MARQUEE MQK2016Recalls & reported problems
Bard Peripheral Vascular, Inc. · MARQUEE® Disposable Core Biopsy Instrument Kit, 20g x 16cm · Class 2
All models of MARQUEE (26)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Bard Marquee disposable core biopsy kit takes tissue samples through a needle during procedures. The 28 reports mostly describe the needle not fitting or getting stuck in the kit's coaxial introducer, sometimes requiring another device to finish; 2 mention a collapsed lung and there were no deaths. Two recalls name this device: an open one from August 2023 over a diameter mismatch between cannula and needle that can block reaching target tissue, and a closed one from January 2021 because the penetration depth switch detached during use.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 28
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 28 adverse event reports that list this exact device (DI 00801741097393), received between May 10, 2023 and Mar 4, 2024 — classified as 2 injuries, 26 malfunctions.
The most frequently reported specific device problem is “Device-Device Incompatibility” (27 reports).
2 FDA recall records name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Aug 10, 2023Class IIOpen, Classified
Diameter mismatch between coaxial and biopsy needle (cutting cannula) in disposable core biopsy instrument kits, where the coaxial cannula inner diameter prohibits the biopsy needle (cutting cannula) from properly fitting into the coaxial cannula and accessing target tissue, which may lead to repeated procedure, and procedural complications, such as pain/discomfort, pneumothorax, and bleeding.
Z-0027-2024 · Root cause: Device Design · Recall details · FDA record ↗
- Jan 14, 2021Class IITerminated
Due to the Penetration Depth Switch used to select the desired penetration depth of the biopsy needle for soft tissue biopsies, detaches from the device during use.
Z-1289-2021 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Device-Device Incompatibility27
- Adverse Event Without Identified Device or Use Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions26
- Pneumothorax2
Recent adverse event reports
The 25 most recent of 28 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 4, 2024Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT DURING A BIOPSY PROCEDURE THROUGH NORMAL DENSITY TISSUE, THE NEEDLE AND THE TROCAR IN THE KIT ALLEGEDLY HAD THE SAME DIAMETER AS THE BIOPSY NEEDLE ITSELF. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18825024 ↗MalfunctionMar 4, 2024Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT PRIOR TO A BIOPSY PROCEDURE THROUGH NORMAL DENSITY TISSUE, THE NEEDLE AND THE TROCAR IN THE KIT ALLEGEDLY HAD THE SAME DIAMETER AS THE BIOPSY NEEDLE ITSELF. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18824962 ↗MalfunctionAug 24, 2023Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT PRIOR TO A BIOPSY PROCEDURE, THE DEVICE NEEDLE ALLEGEDLY WOULD NOT PASS THROUGH THE COAXIAL INTRODUCER. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17610760 ↗MalfunctionAug 24, 2023Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT PRIOR TO A BIOPSY PROCEDURE, THE NEEDLE ALLEGEDLY GOT STUCK AND DID NOT FIT INTO THE DEVICE. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17610667 ↗MalfunctionAug 24, 2023Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT PRIOR TO A BIOPSY PROCEDURE, THE NEEDLE ALLEGEDLY GOT STUCK AND DID NOT FIT INTO THE DEVICE. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17610666 ↗MalfunctionAug 24, 2023Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT PRIOR TO A BIOPSY PROCEDURE, THE NEEDLE ALLEGEDLY GOT STUCK AND DID NOT FIT INTO THE DEVICE. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17610663 ↗MalfunctionAug 24, 2023Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT PRIOR TO A BIOPSY PROCEDURE, THE NEEDLE ALLEGEDLY GOT STUCK AND DID NOT FIT INTO THE DEVICE. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17610578 ↗MalfunctionAug 24, 2023Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT PRIOR TO A BIOPSY PROCEDURE, THE NEEDLE ALLEGEDLY GOT STUCK AND DID NOT FIT INTO THE DEVICE. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17610576 ↗MalfunctionAug 24, 2023Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT PRIOR TO A BIOPSY PROCEDURE, THE NEEDLE ALLEGEDLY GOT STUCK AND DID NOT FIT INTO THE DEVICE. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17610575 ↗MalfunctionAug 24, 2023Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT PRIOR TO A BIOPSY PROCEDURE, THE NEEDLE ALLEGEDLY GOT STUCK AND DID NOT FIT INTO THE DEVICE. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17610508 ↗MalfunctionAug 24, 2023Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT PRIOR TO A BIOPSY PROCEDURE, THE NEEDLE ALLEGEDLY DOES NOT FIT INTO THE COAXIAL. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17610425 ↗MalfunctionAug 24, 2023Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT PRIOR TO A BIOPSY PROCEDURE, THE NEEDLE ALLEGEDLY DOES NOT FIT INTO THE COAXIAL. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17610420 ↗MalfunctionAug 24, 2023Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT PRIOR TO A BIOPSY PROCEDURE, THE NEEDLE ALLEGEDLY DOES NOT FIT INTO THE COAXIAL. