunctionFeb 6, 2024Unsealed Device Packaging; Tear, Rip or Hole in Device Packaging; Device Contamination with Chemical or Other Material; Packaging Problem; Unsealed Device Packaging; Tear, Rip or Hole in Device Packaging; Device Contamination with Chemical or Other Material; Packaging Problem

    IT WAS REPORTED THAT PRIOR TO A BIOPSY PROCEDURE, THERE WAS ALLEGEDLY A LOT OF WATER DAMAGE AND SOME OF THE NEEDLE TRAYS WERE DAMAGED. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18649058 ↗
  • MalfunctionFeb 6, 2024Unsealed Device Packaging; Tear, Rip or Hole in Device Packaging; Device Contamination with Chemical or Other Material; Packaging Problem; Unsealed Device Packaging; Tear, Rip or Hole in Device Packaging; Device Contamination with Chemical or Other Material; Packaging Problem

    IT WAS REPORTED THAT PRIOR TO A BIOPSY PROCEDURE, THERE WAS ALLEGEDLY A LOT OF WATER DAMAGE AND SOME OF THE NEEDLE TRAYS WERE DAMAGED. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18649057 ↗
  • MalfunctionFeb 6, 2024Unsealed Device Packaging; Tear, Rip or Hole in Device Packaging; Device Contamination with Chemical or Other Material; Packaging Problem; Unsealed Device Packaging; Tear, Rip or Hole in Device Packaging; Device Contamination with Chemical or Other Material; Packaging Problem

    IT WAS REPORTED THAT PRIOR TO A BIOPSY PROCEDURE, THERE WAS ALLEGEDLY A LOT OF WATER DAMAGE AND SOME OF THE NEEDLE TRAYS WERE DAMAGED. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18649056 ↗
  • MalfunctionFeb 6, 2024Unsealed Device Packaging; Tear, Rip or Hole in Device Packaging; Device Contamination with Chemical or Other Material; Packaging Problem; Unsealed Device Packaging; Tear, Rip or Hole in Device Packaging; Device Contamination with Chemical or Other Material; Packaging Problem

    IT WAS REPORTED THAT PRIOR TO A BIOPSY PROCEDURE, THERE WAS ALLEGEDLY A LOT OF WATER DAMAGE AND SOME OF THE NEEDLE TRAYS WERE DAMAGED. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18648327 ↗
  • MalfunctionFeb 6, 2024Unsealed Device Packaging; Tear, Rip or Hole in Device Packaging; Device Contamination with Chemical or Other Material; Packaging Problem; Unsealed Device Packaging; Tear, Rip or Hole in Device Packaging; Device Contamination with Chemical or Other Material; Packaging Problem

    IT WAS REPORTED THAT PRIOR TO A BIOPSY PROCEDURE, THERE WAS ALLEGEDLY A LOT OF WATER DAMAGE AND SOME OF THE NEEDLE TRAYS WERE DAMAGED. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18648325 ↗
  • MalfunctionFeb 6, 2024Unsealed Device Packaging; Tear, Rip or Hole in Device Packaging; Device Contamination with Chemical or Other Material; Packaging Problem; Unsealed Device Packaging; Tear, Rip or Hole in Device Packaging; Device Contamination with Chemical or Other Material; Packaging Problem

    IT WAS REPORTED THAT PRIOR TO A BIOPSY PROCEDURE, THERE WAS ALLEGEDLY A LOT OF WATER DAMAGE AND SOME OF THE NEEDLE TRAYS WERE DAMAGED. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18648324 ↗
  • MalfunctionOct 6, 2023Self-Activation or Keying; Self-Activation or Keying

    IT WAS REPORTED THAT DURING A BREAST BIOPSY PROCEDURE, THE DEVICE NEEDLE ALLEGEDLY FIRED INSIDE THE PATIENT BEFORE THE PHYSICIAN TOOK THE ACTION TO FIRE. IT WAS FURTHER REPORTED THAT THE NEEDLE SPONTANEOUSLY FIRED IN THE PATIENT'S BREAST AND WHEN REMOVED, THE SAMPLE NOTCH WAS OPEN. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17884585 ↗
  • MalfunctionOct 6, 2023Self-Activation or Keying; Self-Activation or Keying

    IT WAS REPORTED THAT DURING A BREAST BIOPSY PROCEDURE, THE DEVICE NEEDLE ALLEGEDLY FIRED INSIDE THE PATIENT BEFORE THE PHYSICIAN TOOK THE ACTION TO FIRE. IT WAS FURTHER REPORTED THAT THE NEEDLE SPONTANEOUSLY FIRED IN THE PATIENT'S BREAST AND WHEN REMOVED, THE SAMPLE NOTCH WAS OPEN. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17884584 ↗
  • MalfunctionOct 6, 2023Self-Activation or Keying; Self-Activation or Keying

    IT WAS REPORTED THAT DURING A BREAST BIOPSY PROCEDURE, THE DEVICE NEEDLE ALLEGEDLY FIRED INSIDE THE PATIENT BEFORE THE PHYSICIAN TOOK THE ACTION TO FIRE. IT WAS FURTHER REPORTED THAT THE NEEDLE SPONTANEOUSLY FIRED IN THE PATIENT'S BREAST AND WHEN REMOVED, THE SAMPLE NOTCH WAS OPEN. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17884582 ↗
  • MalfunctionOct 6, 2023Self-Activation or Keying; Self-Activation or Keying

    IT WAS REPORTED THAT DURING A BREAST BIOPSY PROCEDURE, THE DEVICE NEEDLE ALLEGEDLY FIRED INSIDE THE PATIENT BEFORE THE PHYSICIAN TOOK THE ACTION TO FIRE. IT WAS FURTHER REPORTED THAT THE NEEDLE SPONTANEOUSLY FIRED IN THE PATIENT'S BREAST AND WHEN REMOVED, THE SAMPLE NOTCH WAS OPEN. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17884577 ↗
  • MalfunctionOct 6, 2023Self-Activation or Keying; Self-Activation or Keying

    IT WAS REPORTED THAT DURING A BREAST BIOPSY PROCEDURE, THE DEVICE NEEDLE ALLEGEDLY FIRED INSIDE THE PATIENT BEFORE THE PHYSICIAN TOOK THE ACTION TO FIRE. IT WAS FURTHER REPORTED THAT THE NEEDLE SPONTANEOUSLY FIRED IN THE PATIENT'S BREAST AND WHEN REMOVED, THE SAMPLE NOTCH WAS OPEN. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17884511 ↗

Other INSTRUMENT, BIOPSY devices

All INSTRUMENT, BIOPSY devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.