COVERA AVSM07080Recalls & reported problems
Angiomed GmbH & Co. Medizintechnik KG · COVERA™ Vascular Covered Stent 7 mm x 80 mm (80 cm delivery catheter) (straight) · Class 3 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00801741106255), received between Jul 14, 2020 and Oct 8, 2024 — classified as 1 injury, 7 malfunctions.
The most frequently reported specific device problem is “Positioning Failure” (3 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Dec 22, 2022Open, Classified
An increase in the reported complaint rate for inability to deploy was identified and related to poor slide block bonding.
Z-1108-2023 · Root cause: Unknown/Undetermined by firm · FDA record ↗
Most reported problems
Device problems
- Positioning Failure3
- Break2
- Loss of or Failure to Bond2
- Malposition of Device2
- Misfire2
- Adverse Event Without Identified Device or Use Problem1
- Material Deformation1
- Obstruction of Flow1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- No Consequences Or Impact To Patient2
- Restenosis1
- Thrombosis/Thrombus1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 8, 2024Malposition of Device; Malposition of Device
IT WAS REPORTED THAT DURING A STENT GRAFT PLACEMENT PROCEDURE IN THE RIGHT UPPER EXTREMITY ARTERIOVENOUS GRAFT STENOSIS VIA RIGHT UPPER ARM, THE STENT ALLEGEDLY JUMPED. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20394996 ↗MalfunctionJul 22, 2024Loss of or Failure to Bond; Positioning Failure; Misfire; Loss of or Failure to Bond; Positioning Failure; Misfire
IT WAS REPORTED THAT DURING A STENT GRAFT PLACEMENT PROCEDURE IN THE LEFT ARM FISTULA GRAFT, THE DEVICE ALLEGEDLY HAD DID NOT DEPLOY. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19792330 ↗MalfunctionMay 27, 2023Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging
IT WAS REPORTED THAT DURING THE PREPARATION OF A STENT PLACEMENT PROCEDURE, THE INNER BOX ALLEGEDLY HAD A HOLE IN IT. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17015403 ↗MalfunctionNov 1, 2022Break; Positioning Failure; Misfire; Break; Positioning Failure; Misfire
IT WAS REPORTED THAT DURING A STENT PLACEMENT PROCEDURE IN RIGHT THIGH GRAFT, THE CABLE ALLEGEDLY BROKEN. IT WAS FURTHER REPORTED THAT THE STENT ALLEGEDLY FAILED TO DEPLOY. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15706379 ↗InjuryFeb 23, 2022Obstruction of Flow; Adverse Event Without Identified Device or Use Problem; Obstruction of Flow; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THROUGH THE RESULTS OF A CLINICAL TRIAL, THAT APPROXIMATELY THREE MONTHS POST INDEX PROCEDURE, THE SUBJECT DEVELOPED SEVENTY PERCENT OF TARGET LESION STENOSIS. STANDARD PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) WAS USED TO SUCCESSFULLY TREAT THE TARGET LESION. THEN, APPROXIMATELY SIX MONTHS AND FIFTEEN DAYS POST INDEX PROCEDURE, THROMBOSIS WAS DIAGNOSED. STANDARD PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) AND THROMBECTOMY/THROMBOLYSIS WAS USED TO SUCCESSFULLY TREAT THE TARGET LESION. THE CURRENT PATIENT STATUS IS NOT PROVIDED.
Patient problems: Restenosis; Thrombosis/Thrombus
MAUDE report 13583069 ↗MalfunctionSep 30, 2021Loss of or Failure to Bond; Break; Positioning Failure; Loss of or Failure to Bond; Break; Positioning Failure
IT WAS REPORTED THAT DURING A STENT PLACEMENT PROCEDURE, THE DEVICE ALLEGEDLY FAILED TO DEPLOY. THE DEVICE WAS REMOVED AND THE STENT WAS TRIED TO BE DEPLOYED MANUALLY BUT THE STRING INSIDE THE SYSTEM WAS NOT FOUND TO BE ATTACHED TO THE HANDLE. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12555238 ↗MalfunctionJul 14, 2020Malposition of Device; Malposition of Device
THIS REPORT SUMMARIZES ONE MALFUNCTION. A REVIEW OF THE MALFUNCTION INDICATED THAT MODEL AVSM07080 VASCULAR COVERED STENT WAS ALLEGEDLY EXPERIENCED MALPOSITION OF DEVICE. THIS REPORT WAS RECEIVED FROM ONE SOURCE. THE EVENT INVOLVED A PATIENT WITH NO KNOWN IMPACT TO THE PATIENT. THE (B)(6) MALE PATIENT WAS (B)(6).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10271441 ↗MalfunctionJul 14, 2020Material Deformation; Material Deformation
THIS REPORT SUMMARIZES ONE MALFUNCTION. A REVIEW OF THE REPORTED INFORMATION INDICATED THAT MODEL AVSM07080 VASCULAR STENT GRAFT ALLEGEDLY EXPERIENCED MATERIAL DEFORMATION. THIS INFORMATION WAS RECEIVED FROM ONE SOURCE. THE EVENT INVOLVED A PATIENT WITH NO KNOWN IMPACT TO THE PATIENT. THE AGE, WEIGHT AND GENDER OF THE PATIENT WERE NOT PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10271440 ↗
Other System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment devices
- COVERA AVSM10060Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall33 reports
- COVERA AVSM08060Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall21 reports
- COVERA AVSM09060Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall18 reports
- COVERA AVSM10080Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall16 reports
- COVERA AVSM10040Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall15 reports
- COVERA AVFM10060Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall14 reports
- COVERA AVSM09080Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall1 death14 reports
- COVERA AVSM10100Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall13 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page