COVERA AVSM08080Recalls & reported problems
Angiomed GmbH & Co. Medizintechnik KG · COVERA™ Vascular Covered Stent 8 mm x 80 mm (80 cm delivery catheter) (straight) · Class 3 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00801741106309), received between Jul 14, 2020 and Aug 5, 2025 — classified as 2 injuries, 5 malfunctions.
The most frequently reported specific device problem is “Obstruction of Flow” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Dec 22, 2022Open, Classified
An increase in the reported complaint rate for inability to deploy was identified and related to poor slide block bonding.
Z-1108-2023 · Root cause: Unknown/Undetermined by firm · FDA record ↗
Most reported problems
Device problems
- Obstruction of Flow2
- Activation Failure1
- Adverse Event Without Identified Device or Use Problem1
- Difficult to Remove1
- Material Deformation1
- Material Separation1
- Misfire1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Restenosis2
- Hemorrhage/Blood Loss/Bleeding1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 5, 2025Misfire; Misfire
ON (B)(6) 2025, A PATIENT UNDERWENT STENT GRAFT PLACEMENT PROCEDURE USING THE COVERA VASCULAR COVERED STENT IN THE RIGHT MEDIAN CUBITAL VEIN VIA FISTULA. DURING THE PROCEDURE, THE STENT PLACED IN THE VESSEL ALLEGEDLY DEPLOYED PARTIALLY. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22711183 ↗MalfunctionFeb 25, 2025Activation Failure; Activation Failure
IT WAS REPORTED THAT DURING A STENT PLACEMENT PROCEDURE, THE STENT WAS ALLEGEDLY NOT DEPLOYED. IT WAS FURTHER REPORTED THAT THE STENT WAS FOLDED ON THE DISTAL SIDE. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21456927 ↗MalfunctionMar 19, 2024Difficult to Remove; Difficult to Remove
IT WAS REPORTED THAT DURING A STENT GRAFT PLACEMENT PROCEDURE IN THE CEPHALIC ARCH VIA THE LEFT BRACHIAL CEPHALIC FISTULA, THE STENT ALLEGEDLY GOT STUCK IN THE SHEATH AFTER IT WAS DEPLOYED. IT WAS FURTHER REPORTED THAT TO GET THE STENT OUT, THE SHEATH WAS REMOVED WITH THE STENT OVER THE WIRE. THE PROCEDURE WAS COMPLETED BY USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18928710 ↗MalfunctionOct 9, 2023Material Deformation; Material Deformation
IT WAS REPORTED THROUGH THE RESULTS OF A CLINICAL TRIAL, THAT TWENTY-ONE DAYS POST STENT GRAFT PLACEMENT, A CORE LAB ANALYSIS WAS PERFORMED FOR THE STENT DETAILS ANALYSIS AND THE CORE LAB IDENTIFIED THAT THE STENT GRAFT ALLEGEDLY HAD A MODERATE KINKING. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17894060 ↗InjuryAug 10, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THROUGH A CLINICAL STUDY THAT APPROXIMATELY, EIGHT MONTHS AND TEN DAYS, POST STENT PLACEMENT PROCEDURE, THE PATIENT DEVELOPED 70 % OF TARGET LESION STENOSIS. STANDARD PTA WAS USED FOR TREATMENT. THE RE-INTERVENTION PROCEDURE WAS SUCCESSFUL, AND NO NEW ACCESS WAS CREATED. THE CURRENT STATUS OF THE PATIENT IS UNKNOWN.
Patient problems: Restenosis
MAUDE report 15196769 ↗InjuryAug 27, 2021Obstruction of Flow; Obstruction of Flow
IT WAS REPORTED THROUGH THE RESULTS OF A CLINICAL TRIAL, THAT APPROXIMATELY TWO YEARS ONE MONTH POST INDEX PROCEDURE, THE PATIENT DEVELOPED 60% OF TARGET LESION STENOSIS. A STANDARD PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) WAS USED TO SUCCESSFULLY TREAT THE TARGET LESION. THE CURRENT PATIENT STATUS WAS NOT PROVIDED.
Patient problems: Hemorrhage/Blood Loss/Bleeding; Restenosis
MAUDE report 12375998 ↗MalfunctionJul 14, 2020Material Separation; Obstruction of Flow; Material Separation; Obstruction of Flow
THIS REPORT SUMMARIZES ONE MALFUNCTION. A REVIEW OF THE REPORTED INFORMATION INDICATED THAT MODEL AVSM08080 COVERA VASCULAR COVERED STENT SYSTEM ALLEGEDLY EXPERIENCED A MATERIAL SEPARATION/DISSECTION ON IMAGING. THIS INFORMATION WAS RECEIVED FROM ONE SOURCE. THIS MALFUNCTION INVOLVED A PATIENT WITH NO KNOWN IMPACT TO THE PATIENT. THE PATIENT'S AGE, SEX AND WEIGHT WERE UNKNOWN.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10271735 ↗
Other System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment devices
- COVERA AVSM10060Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall33 reports
- COVERA AVSM08060Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall21 reports
- COVERA AVSM09060Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall18 reports
- COVERA AVSM10080Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall16 reports
- COVERA AVSM10040Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall15 reports
- COVERA AVFM10060Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall14 reports
- COVERA AVSM09080Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall1 death14 reports
- COVERA AVSM10100Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall13 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page