COVERA AVSM08040Recalls & reported problems
Angiomed GmbH & Co. Medizintechnik KG · COVERA™ Vascular Covered Stent 8 mm x 40 mm (80 cm delivery catheter) (straight) · Class 3 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 6
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00801741106286), received between Jul 15, 2020 and Aug 27, 2025 — classified as 6 injuries, 4 malfunctions.
The most frequently reported specific device problem is “Misfire” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Dec 22, 2022Open, Classified
An increase in the reported complaint rate for inability to deploy was identified and related to poor slide block bonding.
Z-1108-2023 · Root cause: Unknown/Undetermined by firm · FDA record ↗
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem5
- Misfire2
- Obstruction of Flow2
- Activation Failure1
- Malposition of Device1
- Material Deformation1
- Packaging Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Restenosis2
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Ischemia1
- No Patient Involvement1
- Obstruction/Occlusion1
- Pain1
- Stenosis1
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 27, 2025Malposition of Device; Malposition of Device
ON (B)(6) 2025, A PATIENT UNDERWENT A STENT PLACEMENT PROCEDURE USING THE COVERA VASCULAR COVERED STENT VIA BASILIC VEIN. THE STENT PREMATURELY EXITED THE DELIVERY SHEATH WHEN IT WAS APPROXIMATELY HALFWAY DEPLOYED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22901498 ↗InjuryMar 14, 2025Obstruction of Flow; Obstruction of Flow
IT WAS REPORTED THAT SOMETIME POST A STENT GRAFT PLACEMENT, THE STENT HAD ALLEGEDLY CAUSED OCCLUSION IN THE PATIENT. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THE CURRENT STATUS OF THE PATIENT IS UNKNOWN.
Patient problems: Obstruction/Occlusion
MAUDE report 21603292 ↗MalfunctionFeb 28, 2025Misfire; Material Deformation; Activation Failure; Misfire; Material Deformation; Activation Failure
IT WAS REPORTED THAT DURING A STENT GRAFT PLACEMENT PROCEDURE, THE DISTAL END OF THE STENT ALLEGEDLY DID NOT OPEN UPON DEPLOYMENT. IT WAS FURTHER REPORTED THAT THE STENT WAS BALLOONED WHICH IMPROVED THE RESULT BUT WAS STILL NOT COMPLETELY OPEN. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21488333 ↗InjuryJul 28, 2023Obstruction of Flow; Adverse Event Without Identified Device or Use Problem; Obstruction of Flow; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING A STENT PLACEMENT PROCEDURE IN THE FISTULA VIA FISTULA VENOUS ACCESS, THE STENT WAS ALLEGEDLY PLACED AND IMMEDIATELY CLOTTED THE ENTIRE ACCESS. THE CURRENT STATUS OF THE PATIENT WAS UNKNOWN.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Thrombosis/Thrombus
MAUDE report 17412451 ↗MalfunctionJan 23, 2023Misfire; Misfire
IT WAS REPORTED THAT DURING A STENT GRAFT PLACEMENT PROCEDURE, THE STENT GRAFT ALLEGEDLY PARTIALLY DEPLOYED. IT WAS FURTHER REPORTED THAT THE STENT WAS REMOVED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16222628 ↗InjuryApr 25, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THROUGH THE RESULTS OF A CLINICAL TRIAL, THAT APPROXIMATELY ELEVEN MONTHS POST INDEX PROCEDURE, STENOSIS IN TARGET LESION WAS OBSERVED AND STANDARD PTA WAS USED TO SUCCESSFULLY TREAT THE TARGET LESION. THE CURRENT STATUS OF THE SUBJECT WAS NOT PROVIDED.
Patient problems: Restenosis
MAUDE report 14192377 ↗InjuryOct 19, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THROUGH THE RESULTS OF A CLINICAL TRIAL, THAT APPROXIMATELY SIX MONTHS POST INDEX PROCEDURE, THE SUBJECT DEVELOPED 80% OF TARGET LESION STENOSIS. STANDARD PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) WAS USED TO SUCCESSFULLY TREAT THE TARGET LESION. THE CURRENT PATIENT STATUS WAS NOT PROVIDED.
Patient problems: Restenosis
MAUDE report 12653182 ↗InjurySep 14, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THROUGH THE RESULTS OF A CLINICAL TRIAL, THAT POST INDEX PROCEDURE, THE SUBJECT DEVELOPED LEFT SHOULDER PAIN. APPROXIMATELY SIX MONTHS POST INDEX PROCEDURE, THE SUBJECT DEVELOPED 100% OF TARGET LESION STENOSIS. STANDARD PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) WAS USED TO SUCCESSFULLY TREAT THE TARGET LESION. THE CURRENT PATIENT STATUS WAS NOT PROVIDED.
Patient problems: Pain; Stenosis
MAUDE report 12466022 ↗InjuryAug 26, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THROUGH THE RESULTS OF A CLINICAL TRIAL, THAT APPROXIMATELY TWO YEARS AND FIVE MONTHS POST INDEX PROCEDURE, NEW LESION STENOSIS WAS OBSERVED AND COVERA VASCULAR COVERED STENT WAS USED TO SUCCESSFULLY TREAT THE TARGET LESION. THE PATIENT WAS DIAGNOSED WITH RIGHT HAND ISCHEMIA RELATED TO STUDY PROCEDURE AND DEVICE. THE PATIENT CURRENT STATUS IS UNKNOWN.
Patient problems: Ischemia
MAUDE report 12369763 ↗MalfunctionJul 15, 2020Packaging Problem; Packaging Problem
THIS REPORT SUMMARIZES ONE MALFUNCTION. A REVIEW OF THE REPORTED INFORMATION INDICATED THAT MODEL AVSM08040.A COVERA VASCULAR COVERED STENT SYSTEM ALLEGEDLY EXPERIENCED PACKAGING PROBLEM. THE MALFUNCTION HAD NO PATIENT INVOLVEMENT AND THE PATIENT HAD NO CONSEQUENCES. THE PATIENT WAS A (B)(6) YEAR OLD FEMALE; HOWEVER, THE WEIGHT WAS NOT PROVIDED.
Patient problems: No Patient Involvement
MAUDE report 10276338 ↗
Other System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment devices
- COVERA AVSM10060Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall33 reports
- COVERA AVSM08060Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall21 reports
- COVERA AVSM09060Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall18 reports
- COVERA AVSM10080Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall16 reports
- COVERA AVSM10040Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall15 reports
- COVERA AVFM10060Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall14 reports
- COVERA AVSM09080Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall1 death14 reports
- COVERA AVSM10100Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall13 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page