Skip to content

Products that contain the Pericare Wipes recallZ-3114-2026 · C.R. Bard Inc · initiated Jun 25, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from C.R. Bard Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Due to complaints received indicating presence of contamination on the surface of pericare wipes provided with various Foley and Intermittent Catheter Trays.

FDA root cause category: Under Investigation by firm

What the company is doing

On 06/25/2026, an "URGENT Medical Device Recall" Letter was via regional courier services or other electronic means to inform customers that based on customer feedback, BD has identified that the CliniClean Pericare Wipes provided within Urinary Catheter Trays may be potentially contaminated with an unknown non-biological substance. Customer are instructed to: Do not use CliniClean Pericare Wipes contained within the affected products listed in Attachment 2. The CliniClean Pericare Wipes are the only impacted component within the tray and all remaining components can continue to be safely used. Users should source alternative, cleansing wipes for the impacted product lots listed in Attachment 2 and ensure they continue to follow their facility protocols for providing perineal care for patients with urinary catheters. If affected CliniClean Pericare Wipes have been used without issue, no additional treatment or clinical follow up is required. 1. For affected lots, remove and destroy wipes within the product package at the point of use and do not use. Source alternative cleansing wipes for all impacted urinary catheter trays listed in Attachment 2. 2. Share this notice with any users of the product within your facilities to ensure they are also aware of this Medical Device Correction. 3. Post this notice in all locations where inventory of these trays are stored. 4. Complete and return the attached Customer Response Form to the BD contact noted on the form confirming acknowledgement of the correction notification, whether or not you have any affected product, so that BD may confirm your receipt of this notification. For Questions - contact: BD Medical Information Services: Available 24/7 Select #2 when prompted 1-800-555-7422 or Email: [email protected] BD Technical Support 1-844-8BD-LIFE (1-844-823-5433) Say Product Complaint when prompted M-F 8am - 5pm CT or Email: [email protected]

Recall details

Recall number
Z-3114-2026
Status
Open, Classified
Initiated by the company
Jun 25, 2026
Posted by the FDA
Sep 4, 2026
Recalling firm
C.R. Bard Inc
Quantity
12,463 trays/kits
Product code
EZC
FDA submissions
K910197, K910846
Distribution
Worldwide - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI and WY. The countries of Bahrain, Canada, Costs Rica, Guatemala, Mexico, Saudi Arabia, and United Arab Emirates.

Product description

Products that contain the Pericare Wipes: Bardex Lubricath Foley Catheter Tray REF: 899914 899916 899918 UDI codes: 00801741024801 10801741024808 00801741024818 10801741024815 00801741024825 10801741024822 SureStep" Foley Tray System with Bardex Lubricath Temp Sensing Foley Catheter REF: A119416M UDI code: 00801741073700 10801741073707 SureStep" Foley Tray System Bard Lubricath Foley REF: A800361 UDI code: 00801741073939 10801741073936 SureStep" Foley Tray System Bard Lubricath Foley Catheter Tray REF: A899616 UDI code: 00801741104534 10801741104531 SureStep" Foley Tray System Bardex Lubricath Foley Catheter REF: A899914 UDI code: 00801741074004 10801741074001 SureStep" Foley Tray System Bardex Lubricath Foley Catheter REF: A899916 UDI code: 00801741074011 10801741074018 SureStep" Foley Tray System Bardex Lubricath Temperature Sensing Catheter REF: A900216 UDI code: 00801741120978 10801741120975 SureStep" Foley Tray System BardLubricath Foley Catheter REF: A902914 A902916 A902918 UDI code: 0080174107406610801741074063 00801741074073 10801741074070 0080174107408010801741074087 Bard Urine Meter Foley Catheter Tray With Bardex Lubricath Foley Catheter REF: 304916 UDI code: 00801741019736 01801741019739 Bard Urine Meter Foley Catheter Tray With Bardex Lubricath Foley Catheter REF: 902916 UDI codes: 00801741024863 10801741024860 (416A) SureStep Foley Tray System Bardex IC Complete Care (714A/716A/718A)SureStep Foley Tray System Bardex IC Complete REF: A304914 A304916 A304918 UDI codes: 00801741104350 10801741104357 00801741104367 10801741104364 00801741100499 10801741100496 SureStep" Foley Tray System Bard Lubricath Foley REF: A800365 UDI codes: 00801741073960 10801741073967 SureStep" Foley Tray System Bardex Lubricath Foley Catheter REF: A899918 UDI codes: 00801741074028 10801741074025 SureStep" Foley Tray System BardLubricath Foley Catheter REF: A902618 UDI codes: 00801741104640 10801741104647 SureStep" Foley Tray System BardLubricath Foley Catheter REF: A902616 UDI codes: 00801741104633 10801741104630 SureStep Intermittent Catheter Tray REF: INTL16C UDI codes: 00801741183195 10801741183192 For urological use only, Intended for use in the drainage and/or collection and/or measurement of urine.

