Adverse event reports
4
Deaths reported
0
Injuries reported
3
Recalls naming it
1

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00801741097140), received between Jun 29, 2020 and Jun 3, 2021 — classified as 3 injuries, 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Recalls naming this device

  1. Jan 14, 2021Class IITerminated

    Due to the Penetration Depth Switch used to select the desired penetration depth of the biopsy needle for soft tissue biopsies, detaches from the device during use.

    Z-1289-2021 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem3
  • Break1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Hemorrhage/Blood Loss/Bleeding1
  • Pneumothorax1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 3, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING LUNG BIOPSY PROCEDURE, THE PATIENT ALLEGEDLY HAD PNEUMOTHORAX. PATIENT REPORTED STABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11929193 ↗
  • InjuryMay 12, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THROUGH ANONYMOUS POST-MARKET CLINICAL FOLLOW-UP SURVEY ON CORE BIOPSY INSTRUMENTS, THERE WAS ONE OCCURRENCE OF THORACOSTOMY TUBE PLACEMENT. THE CURRENT STATUS OF THE PATIENT WAS NOT PROVIDED.

    Patient problems: Pneumothorax

    MAUDE report 11810995 ↗
  • MalfunctionMar 16, 2021Break; Break

    IT WAS REPORTED THAT PRIOR TO A BIOPSY PROCEDURE, THE DEVICE BUTTON ALLEGEDLY FELL OFF. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11492235 ↗
  • InjuryJun 29, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED IN A POST-MARKET CLINICAL FOLLOW-UP SURVEY ON BIOPSIES, TWO PATIENTS EXPERIENCED HEMORRHAGE WITH COMPLICATIONS. THE CURRENT PATIENTS' STATUS IS UNKNOWN.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 10210118 ↗

FDA clearance and classification

The FDA regulates product code KNW (instrument, biopsy) under 21 CFR 876.1075 as a Class II device. 21 CFR 876.1075 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the MARQUEE MQ1413

Has the MARQUEE MQ1413 been recalled?

Yes. 1 FDA recall record names this device's identifier, and the FDA has closed (terminated) it. The recall, Z-1289-2021 (Class II), began on Jan 14, 2021: Due to the Penetration Depth Switch used to select the desired penetration depth of the biopsy needle for soft tissue biopsies, detaches from the device during use. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the MARQUEE MQ1413?

The FDA's MAUDE database lists 4 adverse event reports for this device, received between Jun 29, 2020 and Jun 3, 2021. In those reports, the most common specific device problem is “Break” (1); the most common patient problems are “Hemorrhage/Blood Loss/Bleeding” (1) and “Pneumothorax” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the MARQUEE MQ1413?

None of the reports are classified as deaths, 3 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the MARQUEE MQ1413, and is it still sold?

The FDA's device database lists Bard Peripheral Vascular, Inc. as the company responsible for it (brand MARQUEE, model MQ1413). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other INSTRUMENT, BIOPSY devices

All INSTRUMENT, BIOPSY devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page