SureStep™ Foley Tray System Bard® Lubricath® Foley Catheter Tray A899616Recalls & reported problems
C. R. Bard, Inc. · SureStep® Foley Tray System Bard Lubricath Foley Catheter Tray - 2000 ml Drainage Bag · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00801741104534), received between Oct 24, 2019 and Feb 26, 2020 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Gel Leak” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Jun 25, 2026Class IIOpen, Classified
Due to complaints received indicating presence of contamination on the surface of pericare wipes provided with various Foley and Intermittent Catheter Trays.
Z-3114-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Gel Leak1
- Material Discolored1
- Short Fill1
- Use of Device Problem1
Patient problems
- No Consequences Or Impact To Patient2
- Device Embedded In Tissue or Plaque1
- Discomfort1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 26, 2020Material Discolored; Gel Leak; Short Fill; Gel Leak
IT WAS REPORTED THAT THE SYRINGE WAS SHORT FILLED, AND WAS DARK YELLOW AND THICK.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 9757093 ↗MalfunctionOct 24, 2019Use of Device Problem; Adverse Event Without Identified Device or Use Problem; Use of Device Problem
IT WAS REPORT THAT A BETADINE SWAB STICK FROM A FOLEY PREP KIT WAS LEFT IN A PATIENT POST SURGERY. AT THE START OF A SURGERY, AN OR NURSE WAS PREPARING TO INSERT A FOLEY CATHETER INTO AN OBESE FEMALE PATIENT. THE NURSE REPORTED DIFFICULTY VISUALIZING THE URETHRA. AFTER THE NURSE PREPPED THE SITE, SHE PLACED ONE OF THE BETADINE PREP SWABS INTO THE PATIENT'S VAGINA TO ACT AS A LANDMARK. THE PROCEDURE WAS COMPLETED AND THE PATIENT WAS DISCHARGE HOME. POST OPERATION DAY ONE, THE PATIENT CALL THE POST OPERATIVE UNIT COMPLAINING OF SOME VAGINAL PRESSURE. THEN, WHILE SHE WAS URINATING SHE PULLED A SMALL HARD PLASTIC PIECE OUT OF HER VAGINA. THE PATIENT WAS INSTRUCTED TO GO TO HER PHYSICIANS OFFICE. IT WAS DETERMINED THAT THE ITEM REMOVED WAS A BETADINE SWAB STICK FROM A FOLEY PREP KIT . THE PATIENT WAS NOT INJURED. UPON THE INVESTIGATION BY THE HOSPITAL IT WAS DISCOVERED THAT THIS WAS A COMMON PRACTICE AMONG NURSES EVEN THOUGH IT WAS NOT THE INTENDED USE OF THE BETADINE PREP SWAB IN THE FOLEY KIT. THE FACILITY WANTED TO ALERT THE MANUFACTURE TO CONSIDER PUTTING A WARNING LABEL ON PACKAGE OR ALTER THE DESIGN OF THE PREP STICK TO PREVENT IT FROM BEING PLACED IN THE VAGINA.
Patient problems: Device Embedded In Tissue or Plaque; Discomfort; No Consequences Or Impact To Patient
MAUDE report 9233325 ↗
Other CATHETER, COUDE devices
- Bardex® Lubricath® Foley Catheter 0165L14C. R. Bard, Inc. · CATHETER, COUDE562 reports
- Bardex® Lubricath® Foley Catheter 0165L16C. R. Bard, Inc. · CATHETER, COUDE539 reports
- Bardia® Foley Catheter Silicone Coated 123516AC. R. Bard, Inc. · CATHETER, COUDE381 reports
- Bardia® Foley Catheter Silicone Coated 123514AC. R. Bard, Inc. · CATHETER, COUDE195 reports
- SureStep™ Foley Tray System Bardex® Lubricath® Foley Catheter A899916C. R. Bard, Inc. · CATHETER, COUDE3 recalls19 reports
- Bardia® Foley Catheter Silicone Coated 123518AC. R. Bard, Inc. · CATHETER, COUDE153 reports
- SureStep™ Foley Tray System Bardex® Lubricath® Foley Catheter A902916C. R. Bard, Inc. · CATHETER, COUDE2 recalls50 reports
- Bard® Lubricath® Foley Catheter 365716C. R. Bard, Inc. · CATHETER, COUDE117 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page