Code information (models, lots, serial numbers)

Show the full code information (2k characters)

Catalog Numbers / UDI codes: 899914 / 00801741024801¿ 10801741024808 Lot Number: NGKU4611 899916 / 00801741024818¿ 10801741024815 Lot Numbers: NGKU2448 NGKU4610 899918 / 00801741024825¿ 10801741024822 Lot Number: NGKU3555 ***** SureStep" Foley Tray System with Bardex¿ Lubricath Temp Sensing Foley Catheter REF: A119416M UDI code: 00801741073700 10801741073707 SureStep" Foley Tray System Bard¿ Lubricath Foley REF: A800361 UDI code: 00801741073939 10801741073936 SureStep" Foley Tray System Bard¿ Lubricath¿ Foley Catheter Tray REF: A899616 UDI code: 00801741104534¿ 10801741104531 SureStep" Foley Tray System Bardex¿ Lubricath¿ Foley Catheter REF: A899914 UDI code: 00801741074004 10801741074001 SureStep" Foley Tray System Bardex¿ Lubricath¿ Foley Catheter REF: A899916 UDI code: 00801741074011¿ 10801741074018 SureStep" Foley Tray System Bardex¿ Lubricath¿ Temperature Sensing Catheter REF: A900216 UDI code: 00801741120978¿ 10801741120975 SureStep" Foley Tray System Bard¿Lubricath¿ Foley Catheter REF: A902914 A902916 A902918 UDI code: 00801741074066¿10801741074063 00801741074073 10801741074070 00801741074080¿10801741074087 Bard¿ Urine Meter Foley Catheter Tray With Bardex¿ Lubricath¿ Foley Catheter REF: 304916 UDI code: 00801741019736 01801741019739 Bard¿ Urine Meter Foley Catheter Tray With Bardex¿ Lubricath¿ Foley Catheter REF: 902916 UDI codes: 00801741024863 10801741024860 (416A) SureStep Foley Tray System Bardex IC Complete Care (714A/716A/718A)SureStep Foley Tray System Bardex IC Complete REF: A304914 A304916 A304918 UDI codes: 00801741104350¿ 10801741104357 00801741104367¿ 10801741104364 00801741100499¿ 10801741100496 SureStep" Foley Tray System Bard¿ Lubricath Foley REF: A800365 UDI codes: 00801741073960 10801741073967 SureStep" Foley Tray System Bardex¿ Lubricath¿ Foley Catheter REF: A899918 UDI codes: 00801741074028 10801741074025 SureStep" Foley Tray System Bard¿Lubricath¿ Foley Catheter REF: A902618 UDI codes: 00801741104640 10801741104647 SureStep" Foley Tray System Bard¿Lubricath¿ Foley Catheter REF: A902616 UDI codes: 00801741104633¿ 10801741104630 SureStep Intermittent Catheter Tray REF: INTL16C UDI codes: 00801741183195¿ 10801741183192

Devices named in this recall

21 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 98840).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page