COVERA AVSM07100Recalls & reported problems
Angiomed GmbH & Co. Medizintechnik KG · COVERA™ Vascular Covered Stent 7 mm x 100 mm (80 cm delivery catheter) (straight) · Class 3 · Implant
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00801741106262), received between Oct 16, 2019 and Feb 7, 2023 — classified as 2 injuries, 4 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- Dec 22, 2022Open, Classified
An increase in the reported complaint rate for inability to deploy was identified and related to poor slide block bonding.
Z-1108-2023 · Root cause: Unknown/Undetermined by firm · FDA record ↗
Most reported problems
Device problems
- Break2
- Loss of or Failure to Bond2
- Material Deformation2
- Misfire2
- Activation Failure1
- Detachment of Device or Device Component1
- Fracture1
- Malposition of Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Injury1
- No Consequences Or Impact To Patient1
- Stenosis1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 7, 2023Premature Activation; Malposition of Device; Premature Activation; Malposition of Device
IT WAS REPORTED THAT DURING A STENT GRAFT PLACEMENT PROCEDURE VIA DISTAL ACCESS TO TREAT THROMBOSED ARTERIOVENOUS GRAFT, THE STENT GRAFT ALLEGEDLY JUMPED SEVERAL CENTIMETER FORWARD WHICH RESULTED IN MALPOSITION OF DEVICE DURING DEPLOYMENT AND FAILED. IT WAS FURTHER REPORTED THAT, THE SECOND STENT WAS PLACED TO OVERLAP THE FIRST STENT GRAFT. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16316153 ↗MalfunctionJun 15, 2022Loss of or Failure to Bond; Break; Misfire; Loss of or Failure to Bond; Break; Misfire
IT WAS REPORTED THAT DURING A STENT GRAFT PLACEMENT, THE DEVICE ALLEGEDLY PARTIALLY DEPLOYED. IT WAS FURTHER REPORTED THAT THE DEPLOYMENT SYSTEM BROKE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14692402 ↗MalfunctionJun 15, 2022Loss of or Failure to Bond; Break; Misfire; Detachment of Device or Device Component; Material Deformation; Loss of or Failure to Bond; Break; Misfire; Detachment of Device or Device Component; Material Deformation
IT WAS REPORTED THAT DURING A STENT GRAFT PLACEMENT, THE DEVICE ALLEGEDLY PARTIALLY DEPLOYED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14692374 ↗InjuryMar 12, 2020Obstruction of Flow; Obstruction of Flow
IT WAS REPORTED THROUGH THE RESULTS OF A CLINICAL TRIAL APPROXIMATELY 8-MONTHS POST INDEX PROCEDURE, 50% TARGET LESION STENOSIS AND STANDARD PTA WAS PERFORMED TO TREAT LESION. IT WAS FURTHER REPORTED APPROXIMATELY 1 YEAR 11-MONTH POST INDEX PROCEDURE 70% STENOSIS WAS IDENTIFIED AND STANDARD PTA WAS PERFORMED TO TREAT LESION.
Patient problems: Stenosis
MAUDE report 9824446 ↗MalfunctionJan 13, 2020Material Deformation; Activation Failure; Material Deformation; Activation Failure
THIS REPORT SUMMARIZES ONE MALFUNCTION. A REVIEW OF THE REPORTED INFORMATION INDICATED THAT MODEL AVSM07100 VASCULAR COVERED STENT EXPERIENCED FAILURE TO EXPAND. THIS INFORMATION WAS RECEIVED FROM ONE SOURCE. THE EVENT INVOLVED A PATIENT WITH NO KNOWN IMPACT TO THE PATIENT. THE PATIENT AGE, WEIGHT, AND GENDER WERE NOT PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9583847 ↗InjuryOct 16, 2019Fracture; Fracture
IT WAS REPORTED THAT SOME TIME POST STENT GRAFT PLACEMENT THE STENT GRAFT ALLEGEDLY FRACTURED AND A PSEUDO ANEURYSM WAS IDENTIFIED. REPORTEDLY, THE PATIENT ALLEGEDLY DEVELOPED COMPARTMENT SYNDROME AND A FASCIOTOMY AND STENT REPAIR WERE PERFORMED. THERE WAS NO REPORTED PATIENT INJURY POST STENT GRAFT REPAIR PROCEDURE.
Patient problems: Injury
MAUDE report 9200239 ↗
Other System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment devices
- COVERA AVSM10060Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall33 reports
- COVERA AVSM08060Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall21 reports
- COVERA AVSM09060Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall18 reports
- COVERA AVSM10080Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall16 reports
- COVERA AVSM10040Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall15 reports
- COVERA AVFM10060Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall14 reports
- COVERA AVSM09080Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall1 death14 reports
- COVERA AVSM10100Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall13 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.