Skip to content

COVERA AVSM06100Recalls & reported problems

Angiomed GmbH & Co. Medizintechnik KG · COVERA™ Vascular Covered Stent 6 mm x 100 mm (80 cm delivery catheter) (straight) · Class 3 · Implant

All models of COVERA (65)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00801741106217), received between Nov 21, 2025 and Sep 11, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Dec 22, 2022Open, Classified

    An increase in the reported complaint rate for inability to deploy was identified and related to poor slide block bonding.

    Z-1108-2023 · Root cause: Unknown/Undetermined by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Break2
  • Positioning Failure2
  • Loss of or Failure to Bond1
  • Misfire1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 11, 2026Loss of or Failure to Bond; Break; Positioning Failure

    On (b)(6) 2026, a 53-year-old male patient underwent a stent placement procedure using a Covera Vascular Covered Stent for the treatment of severe stenosis at the juxta-anastomotic region of a left radiocephalic arteriovenous fistula (AVF) via the left radiocephalic access site. During the procedure, it was reported that the stent failed to deploy, as the sheath did not retract when the thumbwheel was rotated. It was further reported that the internal wire did not respond to thumbwheel manipulation. The device was successfully retrieved and different brand of stent used and deployed successfully. There was no reported patient injury.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50726391 ↗
  • MalfunctionNov 21, 2025Break; Positioning Failure; Misfire; Break; Positioning Failure; Misfire

    ON (B)(6) 2025, A PATENT UNDERWENT A STENT GRAFT PLACEMENT PROCEDURE USING COVERA VASCULAR COVERED STENT IN THE ACCESS SITE OF COMMON FEMORAL ARTERY APPROACH FROM THE CONTRALATERAL SIDE. DURING THE PROCEDURE THE COVERA STENT GRAFT ALLEGEDLY FAILED TO DEPLOY. THE PROCEDURE WAS COMPLETED BY USING ANOTHER DEVICE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23615225 ↗

Other System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment devices

All System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment devices →