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17610419 ↗MalfunctionAug 24, 2023Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT PRIOR TO A BIOPSY PROCEDURE, THE NEEDLE ALLEGEDLY DOES NOT FIT INTO THE COAXIAL. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17610351 ↗MalfunctionAug 24, 2023Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT DURING THE INVESTIGATIONS OF THE LOT SAMPLES, THE COAXIAL OF THE DEVICE WAS IDENTIFIED TO BE NOT INCOMPATIBLE WITH THE NEEDLE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17610013 ↗MalfunctionAug 24, 2023Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT DURING THE INVESTIGATIONS OF THE LOT SAMPLES, THE COAXIAL OF THE DEVICE WAS IDENTIFIED TO BE NOT INCOMPATIBLE WITH THE NEEDLE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17610012 ↗MalfunctionAug 24, 2023Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT DURING THE INVESTIGATIONS OF THE LOT SAMPLES, THE COAXIAL OF THE DEVICE WAS IDENTIFIED TO BE NOT INCOMPATIBLE WITH THE NEEDLE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17610011 ↗MalfunctionAug 24, 2023Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT DURING THE INVESTIGATIONS OF THE LOT SAMPLES, THE COAXIAL OF THE DEVICE WAS IDENTIFIED TO BE NOT INCOMPATIBLE WITH THE NEEDLE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17610010 ↗MalfunctionAug 24, 2023Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT DURING THE INVESTIGATIONS OF THE LOT SAMPLES, THE COAXIAL OF THE DEVICE WAS IDENTIFIED TO BE NOT INCOMPATIBLE WITH THE NEEDLE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17610009 ↗MalfunctionAug 24, 2023Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT DURING THE INVESTIGATIONS OF THE LOT SAMPLES, THE COAXIAL OF THE DEVICE WAS IDENTIFIED TO BE NOT INCOMPATIBLE WITH THE NEEDLE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17610008 ↗MalfunctionAug 24, 2023Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT DURING THE INVESTIGATIONS OF THE LOT SAMPLES, THE COAXIAL OF THE DEVICE WAS IDENTIFIED TO BE NOT INCOMPATIBLE WITH THE NEEDLE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17610007 ↗MalfunctionAug 24, 2023Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT DURING THE INVESTIGATIONS OF THE LOT SAMPLES, THE COAXIAL OF THE DEVICE WAS IDENTIFIED TO BE NOT INCOMPATIBLE WITH THE NEEDLE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17610006 ↗MalfunctionAug 24, 2023Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT DURING THE INVESTIGATIONS OF THE LOT SAMPLES, THE COAXIAL OF THE DEVICE WAS IDENTIFIED TO BE NOT INCOMPATIBLE WITH THE NEEDLE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17609961 ↗MalfunctionAug 24, 2023Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT DURING THE INVESTIGATIONS OF THE LOT SAMPLES, THE COAXIAL OF THE DEVICE WAS IDENTIFIED TO BE NOT INCOMPATIBLE WITH THE NEEDLE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17609922 ↗MalfunctionAug 24, 2023Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT DURING THE INVESTIGATIONS OF THE LOT SAMPLES, THE COAXIAL OF THE DEVICE WAS IDENTIFIED TO BE NOT INCOMPATIBLE WITH THE NEEDLE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17609908 ↗
FDA clearance and classification
The FDA regulates product code KNW (instrument, biopsy) under 21 CFR 876.1075 as a Class II device. 21 CFR 876.1075 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the MARQUEE MQK2016
Has the MARQUEE MQK2016 been recalled?
Yes. 2 FDA recall records name this device's identifier: 1 still open and 1 closed (terminated) by the FDA. The most recent, Z-0027-2024 (Class II), began on Aug 10, 2023: Diameter mismatch between coaxial and biopsy needle (cutting cannula) in disposable core biopsy instrument kits, where the coaxial cannula inner diameter prohibits the biopsy needle (cutting cannula) from properly… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the MARQUEE MQK2016?
The FDA's MAUDE database lists 28 adverse event reports for this device, received between May 10, 2023 and Mar 4, 2024. In those reports, the most common specific device problem is “Device-Device Incompatibility” (27); the most common patient problem is “Pneumothorax” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the MARQUEE MQK2016?
None of the reports are classified as deaths, 2 as injuries and 26 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the MARQUEE MQK2016, and is it still sold?
The FDA's device database lists Bard Peripheral Vascular, Inc. as the company responsible for it (brand MARQUEE, model MQK2016). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other INSTRUMENT, BIOPSY devices
- MAX-CORE MC1616Bard Peripheral Vascular, Inc. · INSTRUMENT, BIOPSY1 recall4,403 reports
- ENCOR® ECP017GSenorx, Inc. · INSTRUMENT, BIOPSY1 recall4,186 reports
- MAX-CORE MC1820Bard Peripheral Vascular, Inc. · INSTRUMENT, BIOPSY1 recall2,405 reports
- MAX-CORE MC1816Bard Peripheral Vascular, Inc. · INSTRUMENT, BIOPSY1 recall1 death853 reports
- MAX-CORE MC1410Bard Peripheral Vascular, Inc. · INSTRUMENT, BIOPSY1 recall402 reports
- MAX-CORE MC1610Bard Peripheral Vascular, Inc. · INSTRUMENT, BIOPSY1 recall394 reports
- ENCOR® ECP0110GVSenorx, Inc. · INSTRUMENT, BIOPSY392 reports
- ENCOR® ECP0110GSenorx, Inc. · INSTRUMENT, BIOPSY390 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page